RF heating due to conductive wires during MRI depends on the phase distribution of the transmit field.

Christopher J Yeung1, Robert C Susil, Ergin Atalar.   

Abstract

In many studies concerning wire heating during MR imaging, a "resonant wire length" that maximizes RF heating is determined. This may lead to the nonintuitive conclusion that adding more wire, so as to avoid this resonant length, will actually improve heating safety. Through a theoretical analysis using the method of moments, we show that this behavior depends on the phase distribution of the RF transmit field. If the RF transmit field has linear phase, with slope equal to the real part of the wavenumber in the tissue, long wires always heat more than short wires. In order to characterize the intrinsic safety of a device without reference to a specific body coil design, this maximum-tip heating phase distribution must be considered. Finally, adjusting the phase distribution of the electric field generated by an RF transmit coil may lead to an "implant-friendly" coil design. Copyright 2002 Wiley-Liss, Inc.

Mesh:

Year:  2002        PMID: 12465125     DOI: 10.1002/mrm.10310

Source DB:  PubMed          Journal:  Magn Reson Med        ISSN: 0740-3194            Impact factor:   4.668


  28 in total

1.  [Magnetic resonance imaging and implantable cardiac devices. Current status and future perspectives of MR-compatible systems].

Authors:  M Dorenkamp; M Roser; B Hamm; W Haverkamp
Journal:  Herz       Date:  2012-03       Impact factor: 1.443

2.  RF Heating of MRI-Assisted Catheter Steering Coils for Interventional MRI.

Authors:  Fabio Settecase; Steven W Hetts; Alastair J Martin; Timothy P L Roberts; Anthony F Bernhardt; Lee Evans; Vincent Malba; Maythem Saeed; Ronald L Arenson; Walter Kucharzyk; Mark W Wilson
Journal:  Acad Radiol       Date:  2010-11-13       Impact factor: 3.173

3.  Steerable catheter microcoils for interventional MRI reducing resistive heating.

Authors:  Anthony Bernhardt; Mark W Wilson; Fabio Settecase; Leland Evans; Vincent Malba; Alastair J Martin; Maythem Saeed; Timothy P L Roberts; Ronald L Arenson; Steven W Hetts
Journal:  Acad Radiol       Date:  2010-11-13       Impact factor: 3.173

Review 4.  Real-time, Interactive MRI for cardiovascular interventions.

Authors:  Elliot R McVeigh; Michael A Guttman; Peter Kellman; Amish N Raval; Robert J Lederman
Journal:  Acad Radiol       Date:  2005-09       Impact factor: 3.173

5.  Real-time magnetic resonance-guided endovascular repair of experimental abdominal aortic aneurysm in swine.

Authors:  Venkatesh K Raman; Parag V Karmarkar; Michael A Guttman; Alexander J Dick; Dana C Peters; Cengizhan Ozturk; Breno S S Pessanha; Richard B Thompson; Amish N Raval; Ranil DeSilva; Ronnier J Aviles; Ergin Atalar; Elliot R McVeigh; Robert J Lederman
Journal:  J Am Coll Cardiol       Date:  2005-06-21       Impact factor: 24.094

6.  Control of intravascular catheters using an array of active steering coils.

Authors:  N Gudino; J A Heilman; J J Derakhshan; J L Sunshine; J L Duerk; M A Griswold
Journal:  Med Phys       Date:  2011-07       Impact factor: 4.071

7.  Parallel transmit excitation at 1.5 T based on the minimization of a driving function for device heating.

Authors:  N Gudino; M Sonmez; Z Yao; T Baig; S Nielles-Vallespin; A Z Faranesh; R J Lederman; M Martens; R S Balaban; M S Hansen; M A Griswold
Journal:  Med Phys       Date:  2015-01       Impact factor: 4.071

8.  An RF dosimeter for independent SAR measurement in MRI scanners.

Authors:  Di Qian; Abdel-Monem M El-Sharkawy; Paul A Bottomley; William A Edelstein
Journal:  Med Phys       Date:  2013-12       Impact factor: 4.071

9.  In-vitro mapping of E-fields induced near pacemaker leads by simulated MR gradient fields.

Authors:  Howard I Bassen; Gonzalo G Mendoza
Journal:  Biomed Eng Online       Date:  2009-12-15       Impact factor: 2.819

10.  Whole shaft visibility and mechanical performance for active MR catheters using copper-nitinol braided polymer tubes.

Authors:  Ozgur Kocaturk; Christina E Saikus; Michael A Guttman; Anthony Z Faranesh; Kanishka Ratnayaka; Cengizhan Ozturk; Elliot R McVeigh; Robert J Lederman
Journal:  J Cardiovasc Magn Reson       Date:  2009-08-12       Impact factor: 5.364

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.