| Literature DB >> 11288891 |
A J Gawron1, R S Martin, S M Lunte.
Abstract
The first carbon-based dual-electrode detector for microchip capillary electrophoresis (CE) is described. The poly(dimethylsiloxane) (PDMS)-based microchip CE devices were constructed by reversibly sealing a PDMS layer containing separation and injection channels to another PDMS layer containing carbon fiber working electrodes. End-channel amperometric detection was employed and the performance of the chip was evaluated using catechol. The response was found to be linear between 1 and 600 microM with an experimentally determined limit of detection (LOD) of 500 nM and a sensitivity of 30 pA/microM. Collection efficiencies for catechol ranged from 36.0 to 43.7% at field strengths of 260-615 V/cm. The selectivity that can be gained with these devices is demonstrated by the first CE-based dual-electrode detection of a Cu(II) peptide complex. These devices illustrate the potential for a rugged and easily constructed microchip CE system with an integrated carbon-based detector of similar scale.Entities:
Mesh:
Substances:
Year: 2001 PMID: 11288891 DOI: 10.1002/1522-2683(200101)22:2<242::AID-ELPS242>3.0.CO;2-W
Source DB: PubMed Journal: Electrophoresis ISSN: 0173-0835 Impact factor: 3.535