| Literature DB >> 11674825 |
S H Cho1, J R Lowenstein, P A Balter, N H Wells, W F Hanson.
Abstract
A new calibration protocol, developed by the AAPM Task Group 51 (TG-51) to replace the TG-21 protocol, is based on an absorbed-dose to water standard and calibration factor (N(D,w)), while the TG-21 protocol is based on an exposure (or air-kerma) standard and calibration factor (N(x)). Because of differences between these standards and the two protocols, the results of clinical reference dosimetry based on TG-51 may be somewhat different from those based on TG-21. The Radiological Physics Center has conducted a systematic comparison between the two protocols, in which photon and electron beam outputs following both protocols were compared under identical conditions. Cylindrical chambers used in this study were selected from the list given in the TG-51 report, covering the majority of current manufacturers. Measured ratios between absorbed-dose and air-kerma calibration factors, derived from the standards traceable to the NIST, were compared with calculated values using the TG-21 protocol. The comparison suggests that there is roughly a 1% discrepancy between measured and calculated ratios. This discrepancy may provide a reasonable measure of possible changes between the absorbed-dose to water determined by TG-51 and that determined by TG-21 for photon beam calibrations. The typical change in a 6 MV photon beam calibration following the implementation of the TG-51 protocol was about 1%, regardless of the chamber used, and the change was somewhat smaller for an 18 MV photon beam. On the other hand, the results for 9 and 16 MeV electron beams show larger changes up to 2%, perhaps because of the updated electron stopping power data used for the TG-51 protocol, in addition to the inherent 1% discrepancy presented in the calibration factors. The results also indicate that the changes may be dependent on the electron energy.Entities:
Mesh:
Substances:
Year: 2000 PMID: 11674825 PMCID: PMC5726169 DOI: 10.1120/jacmp.v1i3.2643
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Makes and models of the chambers used.
| Ion chamber | Serial number | Wall material | Al electrode |
|
|
|---|---|---|---|---|---|
| NEL 2571 | 1503 | graphite | yes | 4.5873 | 4.7331 |
| NEL 2571 | 1864 | graphite | yes | 4.5257 | 4.6799 |
| PTW N23333 | 1516 | PMMA | yes | 5.0913 | 5.2690 |
| PTW N30001 | 1483 | PMMA | yes | 5.2844 | 5.4712 |
| Capintec PR06C | CII.68624 | C‐552 | no | 4.7445 | 4.9290 |
| Exradin A‐12 | 174 | C‐552 | no | 4.9801 | 5.1249 |
and for electron calibration.
| Electron energy (MeV) |
|
|
|---|---|---|
| 9 | 2.0 | 2.1 |
| 16 | 3.8 | 3.4 |
Comparison between absorbed‐dose and air‐kerma calibration factors. Measured ratios in this table are obtained using measured absorbed‐dose and air‐kerma calibration factors from the MDACC ADCL based on standards traceable to NIST. Note and the numerical parameters (e.g., , etc.) for calculations using Eq. (8) are based on the TG‐21 report (Ref. 2).
| Ion chamber | Serial number |
|
| Meas./Calc. |
|---|---|---|---|---|
| NEL 2571 | 1503 | 1.102 | 1.088 | 1.013 |
| NEL 2571 | 1864 | 1.100 | 1.088 | 1.011 |
| PTW N23333 | 1516 | 1.099 | 1.086 | 1.012 |
| PTW N30001 | 1483 | 1.099 | 1.086 | 1.012 |
| Capintec PR06C | CII.68624 | 1.095 | 1.079 | 1.015 |
| Exradin A‐12 | 174 | 1.105 | 1.093 | 1.011 |
Comparison between TG‐51 and TG‐21 calibrations (photon beam). The ratios represent the comparison at 10 cm depth in water, a recommended depth for photon beam calibration in the TG‐51 protocol. Presented results have an estimated uncertainty of less than excluding the inherent uncertainty associated with the calibration factors.
| Ion chamber |
| 6MV (TG‐51/TG‐21) | 18 MV (TG‐51/TG‐21) |
|---|---|---|---|
| NEL 2571 | 1.012 | 1.010 | 1.007 |
| PTW N23333 & N30001 | 1.012 | 1.010 | 1.006 |
| Capintec PR06C | 1.015 | 1.011 | 1.004 |
| Exradin A‐12 | 1.011 | 1.008 | 1.002 |
Comparison between TG‐51 and TG‐21 calibrations (electron beam). Note the ratios represent the comparison at in water for each electron energy. Presented results have an estimated uncertainty of less than excluding the inherent uncertainty associated with the calibration factors.
| Ion chamber | 9 MeV (TG‐51/TG‐21) | 16 MeV (TG‐51/TG‐21) |
|---|---|---|
| NEL 2571 | 1.015 | 1.021 |
| PTW N23333 & N30001 | 1.014 | 1.017 |
| Capintec PR06C | 1.014 | 1.016 |
| Exradin A‐12 | 1.014 | 1.016 |