Mechanical testing of absorbable suture anchors.

Dominik C Meyer1, Sandro F Fucentese, Kurt Ruffieux, Hilaire A c Jacob, Christian Gerber.   

Abstract

PURPOSE: Absorbable suture anchors offer great advantages but are made of mechanically weak material. The weakest link in the fixation of soft tissue to bone may therefore be the anchor itself. In this study, several commercially available anchors were mechanically tested in vitro. TYPE OF STUDY: Biomechanical bench study.
METHODS: Twelve absorbable suture anchor models were implanted into an artificial test bone according to the recommended technique. Testing temperature was 37 degrees C +/- 1 degrees C. The anchors were loaded with an Instron testing machine with the suture material (USP No. 2, Ethibond, Ethicon, Somerville, NJ) in line with the anchor axis, with and without previous abrasion of the suture at the eyelet. Tensile load at failure and failure mode were recorded. To test creep behavior, a permanent load of 100 N was applied to the anchors, and time to failure was recorded. Suture anchor weight and crystallinity were analyzed.
RESULTS: Mean failure load on tensile testing using a cross-head speed of 60 mm/min ranged from 124 to 244 N. Failure modes were eyelet failure in 5 cases, suture failure in 6 cases, and anchor pullout in 1 case. In creep testing, eyelet failure occurred in 8 anchor models after a mean duration of 0.5 to 99 hours; 3 anchor models remained intact after 300 hours, and 1 anchor model failed by pullout of the test sample. Crystallinity ranged from 0% (amorphous) to 57.2%; anchor weight ranged from 0.036 to 0.161 g. Mechanical properties did not correlate with crystallinity but with anchor weight. Abrasion of the suture material at the eyelet had little effect on failure load.
CONCLUSIONS: At 37 degrees C, structural failure (breaking) of absorbable suture anchors may occur if loaded to the mechanical limit. Absorbable anchors are particularly sensitive to static, long-term loading.

Entities:  

Mesh:

Substances:

Year:  2003        PMID: 12579152     DOI: 10.1053/jars.2003.50015

Source DB:  PubMed          Journal:  Arthroscopy        ISSN: 0749-8063            Impact factor:   4.772


  9 in total

1.  The effect of two nonresorbable suture types on the mechanical performance over a metal suture anchor eyelet.

Authors:  D Acton; A Perry; R Evans; A Butler; P Stephens; W Bruce; J Goldberg; D Sonnabend; W R Walsh
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2003-09-12       Impact factor: 4.342

2.  [Primary stability of the capsule-labrum complex after reconstruction with the Mitek Bioknotless anchor system in human cadaver models].

Authors:  E Erdeljac; E Steinhauser; U Schreiber; A B Imhoff
Journal:  Unfallchirurg       Date:  2007-01       Impact factor: 1.000

3.  Cyclic loading comparison of Bio-SutureTak-#2 FiberWire and Bio Mini-Revo-#2 Hi-Fi suture anchor-sutures in cadaveric scapulae.

Authors:  Brad S Sparks; John Nyland; Akbar Nawab; Ethan Blackburn; Ryan Krupp; Robert Burden
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2007-11-20       Impact factor: 4.342

4.  CORR Insights®: Biocomposite Suture Anchors Remain Visible Two Years After Rotator Cuff Repair.

Authors:  Xavier Ampuero Duralde
Journal:  Clin Orthop Relat Res       Date:  2019-06       Impact factor: 4.176

5.  Pullout strength of suture anchors: effect of mechanical properties of trabecular bone.

Authors:  Mariya Poukalova; Christopher M Yakacki; Robert E Guldberg; Angela Lin; Minn Saing; Scott D Gillogly; Ken Gall
Journal:  J Biomech       Date:  2010-02-01       Impact factor: 2.712

6.  Mechanical properties of all-suture anchors for rotator cuff repair.

Authors:  N S Nagra; N Zargar; R D J Smith; A J Carr
Journal:  Bone Joint Res       Date:  2017-02       Impact factor: 5.853

7.  Initial Experience With a Bioresorbable Polymer Anchor.

Authors:  Robert R Burnham; Jayanth Kumar; Michael Pinzur; Adam Schiff
Journal:  Cureus       Date:  2020-12-29

8.  Outcomes of Operative Management of Insertional Achilles Tendinopathy in the Young vs Elderly.

Authors:  Huai Ming Phen; Wesley J Manz; Danielle Mignemi; Joel T Greenshields; Jason T Bariteau
Journal:  Foot Ankle Orthop       Date:  2020-07-30

9.  Knotless PEEK and double-loaded biodegradable suture anchors ensure comparable clinical outcomes in the arthroscopic treatment of traumatic anterior shoulder instability: a prospective randomized study.

Authors:  Maristella F Saccomanno; Simone Cerciello; Marco Adriani; Marcello Motta; Almerico Megaro; Stefano Galli; Alessandra Scaini; Giuseppe Milano
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2022-04-18       Impact factor: 4.114

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.