Literature DB >> 12004013

Survival and polyethylene wear of porous-coated acetabular components in patients less than fifty years old: results at nine to fourteen years.

James D Crowther1, Paul F Lachiewicz.   

Abstract

BACKGROUND: Younger patients (those who are less than fifty years old) have been shown to have a high rate of failure of cemented acetabular components following total hip arthroplasty. In this report, we present the results associated with the use of an uncemented acetabular component in young patients who were evaluated at a minimum of nine years postoperatively.
METHODS: Between December 1984 and December 1989, the senior author performed 174 primary total hip arthroplasties with use of a single design of porous-coated acetabular component. Seventy-one of these procedures were performed in fifty-six patients who were younger than fifty years old. Fifty-six of the seventy-one hips were available for radiographic and clinical analysis after a mean duration of follow-up of eleven years. All hips had been treated with a Harris-Galante-I porous-coated acetabular component that had been placed with a line-to-line fit and fixed with a mean of four screws. Clinical analysis was performed with use of the Harris hip score. Standardized anteroposterior radiographs were analyzed with regard to migration, radiolucent lines, pelvic osteolysis, and two-dimensional linear wear of the polyethylene.
RESULTS: No metal shell was revised because of aseptic loosening, and no shell was loose at the time of the latest follow-up. A nonprogressive radiolucent line was seen in one zone in ten hips (18%) and in two zones in six hips (11%). No hip had a radiolucent line in all three zones. Pelvic osteolysis was noted in thirteen hips (23%); the osteolysis was observed in the ischium in eleven hips and around the screws in two. Survivorship analysis revealed that the probability of survival of the metal shell was 98% (95% confidence interval, 96.9% to 99.9%) at ten years. The mean rate of linear polyethylene wear (and standard deviation) was 0.15 +/- 0.10 mm/yr (range, 0.02 to 0.59 mm/yr). The wear rate was significantly increased in patients with an excellent Harris hip score (p = 0.004) and a younger age (less than thirty-eight years) (p = 0.026). With the numbers available, no relationship could be detected between the wear rate and the gender or weight of the patient, the polyethylene thickness, the abduction angle, or the femoral neck length.
CONCLUSIONS: The fixation and survival of porous-coated acetabular metal shells in patients less than fifty years old was excellent after a mean duration of follow-up of eleven years. The high rate of linear polyethylene wear and the high prevalence of pelvic osteolysis are of serious concern in this patient population. Continued follow-up will be necessary to evaluate the influence of these findings on the longevity of the fixation of this prosthesis.

Entities:  

Mesh:

Substances:

Year:  2002        PMID: 12004013     DOI: 10.2106/00004623-200205000-00005

Source DB:  PubMed          Journal:  J Bone Joint Surg Am        ISSN: 0021-9355            Impact factor:   5.284


  24 in total

1.  Is routine mid-term total hip arthroplasty surveillance beneficial?

Authors:  James A Keeney; Bradley S Ellison; William J Maloney; John C Clohisy
Journal:  Clin Orthop Relat Res       Date:  2012-06-06       Impact factor: 4.176

Review 2.  Hip resurfacing: a technology reborn.

Authors:  Steven Cutts; Paul B Carter
Journal:  Postgrad Med J       Date:  2006-12       Impact factor: 2.401

Review 3.  Sports activity after total hip and knee arthroplasty : specific recommendations concerning tennis.

Authors:  Thorsten M Seyler; Michael A Mont; Phillip S Ragland; Munaf M Kachwala; Ronald E Delanois
Journal:  Sports Med       Date:  2006       Impact factor: 11.136

4.  Acetabular revision surgery with impacted bone allografts and cemented cups in patients younger than 55 years.

Authors:  Fernando Comba; Martín Buttaro; Rodolfo Pusso; Francisco Piccaluga
Journal:  Int Orthop       Date:  2008-02-09       Impact factor: 3.075

5.  Long-term results of the standard Wagner cup.

Authors:  Uwe Demmelmeyer; Annemarie Schraml; Wolfgang Hönle; Alexander Schuh
Journal:  Int Orthop       Date:  2009-04-08       Impact factor: 3.075

6.  Subcapital correction osteotomy for malunited slipped capital femoral epiphysis.

Authors:  Lucas A Anderson; Jeremy M Gililland; Christoper E Pelt; Christopher L Peters
Journal:  J Pediatr Orthop       Date:  2013-06       Impact factor: 2.324

7.  Cementless Total Hip Arthroplasty With Metasul Bearings Provides Good Results in Active Young Patients: A Concise Followup.

Authors:  Christian P Delaunay; Sophie Putman; Benjamin Puliéro; Matthieu Bégin; Henri Migaud; François Bonnomet
Journal:  Clin Orthop Relat Res       Date:  2016-10       Impact factor: 4.176

8.  Chronic renal failure due to amyloid nephropathy caused by chronic infection after total hip replacement.

Authors:  Shinsuke Nishimura; Tomoji Matsumae; Yuji Murakami; Yasuhiro Abe; Yoshie Sasatomi; Ikufumi Nagayoshi; Kazuo Ueda; Hitoshi Nakashima
Journal:  CEN Case Rep       Date:  2014-04-16

9.  Metal on metal hip resurfacing versus uncemented custom total hip replacement--early results.

Authors:  Nemandra A Sandiford; Sarah K Muirhead-Allwood; John A Skinner; Jia Hua
Journal:  J Orthop Surg Res       Date:  2010-02-18       Impact factor: 2.359

10.  Cementless total hip replacement: past, present, and future.

Authors:  Harumoto Yamada; Yasuo Yoshihara; Osamu Henmi; Mitsuhiro Morita; Yuichiro Shiromoto; Tomoki Kawano; Arihiko Kanaji; Kennichi Ando; Masato Nakagawa; Naoto Kosaki; Eiichi Fukaya
Journal:  J Orthop Sci       Date:  2009-04-01       Impact factor: 1.601

View more