Literature DB >> 12766063

Fourier analysis of scanning laser polarimetry measurements with variable corneal compensation in glaucoma.

Felipe A Medeiros1, Linda M Zangwill, Christopher Bowd, Antje S Bernd, Robert N Weinreb.   

Abstract

PURPOSE: To apply Fourier analysis to the retinal nerve fiber layer (RNFL) thickness measurements obtained with scanning laser polarimetry (SLP), by using variable corneal compensation, and to evaluate the ability of this method to discriminate glaucomatous from normal eyes.
METHODS: The study included one eye each of 55 patients with glaucoma and 52 healthy subjects. RNFL thickness measurements were obtained with a modified commercial scanning laser polarimeter (GDx Nerve Fiber Analyzer; Laser Diagnostic Technologies, Inc., San Diego, CA) so that corneal birefringence could be corrected on a subject-specific variable basis. The shape of the RNFL thickness double-hump pattern was analyzed by Fourier analysis of polarimetry data. Fourier coefficients and GDx parameters were compared between the two groups. A linear discriminant function was developed to identify and combine the most useful Fourier coefficients to separate the two groups. Receiver operating characteristic (ROC) curves were obtained for each measurement, and sensitivity values (at fixed specificities) were calculated.
RESULTS: The Fourier-based linear discriminant function (LDF Fourier) resulted in a sensitivity of 84% for a specificity set at 92%. For similar specificity, the GDx software-provided parameters had sensitivities ranging from 24% to 69%. The area under ROC curve for the LDF Fourier was 0.949, significantly larger than the ROC curve area for the single best GDx software-provided parameter, superior average (0.870).
CONCLUSIONS: The combination of Fourier RNFL thickness measures in an LDF, obtained using SLP with variable corneal compensation, improved the ability to discriminate glaucomatous from healthy eyes, compared with the GDx software-provided parameters.

Entities:  

Mesh:

Year:  2003        PMID: 12766063     DOI: 10.1167/iovs.02-0814

Source DB:  PubMed          Journal:  Invest Ophthalmol Vis Sci        ISSN: 0146-0404            Impact factor:   4.799


  16 in total

Review 1.  Imaging in glaucoma.

Authors:  Daniel M Stein; Gadi Wollstein; Joel S Schuman
Journal:  Ophthalmol Clin North Am       Date:  2004-03

2.  Repeatability and reproducibility of optic nerve head topography using the retinal thickness analyzer.

Authors:  Esther M Hoffmann; Felipe A Medeiros; Christina Kramann; Norbert Pfeiffer; Franz H Grus
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2005-08-05       Impact factor: 3.117

3.  Automated keratoconus detection using height data of anterior and posterior corneal surfaces.

Authors:  Kenichiro Bessho; Naoyuki Maeda; Teruhito Kuroda; Takashi Fujikado; Yasuo Tano; Tetsuro Oshika
Journal:  Jpn J Ophthalmol       Date:  2006 Sep-Oct       Impact factor: 2.447

4.  GDx-VCC performance in discriminating normal from glaucomatous eyes with early visual field loss.

Authors:  Stefano Da Pozzo; Mirko Fuser; Odilla Vattovani; Giuseppe Di Stefano; Giuseppe Ravalico
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2005-11-16       Impact factor: 3.117

5.  Scanning laser polarimetry with variable corneal compensation and detection of glaucomatous optic neuropathy.

Authors:  Stefano Da Pozzo; Pierluigi Iacono; Roberta Marchesan; Anna Fantin; Giuseppe Ravalico
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2005-03-09       Impact factor: 3.117

6.  The relationship between optical coherence tomography and scanning laser polarimetry measurements in glaucoma.

Authors:  Yun Suk Chung; Yong Ho Sohn
Journal:  Korean J Ophthalmol       Date:  2006-12

7.  Impact of atypical retardation patterns on detection of glaucoma progression using the GDx with variable corneal compensation.

Authors:  Felipe A Medeiros; Luciana M Alencar; Linda M Zangwill; Pamela A Sample; Remo Susanna; Robert N Weinreb
Journal:  Am J Ophthalmol       Date:  2009-04-17       Impact factor: 5.258

Review 8.  Diagnostic tools for glaucoma detection and management.

Authors:  Pooja Sharma; Pamela A Sample; Linda M Zangwill; Joel S Schuman
Journal:  Surv Ophthalmol       Date:  2008-11       Impact factor: 6.048

9.  Optic nerve head and retinal nerve fiber layer analysis: a report by the American Academy of Ophthalmology.

Authors:  Shan C Lin; Kuldev Singh; Henry D Jampel; Elizabeth A Hodapp; Scott D Smith; Brian A Francis; David K Dueker; Robert D Fechtner; John S Samples; Joel S Schuman; Don S Minckler
Journal:  Ophthalmology       Date:  2007-10       Impact factor: 12.079

10.  Structural and functional relationships in glaucoma using standard automated perimetry and the Humphrey Matrix.

Authors:  Seong Bae Park; Yoon Pyo Nam; Kyung Rim Sung; Michael S Kook
Journal:  Korean J Ophthalmol       Date:  2009-09-08
View more