| Literature DB >> 12777144 |
R C Tailor1, W F Hanson, G S Ibbott.
Abstract
The Radiological Physics Center (RPC) is a resource to the medical physics community for assistance regarding dosimetry procedures. Since the publication of the AAPM TG-51 calibration protocol, the RPC has responded to numerous phone calls raising questions and describing areas in the protocol where physicists have had problems. At the beginning of the year 2000, the RPC requested that institutions participating in national clinical trials provide the change in measured beam output resulting from the conversion from the TG-21 protocol to TG-51. So far, the RPC has received the requested data from approximately 150 of the approximately 1300 institutions in the RPC program. The RPC also undertook a comparison of TG-21 and TG-51 and determined the expected change in beam calibration for ion chambers in common use, and for the range of photon and electron beam energies used clinically. Analysis of these data revealed two significant outcomes: (i) a large number (approximately 1/2) of the reported calibration changes for photon and electron beams were outside the RPC's expected values, and (ii) the discrepancies in the reported versus the expected dose changes were as large as 8%. Numerous factors were determined to have contributed to these deviations. The most significant factors involved the use of plane-parallel chambers, the mixing of phantom materials and chambers between the two protocols, and the inconsistent use of depth-dose factors for transfer of dose from the measurement depth to the depth of dose maximum. In response to these observations, the RPC has identified a number of circumstances in which physicists might have difficulty with the protocol, including concerns related to electron calibration at low energies (R50<2 cm), and the use of a cylindrical chamber at 6 MeV electrons. In addition, helpful quantitative hints are presented, including the effect of the prescribed lead filter for photon energy measurements, the impact of shifting the chamber depth for photon depth-dose measurements, and the impact of updated stopping-power data used in TG-51 versus that used in TG-21, particularly for electron calibrations. 2003 American College of Medical Physics.Entities:
Mesh:
Year: 2003 PMID: 12777144 PMCID: PMC5724471 DOI: 10.1120/jacmp.v4i2.2524
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Figure 1Ratio of dose calculated by TG‐51 vs the dose calculated by TG‐21 for 163 beams from 151 institutions for electrons with nominal energies of 18 MeV or more. The hatched area represents the ratios expected by the RPC for these energies for most chambers in clinical use.
Figure 4Ratio of dose calculated by TG‐51 vs the dose calculated by TG‐21 for 221 beams from 151 institutions for photons with nominal energy of 6 MV. The hatched area represents the ratios expected by the RPC for these energies for most chambers in clinical use.
Figure 2Ratio of dose calculated by TG‐51 vs the dose calculated by TG‐21 for 166 beams from 151 institutions for electrons with nominal energies of 6 MeV or less. The hatched area represents the ratios expected by the RPC for these energies for most chambers in clinical use.
Figure 3Ratio of dose calculated by TG‐51 vs the dose calculated by TG‐21 for 100 beams from 151 institutions for photons with nominal energies of 18 MV or more. The hatched area represents the ratios expected by the RPC for these energies for most chambers in clinical use.
Error in if lead sheet is not used.
| Nominal MV |
|
|
|
| |
|---|---|---|---|---|---|
| Varian machines | 23 | 80.0 | 1.010 | 1.018 | 1.002 |
| 18 | 79.7 | 1.008 | 1.015 | 1.002 | |
| 10 | 73.4 | 1.004 | 1.009 | 1.001 |
Impact of electron stopping powers (TG‐51 vs TG‐21) on TG‐51 output calibration at due to differences in depth‐dose factor at . Data are provided for typical beam characteristics.
| Nominal energy (MeV) | Beam characteristics | Effective depth (cm) |
| Stopping powers | Ratio of fdd at | ||||
|---|---|---|---|---|---|---|---|---|---|
|
|
| TG51 | TG21 | TG51 | TG21 | ||||
| 5 | 4.59 | 1.96 |
|
| 1.078 | 1.088 | ‐ | ‐ | ‐ |
|
|
| 1.082 | 1.090 | 1.004 | 1.002 | 1.002 | |||
| 12 | 10.91 | 4.76 |
|
| 1.047 | 1.051 | ‐ | ‐ | ‐ |
|
|
| 1.046 | 1.050 | 0.999 | 0.999 | 1.000 | |||
| 20 | 19.15 | 8.40 |
|
| 0.980 | 0.978 | ‐ | ‐ | ‐ |
|
|
| 1.019 | 1.021 | 1.039 | 1.044 | 0.996 | |||
Data from Burns et al.
at at .