Abstract
An on-chip electrochemical detector for micromachined capillary electrophoresis (CE) systems, based on sputtering a gold working electrode directly onto the capillary outlet, is described. The new on-chip detector preparation requires no microfabrication or alignment procedures nor a decoupling mechanism. The attractive performance of the integrated electrophoresis chips/amperometric detection was demonstrated for the anodic detection of neurotransmitters. The response for dopamine was linear from 20 to 200 μM, with a LOD of 1.0 μM and a sensitivity of 52 pA/μM. Such intimate coupling of capillary electrophoresis chips and electrochemical detection facilitates the realization of complete integrated microanalytical devices.Entities:
Year: 1999 PMID: 21662891 DOI: 10.1021/ac9904720
Source DB: PubMed Journal: Anal Chem ISSN: 0003-2700 Impact factor: 6.986