Pulse oximetry during low perfusion caused by emerging pneumonia and sepsis in rabbits.

Helmut D Hummler1, Frank Pohlandt, Axel R Franz.   

Abstract

OBJECTIVE: This study tested the effects of low perfusion caused by emerging sepsis on the reliability of a new pulse oximetry technology (Masimo SET; IVY 405T) compared with a standard pulse oximeter (Nellcor N-200).
DESIGN: Randomized trial.
SETTING: University animal research facility.
SUBJECTS: Twenty-six anesthetized, ventilated (Fio, 1.0), adult rabbits.
INTERVENTIONS: Pneumonia/sepsis was induced by tracheal instillation of Escherichia coli. Oxygen saturation was measured by pulse oximetry (Spo ) and recorded continuously until death. Arterial oxygen saturation (Sao2) was measured hourly by oximetry and whenever Spo dropped to </=95%, or whenever a difference of >/=5% between devices occurred. Spo2 sensors were positioned at both forelegs and switched hourly.
MEASUREMENTS AND MAIN RESULTS: The total time of signal loss was longer with the N-200 vs. the IVY: 65 (4-299) mins vs. 7 (0-97) mins [median (range)], p < 0.001. Signal loss was more prevalent during the first 80% of the experimental time with the N-200 compared with the IVY. Nineteen of 26 animals had a total of 62 episodes of a falsely low Spo2 value with either one of the two devices associated with hemodynamic deterioration. Median bias (Spo2 - Sao2) was small, but variability of bias values increased toward the end of the experimental time with both devices.
CONCLUSIONS: The pulse oximeter equipped with Masimo SET was less prone to signal loss than the standard pulse oximeter in this sepsis model. Episodes of falsely low Spo2 readings may occur, and deviation of Spo2 from Sao2 may be increased with deteriorating hemodynamics with both devices.

Entities:  

Mesh:

Year:  2002        PMID: 12441761     DOI: 10.1097/00003246-200211000-00016

Source DB:  PubMed          Journal:  Crit Care Med        ISSN: 0090-3493            Impact factor:   7.598


  6 in total

1.  Decreased accuracy of pulse oximetry measurements during low perfusion caused by sepsis: Is the perfusion index of any value?

Authors:  Helmut D Hummler; Anja Engelmann; Frank Pohlandt; Josef Högel; Axel R Franz
Journal:  Intensive Care Med       Date:  2006-06-30       Impact factor: 17.440

2.  Drug treatment combined with BCG vaccination reduces disease reactivation in guinea pigs infected with Mycobacterium tuberculosis.

Authors:  Shaobin Shang; Crystal A Shanley; Megan L Caraway; Eileen A Orme; Marcela Henao-Tamayo; Laurel Hascall-Dove; David Ackart; Ian M Orme; Diane J Ordway; Randall J Basaraba
Journal:  Vaccine       Date:  2012-01-11       Impact factor: 3.641

3.  The accuracy of pulse oximetry in emergency department patients with severe sepsis and septic shock: a retrospective cohort study.

Authors:  Ben J Wilson; Hamish J Cowan; Jason A Lord; Dan J Zuege; David A Zygun
Journal:  BMC Emerg Med       Date:  2010-05-05

4.  Accuracy of pulse oximetry readings in an animal model of low perfusion caused by emerging pneumonia and sepsis.

Authors:  Helmut D Hummler; Anja Engelmann; Frank Pohlandt; Josef Högel; Axel R Franz
Journal:  Intensive Care Med       Date:  2004-01-13       Impact factor: 17.440

5.  Forehead reflectance oximetry: a clinical comparison with conventional digit sensors during laparotomic and laparoscopic abdominal surgery.

Authors:  Andrea Casati; Grazia Squicciarini; Marco Baciarello; Marta Putzu; Alessandra Salvadori; Guido Fanelli
Journal:  J Clin Monit Comput       Date:  2007-08-16       Impact factor: 2.502

6.  The increase of vasomotor tone avoids the ability of the dynamic preload indicators to estimate fluid responsiveness.

Authors:  Juan P Bouchacourt; Juan A Riva; Juan C Grignola
Journal:  BMC Anesthesiol       Date:  2013-11-11       Impact factor: 2.217

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.