Endotracheal tube intraluminal diameter narrowing after mechanical ventilation: use of acoustic reflectometry.

M C Boqué1, B Gualis, A Sandiumenge, J Rello.   

Abstract

OBJECTIVE: To quantify the incidence and degree of endotracheal tube intraluminal obstruction after mechanical ventilation and its relation to time of intubation.
DESIGN: Prospective observational study.
SETTING: A 14-bed medical-surgical intensive care unit at a university-affiliated teaching hospital. PATIENTS: Ninety-four endotracheal tubes used in 80 patients requiring mechanical ventilation for more than 12 h. INTERVENTIONS AND
RESULTS: Acoustic reflectometry was performed in every endotracheal tube after patient extubation to measure its volume reduction. The intraluminal volumes of used endotracheal tubes in mechanically ventilated patients were significantly lower than those of unused tubes of the same size (5.52+/-0.92 ml(3) versus 6.54+/-0.79 ml(3), p<0.05). The mean difference in endotracheal tube segment volumes was 15.2% (range 0-66%). Volume reduction was above 10% in 60.8% of the tubes. In 22% of endotracheal tubes the remaining inner diameter was less than 7 mm. Reduction below this figure was less frequent (9.3%) in tubes 8 mm or more (p<0.05). The percentage of endotracheal tube volume reduction was not associated with the duration of intubation (r=-0.09, p= n.s.) Peak pressure measured before extubation did not predict obstruction (r=0.11, p= n.s.).
CONCLUSIONS: Inadvertent endotracheal tube obstruction was common in patients requiring mechanical ventilation and may be significant as early as at 24 h. Moderate obstruction in endotracheal tube lumens should be suspected in cases of difficulties in weaning, even in patients who were ventilated for less than 1 day.

Entities:  

Mesh:

Year:  2004        PMID: 15580475     DOI: 10.1007/s00134-004-2465-4

Source DB:  PubMed          Journal:  Intensive Care Med        ISSN: 0342-4642            Impact factor:   17.440


  31 in total

1.  Reduction in tracheal lumen due to endotracheal intubation and its calculated clinical significance.

Authors:  K R Bock; P Silver; M Rom; M Sagy
Journal:  Chest       Date:  2000-08       Impact factor: 9.410

2.  Randomized clinical trial of extended use of a hydrophobic condenser humidifier: 1 vs. 7 days.

Authors:  Laurent Thomachot; Marc Leone; Karim Razzouk; François Antonini; Renaud Vialet; Claude Martin
Journal:  Crit Care Med       Date:  2002-01       Impact factor: 7.598

3.  Detection of positional airway obstruction in neonates by acoustic reflection.

Authors:  P H Jarreau; B Louis; L Desfrère; P W Blanchard; D Isabey; A Harf; G Moriette
Journal:  Am J Respir Crit Care Med       Date:  2000-05       Impact factor: 21.405

4.  In vitro versus in vivo comparison of endotracheal tube airflow resistance.

Authors:  P E Wright; J J Marini; G R Bernard
Journal:  Am Rev Respir Dis       Date:  1989-07

5.  Work of breathing through different sized endotracheal tubes.

Authors:  M Shapiro; R K Wilson; G Casar; K Bloom; R B Teague
Journal:  Crit Care Med       Date:  1986-12       Impact factor: 7.598

6.  Noninvasive estimate of work of breathing due to the endotracheal tube.

Authors:  L Heyer; B Louis; D Isabey; F Lofaso; L Brochard; J J Fredberg; A Harf
Journal:  Anesthesiology       Date:  1996-12       Impact factor: 7.892

7.  A comparison of four methods of weaning patients from mechanical ventilation. Spanish Lung Failure Collaborative Group.

Authors:  A Esteban; F Frutos; M J Tobin; I Alía; J F Solsona; I Valverdú; R Fernández; M A de la Cal; S Benito; R Tomás
Journal:  N Engl J Med       Date:  1995-02-09       Impact factor: 91.245

8.  Influence of gender and endotracheal tube size on preextubation breathing pattern.

Authors:  S K Epstein; R L Ciubotaru
Journal:  Am J Respir Crit Care Med       Date:  1996-12       Impact factor: 21.405

9.  Long-term effects of different humidification systems on endotracheal tube patency: evaluation by the acoustic reflection method.

Authors:  Samir Jaber; Jérôme Pigeot; Redouane Fodil; Salvatore Maggiore; Alain Harf; Daniel Isabey; Laurent Brochard; Bruno Louis
Journal:  Anesthesiology       Date:  2004-04       Impact factor: 7.892

10.  Comparing two heat and moisture exchangers, one hydrophobic and one hygroscopic, on humidifying efficacy and the rate of nosocomial pneumonia.

Authors:  L Thomachot; X Viviand; S Arnaud; C Boisson; C D Martin
Journal:  Chest       Date:  1998-11       Impact factor: 9.410

View more
  13 in total

1.  Antibacterial-coated tracheal tubes cleaned with the Mucus Shaver : a novel method to retain long-term bactericidal activity of coated tracheal tubes.

Authors:  Lorenzo Berra; Francesco Curto; Gianluigi Li Bassi; Patrice Laquerriere; Andrea Baccarelli; Theodor Kolobow
Journal:  Intensive Care Med       Date:  2006-04-19       Impact factor: 17.440

2.  Abnormally increased power of breathing as a complication of closed endotracheal suction catheter systems.

Authors:  Mehmet S Ozcan; Steven W Bonett; A Daniel Martin; Andrea Gabrielli; A Joseph Layon; Michael J Banner
Journal:  Respir Care       Date:  2006-04       Impact factor: 2.258

3.  Sound level analysis in endotracheal tube obstruction in spontaneous breathing and mechanical ventilation-an animal model study.

Authors:  Behzad Maghsoodi; Golnar Sabetian; Aram Azimi; Nader Tanideh; Alireza Mehdizade
Journal:  J Clin Monit Comput       Date:  2016-12-26       Impact factor: 2.502

4.  Comparison of the effects of two humidifier systems on endotracheal tube resistance.

Authors:  Indalecio Morán; Belén Cabello; Elizabeth Manero; Jordi Mancebo
Journal:  Intensive Care Med       Date:  2011-09-27       Impact factor: 17.440

5.  Reduction of Endotracheal Tube Biofilms Using Antimicrobial Photodynamic Therapy.

Authors:  Merrill A Biel; Chet Sievert; Marina Usacheva; Matthew Teichert; Eric Wedell; Nicolas Loebel; Andreas Rose; Ron Zimmermann
Journal:  Lasers Surg Med       Date:  2011-09-01       Impact factor: 4.025

6.  A clinical assessment of the Mucus Shaver: a device to keep the endotracheal tube free from secretions.

Authors:  Lorenzo Berra; Andrea Coppadoro; Edward A Bittner; Theodor Kolobow; Patrice Laquerriere; Joshua R Pohlmann; Simone Bramati; Joel Moss; Antonio Pesenti
Journal:  Crit Care Med       Date:  2012-01       Impact factor: 7.598

7.  Tracheal pressure and endotracheal tube obstruction can be detected by continuous cuff pressure monitoring: in vitro pilot study.

Authors:  Shai Efrati; Israel Deutsch; Gabriel M Gurman; Matitiau Noff; Giorgio Conti
Journal:  Intensive Care Med       Date:  2010-03-16       Impact factor: 17.440

8.  Sound analysis in an in vitro endotracheal tube model.

Authors:  Young Sik Park; Young Wook Kee; Kwang Suk Park; Jinwoo Lee; Sang-Min Lee; Jae-Joon Yim; Chul-Gyu Yoo; Young Whan Kim; Sung Koo Han; Seok-Chul Yang
Journal:  Korean J Intern Med       Date:  2011-11-28       Impact factor: 2.884

9.  Measurement of endotracheal tube secretions volume by micro computed tomography (MicroCT) scan: an experimental and clinical study.

Authors:  Andrea Coppadoro; Giacomo Bellani; Alfio Bronco; Roberto Borsa; Alberto Lucchini; Simone Bramati; Leonello Avalli; Roberto Marcolin; Antonio Pesenti
Journal:  BMC Anesthesiol       Date:  2014-03-28       Impact factor: 2.217

10.  Ex Vivo Evaluation of Secretion-Clearing Device in Reducing Airway Resistance within Endotracheal Tubes.

Authors:  Christopher Waters; R Constance Wiener; Hamed M Motlagh
Journal:  Crit Care Res Pract       Date:  2018-12-10
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.