BACKGROUND:Tibial fixation is more problematic than femoral fixation in anterior cruciate ligament reconstruction. HYPOTHESIS: There is no difference in initial fixation strength among hamstring tendon graft tibial fixation devices. STUDY DESIGN: Randomized experimental study. METHODS: Each of six devices used to fix 120 quadrupled human semitendinosus-gracilis tendon grafts into porcine tibiae was tested 10 times with a single-cycle load-to-failure test and 10 times with a 1500-cycle loading test. Specimens surviving cyclic loading were subjected to a single-cycle load-to-failure test. RESULTS: Intrafix (1332 N) was the strongest in the single-cycle load-to-failure test, followed by WasherLoc (975 N), tandem spiked washer (769 N), SmartScrew ACL (665 N), BioScrew (612 N), and SoftSilk (471 N). After cyclic-loading tests, Intrafix showed the lowest residual displacement (1.5 mm) and was also strongest (1309 N) in the single-cycle load-to-failure test after the cyclic-loading test, followed by WasherLoc (3.2 mm; 917 N). CONCLUSION: The Intrafix provided clearly superior strength in the fixation of hamstring tendon grafts to the tibial drill hole. CLINICAL RELEVANCE: Some caution may be warranted when using the implants that showed increased residual displacement, especially if aggressive rehabilitation is to be used. Preconditioning of the hamstring tendon graft-implant complex before tibial fixation is needed. Copyright 2003 American Orthopaedic Society for Sports Medicine
RCT Entities:
BACKGROUND: Tibial fixation is more problematic than femoral fixation in anterior cruciate ligament reconstruction. HYPOTHESIS: There is no difference in initial fixation strength among hamstring tendon graft tibial fixation devices. STUDY DESIGN: Randomized experimental study. METHODS: Each of six devices used to fix 120 quadrupled human semitendinosus-gracilis tendon grafts into porcine tibiae was tested 10 times with a single-cycle load-to-failure test and 10 times with a 1500-cycle loading test. Specimens surviving cyclic loading were subjected to a single-cycle load-to-failure test. RESULTS: Intrafix (1332 N) was the strongest in the single-cycle load-to-failure test, followed by WasherLoc (975 N), tandem spiked washer (769 N), SmartScrew ACL (665 N), BioScrew (612 N), and SoftSilk (471 N). After cyclic-loading tests, Intrafix showed the lowest residual displacement (1.5 mm) and was also strongest (1309 N) in the single-cycle load-to-failure test after the cyclic-loading test, followed by WasherLoc (3.2 mm; 917 N). CONCLUSION: The Intrafix provided clearly superior strength in the fixation of hamstring tendon grafts to the tibial drill hole. CLINICAL RELEVANCE: Some caution may be warranted when using the implants that showed increased residual displacement, especially if aggressive rehabilitation is to be used. Preconditioning of the hamstring tendon graft-implant complex before tibial fixation is needed. Copyright 2003 American Orthopaedic Society for Sports Medicine
Authors: M Jagodzinski; M Ettinger; C Haasper; S Hankemeier; D Breitmeier; C Hurschler; C Krettek Journal: Unfallchirurg Date: 2010-07 Impact factor: 1.000
Authors: M Jagodzinski; K Scheunemann; K Knobloch; K Albrecht; C Krettek; C Hurschler; J Zeichen Journal: Knee Surg Sports Traumatol Arthrosc Date: 2006-06-09 Impact factor: 4.342
Authors: Dietmar Krappinger; Franz Sebastian Kralinger; Rene El Attal; Wolfgang Hackl; Christian Haid Journal: Knee Surg Sports Traumatol Arthrosc Date: 2006-08-15 Impact factor: 4.342
Authors: Andrew R Duffee; Jeffrey A Brunelli; John Nyland; Robert Burden; Akbar Nawab; David Caborn Journal: Knee Surg Sports Traumatol Arthrosc Date: 2006-09-09 Impact factor: 4.342