| Literature DB >> 12199574 |
Ulli Backofen1, Frank-Michael Matysik, Craig E Lunte.
Abstract
A simple capillary electrophoresis system in the planar format that uses a new, hydrodynamic injection principle is described. The system was realized with poly(dimethylsiloxane)-glass chips and microdisk electrodes for amperometric detection. Using a double-tee injector, no precise voltage control of the electrolyte reservoirs was needed, thus making the microchip CE system more user-friendly. The analytical characteristics of chip-based CE-EC were evaluated using ascorbic acid as the model analyte. The reproducibility of migration time and signal height was expressed by relative standard deviations of 1.2% and 5.1%, respectively (n = 5). The limit of detection for ascorbic acid was approximately 5 microM at a signal-to-noise ratio of 3. Practical application concerning the determination of physiologically relevant compounds such as noradrenaline and L-dopa are discussed.Entities:
Mesh:
Substances:
Year: 2002 PMID: 12199574 PMCID: PMC2519826 DOI: 10.1021/ac020110j
Source DB: PubMed Journal: Anal Chem ISSN: 0003-2700 Impact factor: 6.986