An investigation of a new amorphous silicon electronic portal imaging device for transit dosimetry.

Ellen E Grein1, Richard Lee, Kurt Luchka.   

Abstract

The relationship between the pixel value and exit dose was investigated for a new commercially available amorphous silicon electronic portal imaging device. The pixel to dose mapping function was established to be linear for detector distances between 116.5 cm to 150 cm from the source, radiation field sizes from 5 x 5 cm2 to 20 x 20 cm2 and beam energies of 6 to 18 MV. Coefficients in the mapping function were found to be dependent on beam energy and field size. Open and wedged field profiles measured with the device showed agreement to a maximum of 5% and 8%, respectively, as compared to film. A comparison of relative transmission measurements between the EPID and ion chamber indicate a maximum deviation of 6% and 2% at 6 and 18 MV, respectively, for an attenuator thickness of 21 cm and SDD > or = 130 cm. It was found that accuracies of better than 1% could be obtained if detector position and field size specific fitting parameters were used to generate unique mapping functions for each configuration.

Mesh:

Substances:

Year:  2002        PMID: 12408300     DOI: 10.1118/1.1508108

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  11 in total

1.  Practical guidelines for routine intensity-modulated radiotherapy verification: pre-treatment verification with portal dosimetry and treatment verification with in vivo dosimetry.

Authors:  A J Vinall; A J Williams; V E Currie; A Van Esch; D Huyskens
Journal:  Br J Radiol       Date:  2010-11       Impact factor: 3.039

2.  Monte Carlo-based adaptive EPID dose kernel accounting for different field size responses of imagers.

Authors:  Song Wang; Joseph K Gardner; John J Gordon; Weidong Li; Luke Clews; Peter B Greer; Jeffrey V Siebers
Journal:  Med Phys       Date:  2009-08       Impact factor: 4.071

3.  In patient dose reconstruction using a cine acquisition for dynamic arc radiation therapy.

Authors:  Angelo Piermattei; Andrea Fidanzio; Luigi Azario; Francesca Greco; Alessandra Mameli; Savino Cilla; Luca Grimaldi; Guido D'Onofrio; Boris Giuseppe Augelli; Gerardina Stimato; Diego Gaudino; Sara Ramella; Rolando D'Angelillo; Francesco Cellini; Lucio Trodella
Journal:  Med Biol Eng Comput       Date:  2009-02-17       Impact factor: 2.602

4.  Transit dosimetry in dynamic IMRT with an a-Si EPID.

Authors:  Mahsheed Sabet; Pejman Rowshanfarzad; Fred W Menk; Peter B Greer
Journal:  Med Biol Eng Comput       Date:  2014-05-31       Impact factor: 2.602

5.  Testing the portal imager GLAaS algorithm for machine quality assurance.

Authors:  G Nicolini; E Vanetti; A Clivio; A Fogliata; G Boka; L Cozzi
Journal:  Radiat Oncol       Date:  2008-05-21       Impact factor: 3.481

6.  A method for deconvolution of integrated electronic portal images to obtain incident fluence for dose reconstruction.

Authors:  Wendel Dean Renner; Kevin Norton; Timothy Holmes
Journal:  J Appl Clin Med Phys       Date:  2005-11-21       Impact factor: 2.102

7.  Feasibility of using two-dimensional array dosimeter for in vivo dose reconstruction via transit dosimetry.

Authors:  Heeteak Chung; Jonathan Li; Sanjiv Samant
Journal:  J Appl Clin Med Phys       Date:  2011-04-08       Impact factor: 2.102

8.  Portal dosimetry for pretreatment verification of IMRT plan: a comparison with 2D ion chamber array.

Authors:  Dayananda Shamurailatpam Sharma; Vaibav Mhatre; Malhotra Heigrujam; Kaustav Talapatra; Suman Mallik
Journal:  J Appl Clin Med Phys       Date:  2010-08-19       Impact factor: 2.102

9.  The GLAaS algorithm for portal dosimetry and quality assurance of RapidArc, an intensity modulated rotational therapy.

Authors:  Giorgia Nicolini; Eugenio Vanetti; Alessandro Clivio; Antonella Fogliata; Stine Korreman; Jiri Bocanek; Luca Cozzi
Journal:  Radiat Oncol       Date:  2008-09-09       Impact factor: 3.481

10.  Primo software as a tool for Monte Carlo simulations of intensity modulated radiotherapy: a feasibility study.

Authors:  Alessandro Esposito; Sofia Silva; Jorge Oliveira; Joana Lencart; João Santos
Journal:  Radiat Oncol       Date:  2018-05-15       Impact factor: 3.481

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.