Accuracy of caries detection with four storage phosphor systems and E-speed radiographs.

H Hintze1, A Wenzel, M Frydenberg.   

Abstract

OBJECTIVE: To evaluate the accuracy of caries detection using four intra-oral storage phosphor plate systems exposed with two different exposure times and one film system.
METHODS: Under in vitro and standardised conditions 365 approximal and 159 occlusal surfaces were radiographed using four storage phosphor plate systems: DenOptix, Cd-dent, Digora (blue and white plates) and one film system: Ektaspeed Plus. The plates were exposed twice with 10% and 25% of the time needed for the film. Four observers recorded approximal enamel/dentine and occlusal dentine caries using a five-point confidence scale. Caries presence was validated histologically and diagnostic accuracy of the systems by using ROC curve areas (Az).
RESULTS: The shorter exposure time for approximal caries, Digora(white)(Az=0.68) and Ekta Plus (Az=0.69) were significantly more accurate than the other systems (Az=0.64-0.65). For occlusal caries using the same exposure time, no significant difference was found between the digital systems (Az=0.76-0.77), which all showed to be significantly less accurate than Ekta Plus (Az=0.81). Using the longer exposure time for approximal caries, no significant difference was found between DenOptix (Az=0.68), Digora(blue) (Az=0.69), Digora(white) (Az=0.69) and Ekta Plus (Az=0.69) which were all significantly more accurate than Cd-dent (Az=0.65). For occlusal caries using the same exposure time, Digora(blue) (Az=0.81) was the most significantly accurate of the digital systems. When approximal caries was the diagnostic task, the longer exposure time resulted in significantly higher accuracy with the DenOptix and Digora(blue) systems but not with the Cd-dent and Digora(white) systems. For occlusal caries, the longer exposure time resulted in a significant improvement only with the Digora(blue) system.
CONCLUSION: For approximal caries using the longer exposure time, no significant difference in diagnostic accuracy was found between DenOptix, Digora(blue), Digora(white) and Ekta Plus which were all significantly more accurate than the Cd-dent. For occlusal caries using the longer exposure time, Digora(blue) was found to be the most accurate of the digital systems. The exposure time had an influence on the diagnostic accuracy with the DenOptix and Digora(blue) systems for approximal and with the Digora(blue) system for occlusal caries detection.

Entities:  

Mesh:

Year:  2002        PMID: 12058264     DOI: 10.1038/sj/dmfr/4600686

Source DB:  PubMed          Journal:  Dentomaxillofac Radiol        ISSN: 0250-832X            Impact factor:   2.419


  19 in total

1.  Diagnostic accuracy of proximal enamel subsurface demineralization and its relationship with calcium loss and lesion depth.

Authors:  E Onem; B G Baksi; B H Sen; O Sögüt; A Mert
Journal:  Dentomaxillofac Radiol       Date:  2011-11-10       Impact factor: 2.419

2.  Comparison of digital systems and conventional dental film for the detection of approximal enamel caries.

Authors:  A A Pontual; D P de Melo; S M de Almeida; F N Bóscolo; F Haiter Neto
Journal:  Dentomaxillofac Radiol       Date:  2010-10       Impact factor: 2.419

3.  Effect of tube potential and image receptor on the detection of natural proximal caries in primary teeth.

Authors:  Elif Sogur; B Güniz Baksı; Kaan Orhan; S Candan Paksoy; Salih Dogan; Yılmaz S Erdal; Ali Mert
Journal:  Clin Oral Investig       Date:  2010-09-14       Impact factor: 3.573

Review 4.  Modern dental imaging: a review of the current technology and clinical applications in dental practice.

Authors:  Bart Vandenberghe; Reinhilde Jacobs; Hilde Bosmans
Journal:  Eur Radiol       Date:  2010-06-11       Impact factor: 5.315

5.  Diagnostic accuracy of different imaging modalities in detection of proximal caries.

Authors:  B Senel; K Kamburoglu; O Uçok; S P Yüksel; T Ozen; H Avsever
Journal:  Dentomaxillofac Radiol       Date:  2010-12       Impact factor: 2.419

6.  The effect of delayed scanning of storage phosphor plates on occlusal caries detection.

Authors:  E Sogur; B G Baksi; A Mert
Journal:  Dentomaxillofac Radiol       Date:  2012-01-26       Impact factor: 2.419

7.  Enterprise-wide implementation of digital radiography in oral and maxillofacial imaging: the University of Florida Dentistry System.

Authors:  Madhu K Nair; James C Pettigrew; Jeffrey S Loomis; Robert E Bates; Stephen Kostewicz; Boyd Robinson; Jean Sweitzer; Teresa A Dolan
Journal:  J Digit Imaging       Date:  2008-09-03       Impact factor: 4.056

8.  The usage of digital radiography and cone beam computed tomography among Turkish dentists.

Authors:  Semanur Dölekoğlu; E Fişekçioğlu; M İlgüy; D İlgüy
Journal:  Dentomaxillofac Radiol       Date:  2011-09       Impact factor: 2.419

9.  A comparison of the diagnostic accuracy of in vivo and in vitro photostimulable phosphor digital images in the detection of occlusal caries lesions.

Authors:  K Kamburoglu; B Senel; S P Yüksel; T Ozen
Journal:  Dentomaxillofac Radiol       Date:  2010-01       Impact factor: 2.419

10.  The efficiency of operating microscope compared with unaided visual examination, conventional and digital intraoral radiography for proximal caries detection.

Authors:  Ilkay Peker; Meryem Toraman Alkurt; Oya Bala; Bulent Altunkaynak
Journal:  Int J Dent       Date:  2009-01-29
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.