Factors affecting the pullout strength of self-drilling and self-tapping anterior cervical screws.

Patrick W Hitchon1, Matthew D Brenton, Justin K Coppes, Aaron M From, James C Torner.   

Abstract

STUDY
DESIGN: A biomechanical comparison of 12-, 14-, and 16-mm self-tapping and self-drilling screws in cadaveric cervical spines was performed.
OBJECTIVE: This study compared self-drilling and self-tapping screw designs used in anterior cervical locking plates in terms of insertion torque and pullout strength. In addition to screw design, the effects of screw length (12, 14, and 16 mm) and bone mineral density were examined.
METHODS: A total of 201 screws were implanted and tested. Nineteen cadaver spine specimens (C2 through C7) were radiographed and their bone mineral density measured. In each vertebral level, two screws of the same length were implanted and their insertion torque measured. One of each screw design (self-drilling or self-tapping) was randomly placed in each vertebra for side-by-side comparison. Vertebrae then were individually potted and mounted on an MTS machine for accurate measurement of maximum pullout strength.
RESULTS: Pullout strength was strongly related to screw length ( = 0.0002). There was no significant difference in pullout strength between the self-drilling and self-tapping screw designs at any length ( = 0.9352). A significant correlation ( </= 0.0077) exists between insertion torque and bone mineral density for the self-drilling and self-tapping 14- and 16-mm screws. A significant correlation exists between pullout strength and bone mineral density for 14- and 16-mm screw lengths of both screw types ( < 0.0001).
CONCLUSIONS: It would appear that the longest screw feasible should always be used clinically for maximum pullout strength. There is no difference in pullout strength between self-drilling and self-tapping screws. Bone mineral density significantly influences screw performance.

Mesh:

Year:  2003        PMID: 12544947     DOI: 10.1097/00007632-200301010-00004

Source DB:  PubMed          Journal:  Spine (Phila Pa 1976)        ISSN: 0362-2436            Impact factor:   3.468


  28 in total

1.  Morphological and structural characteristics of orthodontic mini-implants.

Authors:  Saeed AlSamak; Elias Bitsanis; Margarita Makou; George Eliades
Journal:  J Orofac Orthop       Date:  2012-01-12       Impact factor: 1.938

2.  Experimental in vivo acute and chronic biomechanical and histomorphometrical comparison of self-drilling and self-tapping anterior cervical screws.

Authors:  Maximiliano Aguiar Porto; Patrícia Silva; Rodrigo Rosa; José Batista Volpon; Antônio Carlos Shimano; Francisco José Albuquerque de Paula; Helton Luiz Aparecido Defino
Journal:  Eur Spine J       Date:  2011-12-16       Impact factor: 3.134

3.  [Spinal column: implants and revisions].

Authors:  S M Krieg; H S Meyer; B Meyer
Journal:  Chirurg       Date:  2016-03       Impact factor: 0.955

4.  Maintenance of graft compression in the adult cervical spine.

Authors:  Ciaran Bolger; Maurice Bourlion; Xavier Leroy; Dominique Petit; Gerard Vanacker; Linda McEvoy; Jabir Nagaria
Journal:  Eur Spine J       Date:  2006-01-19       Impact factor: 3.134

5.  Biomechanical study of anterior spinal instrumentation configurations.

Authors:  Luc P Cloutier; Carl-Eric Aubin; Guy Grimard
Journal:  Eur Spine J       Date:  2007-01-05       Impact factor: 3.134

6.  Cervical anterior transpedicular screw fixation (ATPS)--Part II. Accuracy of manual insertion and pull-out strength of ATPS.

Authors:  Heiko Koller; Frank Acosta; Mark Tauber; Michael Fox; Hudelmaier Martin; Rosmarie Forstner; Peter Augat; Rainer Penzkofer; Christian Pirich; H Kässmann; Herbert Resch; Wolfgang Hitzl
Journal:  Eur Spine J       Date:  2008-01-26       Impact factor: 3.134

7.  Effect of surgical factors on the augmentation of cement-injectable cannulated pedicle screw fixation by a novel calcium phosphate-based nanocomposite.

Authors:  Haolin Sun; Chun Liu; Shunlun Chen; Yanjie Bai; Huilin Yang; Chunde Li; Lei Yang
Journal:  Front Med       Date:  2019-09-23       Impact factor: 4.592

8.  Pullout Strength After Multiple Reinsertions in Radial Bone Fixation.

Authors:  Alfonso Mejia; Giovanni Solitro; Elena Gonzalez; Amit Parekh; Mark Gonzalez; Farid Amirouche
Journal:  Hand (N Y)       Date:  2018-09-06

9.  Biomechanics of Cervical "Skip" Corpectomy Versus Standard Multilevel Corpectomy.

Authors:  Murat Yilmaz; Kasim Zafer Yüksel; Seungwon Baek; Anna G U S Newcomb; Sedat Dalbayrak; Volker K H Sonntag; Neil R Crawford
Journal:  Clin Spine Surg       Date:  2017-04       Impact factor: 1.876

10.  Serious complication of cement augmentation for damaged pilot hole.

Authors:  Moon Young Jung; Dong Ah Shin; In Bo Hahn; Tae Gon Kim; Ryoong Huh; Sang Sup Chung
Journal:  Yonsei Med J       Date:  2010-05       Impact factor: 2.759

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.