Repeatability and reproducibility of optic nerve head topography using the retinal thickness analyzer.

Esther M Hoffmann1, Felipe A Medeiros, Christina Kramann, Norbert Pfeiffer, Franz H Grus.   

Abstract

OBJECTIVE: The objective was to determine the reproducibility and repeatability of topographic disk parameters with the retinal thickness analyzer (RTA).
METHODS: Thirty-nine eyes of 39 healthy subjects were included. Three observers performed three consecutive measurements of optic disk topography independently. Combined repeatability (intra-observer component of variability) and reproducibility (inter-observer component of variability, R and R) was assessed by variance component analysis.
RESULTS: Six topographic parameters (disk area, cup area, cup volume, mean cup depth, maximum depth, and cup shape measure) revealed an acceptable repeatability and reproducibility expressed as the percentage of the total measurement variability (range 17.6-28.4%). The average reproducibility and repeatability of all optic disk parameters measured using the RTA was not acceptable (combined R and R = 33.4%).
CONCLUSIONS: Repeatability and reproducibility of the optic disk topography analysis using the RTA reveals a high variation, and should, therefore, be interpreted carefully for glaucoma diagnosis and for monitoring the disease.

Entities:  

Mesh:

Year:  2005        PMID: 16082558     DOI: 10.1007/s00417-005-0027-2

Source DB:  PubMed          Journal:  Graefes Arch Clin Exp Ophthalmol        ISSN: 0721-832X            Impact factor:   3.117


  34 in total

1.  Application of rapid scanning retinal thickness analysis in retinal diseases.

Authors:  S Asrani; R Zeimer; M F Goldberg; S Zou
Journal:  Ophthalmology       Date:  1997-07       Impact factor: 12.079

2.  Reproducibility of nerve fiber layer thickness measurements by use of optical coherence tomography.

Authors:  E Z Blumenthal; J M Williams; R N Weinreb; C A Girkin; C C Berry; L M Zangwill
Journal:  Ophthalmology       Date:  2000-12       Impact factor: 12.079

3.  Validity of rim area measurements by different reference planes.

Authors:  James C H Tan; Edward White; Darmalingum Poinoosawmy; Roger A Hitchings
Journal:  J Glaucoma       Date:  2004-06       Impact factor: 2.503

4.  Reproducibility of nerve fiber thickness, macular thickness, and optic nerve head measurements using StratusOCT.

Authors:  Lelia A Paunescu; Joel S Schuman; Lori Lyn Price; Paul C Stark; Siobahn Beaton; Hiroshi Ishikawa; Gadi Wollstein; James G Fujimoto
Journal:  Invest Ophthalmol Vis Sci       Date:  2004-06       Impact factor: 4.799

5.  A new method for rapid mapping of the retinal thickness at the posterior pole.

Authors:  R Zeimer; M Shahidi; M Mori; S Zou; S Asrani
Journal:  Invest Ophthalmol Vis Sci       Date:  1996-09       Impact factor: 4.799

6.  Detection of glaucoma with scanning laser polarimetry.

Authors:  R N Weinreb; L Zangwill; C C Berry; R Bathija; P A Sample
Journal:  Arch Ophthalmol       Date:  1998-12

7.  Comparison of optic disc features in low-tension and typical open-angle glaucoma.

Authors:  K M Miller; H A Quigley
Journal:  Ophthalmic Surg       Date:  1987-12

Review 8.  Optical coherence tomography: a new tool for glaucoma diagnosis.

Authors:  J S Schuman; M R Hee; A V Arya; T Pedut-Kloizman; C A Puliafito; J G Fujimoto; E A Swanson
Journal:  Curr Opin Ophthalmol       Date:  1995-04       Impact factor: 3.761

9.  The mode of progressive disc cupping in ocular hypertension and glaucoma.

Authors:  J E Pederson; D R Anderson
Journal:  Arch Ophthalmol       Date:  1980-03

10.  Differentiating patients with glaucoma from glaucoma suspects and normal subjects by nerve fiber layer assessment with scanning laser polarimetry.

Authors:  N T Choplin; D C Lundy; A W Dreher
Journal:  Ophthalmology       Date:  1998-11       Impact factor: 12.079

View more
  1 in total

1.  Comparison of optic nerve head topography in healthy adults using a Heidelberg retina tomograph and retinal thickness analyzer.

Authors:  Andreja Rekic; Marjanca Breznik; Barbara Cvenkel
Journal:  Int Ophthalmol       Date:  2007-04-11       Impact factor: 2.029

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.