Validation study of two-microphone acoustic reflectometry for determination of breathing tube placement in 200 adult patients.

David T Raphael1, Maxim Benbassat, Dimiter Arnaudov, Alex Bohorquez, Bita Nasseri.   

Abstract

BACKGROUND: Acoustic reflectometry allows the construction of a one-dimensional image of a cavity, such as the airway or the esophagus. The reflectometric area-distance profile consists of a constant cross-sectional area segment (length of endotracheal tube), followed either by a rapid increase in the area beyond the carina (tracheal intubation) or by an immediate decrease in the area (esophageal intubation).
METHODS: Two hundred adult patients were induced and intubated, without restrictions on anesthetic agents or airway adjunct devices. A two-microphone acoustic reflectometer was used to determine whether the breathing tube was placed in the trachea or esophagus. A blinded reflectometer operator, seated a distance away from the patient, interpreted the acoustic area-distance profile alone to decide where the tube was placed. Capnography was used as the gold standard.
RESULTS: Of 200 tracheal intubations confirmed by capnography, the reflectometer operator correctly identified 198 (99% correct tracheal intubation identification rate). In two patients there were false-negative results, patients with a tracheal intubation were interpreted as having an esophageal intubation. A total of 14 esophageal intubations resulted, all correctly identified by reflectometry, for a 100% esophageal intubation identification rate.
CONCLUSIONS: Acoustic reflectometry is a rapid, noninvasive method by which to determine whether breathing tube placement is correct (tracheal) or incorrect (esophageal). Reflectometry determination of tube placement may be useful in airway emergencies, particularly in cases where visualization of the glottic area is not possible and capnography may fail, as in patients with cardiac arrest.

Entities:  

Mesh:

Year:  2002        PMID: 12459661     DOI: 10.1097/00000542-200212000-00007

Source DB:  PubMed          Journal:  Anesthesiology        ISSN: 0003-3022            Impact factor:   7.892


  7 in total

1.  A novel airway device with tactile sensing capabilities for verifying correct endotracheal tube placement.

Authors:  Pauwel Goethals; Harshu Chaobal; Dominiek Reynaerts; David Schaner
Journal:  J Clin Monit Comput       Date:  2013-09-25       Impact factor: 2.502

2.  Breath sound changes associated with malpositioned endotracheal tubes.

Authors:  H A Mansy; C J O'Connor; R A Balk; R H Sandler
Journal:  Med Biol Eng Comput       Date:  2005-03       Impact factor: 2.602

3.  Endotracheal tube intraluminal diameter narrowing after mechanical ventilation: use of acoustic reflectometry.

Authors:  M C Boqué; B Gualis; A Sandiumenge; J Rello
Journal:  Intensive Care Med       Date:  2004-10-02       Impact factor: 17.440

4.  Detection of endobronchial intubation by monitoring the CO2 level above the endotracheal cuff.

Authors:  Shai Efrati; Israel Deutsch; Nathan Weksler; Gabriel M Gurman
Journal:  J Clin Monit Comput       Date:  2014-05-29       Impact factor: 2.502

5.  Endobronchial intubation detected by insertion depth of endotracheal tube, bilateral auscultation, or observation of chest movements: randomised trial.

Authors:  Christian Sitzwohl; Angelika Langheinrich; Andreas Schober; Peter Krafft; Daniel I Sessler; Harald Herkner; Christopher Gonano; Christian Weinstabl; Stephan C Kettner
Journal:  BMJ       Date:  2010-11-09

6.  Sound analysis in an in vitro endotracheal tube model.

Authors:  Young Sik Park; Young Wook Kee; Kwang Suk Park; Jinwoo Lee; Sang-Min Lee; Jae-Joon Yim; Chul-Gyu Yoo; Young Whan Kim; Sung Koo Han; Seok-Chul Yang
Journal:  Korean J Intern Med       Date:  2011-11-28       Impact factor: 2.884

7.  Reliability of Ultrasonography in Confirming Endotracheal Tube Placement in an Emergency Setting.

Authors:  Vimal Koshy Thomas; Cherish Paul; Punchalil Chathappan Rajeev; Babu Urumese Palatty
Journal:  Indian J Crit Care Med       Date:  2017-05
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.