Evaluation of two novel methods for the direct and continuous measurement of the intra-abdominal pressure in a porcine model.

Alexander Schachtrupp1, Christian Tons, Volker Fackeldey, Joerg Hoer, Marcus Reinges, Volker Schumpelick.   

Abstract

OBJECTIVE: Intravesical bladder pressure (IVP) measurement is considered to be the gold standard for the assessment of intra-abdominal pressure (IAP). However, this method is indirect, discontinuous, and potentially infectious and relies on a physiological bladder function. This study evaluated two novel methods for direct, continuous IAP measurement. DESIGN AND
SETTING: Experimental study in an animal research laboratory.
SUBJECTS: 18 male domestic pigs.
INTERVENTIONS: CO(2) was insufflated to increase the IAP to 30 mmHg for 18 and 24 h in six animals each. Another six animals served as controls. A piezoresistive (PRM) and an air-capsule (ACM) pressure measurement probe were placed intra-abdominally and of IAP was measured every 1 h (PRM/ACM) or every 2 h (IVP). The mean difference between insufflator readings and IAP values and limits of agreement (mean difference +/-2 SD) were calculated. MEASUREMENTS AND
RESULTS: In the presence of applied pressure IVP and PRM remained significantly below insufflator readings while ACM values showed no difference. Mean difference (and limits of agreement) were 4.5 (-2.1 to 11.1 mmHg), 1.6 (-8.0 to 11.2 mmHg), and 0.5 (-4.5 to 5.4 mmHg) for IVP, PRM, and ACM. The mean measurement-to-measurement drift of the ACM values was 9.0+/-10.2 mmHg.
CONCLUSIONS: In this model agreement of PRM and ACM with insufflator readings was comparable to IVP. As both methods may be advantageous regarding continuous straightforward measurement of IAP, the employment in further experimental and clinical investigations is suggested.

Entities:  

Mesh:

Year:  2003        PMID: 12920511     DOI: 10.1007/s00134-003-1847-3

Source DB:  PubMed          Journal:  Intensive Care Med        ISSN: 0342-4642            Impact factor:   17.440


  14 in total

1.  Estimation of intra-abdominal pressure by bladder pressure measurement: validity and methodology.

Authors:  M A Fusco; R S Martin; M C Chang
Journal:  J Trauma       Date:  2001-02

2.  Experimental evaluation of the Spiegelberg intracranial pressure and intracranial compliance monitor. Technical note.

Authors:  Y H Yau; I R Piper; R E Clutton; I R Whittle
Journal:  J Neurosurg       Date:  2000-12       Impact factor: 5.115

3.  Intra-abdominal pressure: time for clinical practice guidelines?

Authors:  Michael Sugrue
Journal:  Intensive Care Med       Date:  2002-03-20       Impact factor: 17.440

Review 4.  [Abdominal compartment syndrome: prevention and treatment].

Authors:  C Töns; A Schachtrupp; M Rau; T Mumme; V Schumpelick
Journal:  Chirurg       Date:  2000-08       Impact factor: 0.955

5.  A porcine model of the abdominal compartment syndrome.

Authors:  Christian Toens; Alexander Schachtrupp; Joerg Hoer; Karsten Junge; Bernd Klosterhalfen; Violker Schumpelick
Journal:  Shock       Date:  2002-10       Impact factor: 3.454

6.  Determination of intra-abdominal pressure using a transurethral bladder catheter: clinical validation of the technique.

Authors:  T J Iberti; C E Lieber; E Benjamin
Journal:  Anesthesiology       Date:  1989-01       Impact factor: 7.892

7.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

8.  A simplified approach to the diagnosis of elevated intra-abdominal pressure.

Authors:  Steven L Lee; John T Anderson; Eric J Kraut; David H Wisner; Bruce M Wolfe
Journal:  J Trauma       Date:  2002-06

9.  Intra-abdominal pressure measurement using a modified nasogastric tube: description and validation of a new technique.

Authors:  M Sugrue; M D Buist; A Lee; D J Sanchez; K M Hillman
Journal:  Intensive Care Med       Date:  1994-11       Impact factor: 17.440

10.  Abdominal compartment syndrome: does intra-cystic pressure reflect actual intra-abdominal pressure? A prospective study in surgical patients.

Authors: 
Journal:  Crit Care       Date:  1999       Impact factor: 9.097

View more
  21 in total

1.  Measurement of compartment pressure of the rectus sheath during intra-abdominal hypertension in rats.

Authors:  Christoph Meier; René Schramm; Joerg H Holstein; Burkhardt Seifert; Otmar Trentz; Michael D Menger
Journal:  Intensive Care Med       Date:  2006-08-29       Impact factor: 17.440

Review 2.  [Abdominal compartment syndrome: significance, diagnosis and treatment].

Authors:  A Schachtrupp; M Jansen; P Bertram; R Kuhlen; V Schumpelick
Journal:  Anaesthesist       Date:  2006-06       Impact factor: 1.041

3.  Why we need guidelines and recommendations for research on intra-abdominal hypertension.

Authors:  Manu L N G Malbrain; Inneke De Laet
Journal:  Intensive Care Med       Date:  2009-10-20       Impact factor: 17.440

4.  Intra- and interobserver variability during in vitro validation of two novel methods for intra-abdominal pressure monitoring.

Authors:  Tom J R De Potter; Hilde Dits; Manu L N G Malbrain
Journal:  Intensive Care Med       Date:  2005-04-05       Impact factor: 17.440

5.  Results from the International Conference of Experts on Intra-abdominal Hypertension and Abdominal Compartment Syndrome. I. Definitions.

Authors:  Manu L N G Malbrain; Michael L Cheatham; Andrew Kirkpatrick; Michael Sugrue; Michael Parr; Jan De Waele; Zsolt Balogh; Ari Leppäniemi; Claudia Olvera; Rao Ivatury; Scott D'Amours; Julia Wendon; Ken Hillman; Kenth Johansson; Karel Kolkman; Alexander Wilmer
Journal:  Intensive Care Med       Date:  2006-09-12       Impact factor: 17.440

6.  Abdominal compartment syndrome in patients with severe acute pancreatitis in early stage.

Authors:  Hong Chen; Fei Li; Jia-Bang Sun; Jian-Guo Jia
Journal:  World J Gastroenterol       Date:  2008-06-14       Impact factor: 5.742

7.  Direct intra-abdominal pressure monitoring via piezoresistive pressure measurement: a technical note.

Authors:  Jens Otto; Daniel Kaemmer; Marcel Binnebösel; Marc Jansen; Rolf Dembinski; Volker Schumpelick; Alexander Schachtrupp
Journal:  BMC Surg       Date:  2009-04-21       Impact factor: 2.102

Review 8.  [Problematic of intra-abdominal pressure measurement].

Authors:  P Neumann
Journal:  Anaesthesist       Date:  2009-05       Impact factor: 1.041

Review 9.  Different techniques to measure intra-abdominal pressure (IAP): time for a critical re-appraisal.

Authors:  Manu L N G Malbrain
Journal:  Intensive Care Med       Date:  2004-01-17       Impact factor: 17.440

10.  In vitro validation of a novel method for continuous intra-abdominal pressure monitoring.

Authors:  Manu L N G Malbrain; Inneke De laet; Dries Viaene; Karen Schoonheydt; Hilde Dits
Journal:  Intensive Care Med       Date:  2007-12-13       Impact factor: 17.440

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.