Attenuation correction and gating synergistically improve the diagnostic accuracy of myocardial perfusion SPECT.

Jonathan M Links1, E Gordon DePuey, Raymond Taillefer, Lewis C Becker.   

Abstract

BACKGROUND: The diagnostic accuracy of myocardial perfusion single photon emission computed tomography is limited by soft tissue attenuation. Artifacts may be reduced by attenuation correction (AC) or compensated for by assessment of wall motion in gated images. We studied the benefit of gating and AC, both separately and combined, in improving accuracy. METHODS AND
RESULTS: Sixty-six subjects (27 with > or =50% angiographic stenosis, 5 with <50% stenosis, and 34 Bayesian normal volunteers) underwent gated AC single photon emission computed tomography. Images were reconstructed and independently viewed in 4 ways: static with motion correction (MC) only, gated with MC only, static with MC plus blur correction plus AC (referred to as combined corrections [CC]), and gated with CC. Images were interpreted by 2 blinded observers for overall presence of coronary disease and for vascular territory (left anterior descending [LAD], left circumflex [LCx], and right coronary artery [RCA]). Statistical analysis of sensitivity and normalcy was done by means of the Cochran Q test. Overall diagnostic accuracy showed statistically significant improvement (P =.05 for sensitivity, P <.001 for normalcy), progressing from static MC to gated MC to static CC to gated CC (sensitivity/normalcy = 85%/54%, 78%/62%, 93%/77%, and 96%/85%, respectively, for the 4 reconstruction and viewing approaches). Sensitivity was highest in all 3 vascular territories for the combination of gating and CC; normalcy was also highest with the same combination for the LAD and RCA territories; sensitivity/normalcy with this combination was 85%/87% for the LAD, 69%/87% for the LCx, and 89%/87% for the RCA territory.
CONCLUSIONS: The combination of gating and CC provides the highest diagnostic accuracy, and gating and AC should thus be considered complementary and synergistic.

Entities:  

Mesh:

Year:  2002        PMID: 11986563     DOI: 10.1067/mnc.2002.120163

Source DB:  PubMed          Journal:  J Nucl Cardiol        ISSN: 1071-3581            Impact factor:   5.952


  15 in total

1.  Optimal specificity of thallium-201 SPECT through recognition of imaging artifacts.

Authors:  E G DePuey; E V Garcia
Journal:  J Nucl Med       Date:  1989-04       Impact factor: 10.057

2.  Limited incremental diagnostic values of attenuation-noncorrected gating and ungated attenuation correction to rest/stress myocardial perfusion SPECT in patients with an intermediate likelihood of coronary artery disease.

Authors:  D S Lee; Y So; G J Cheon; K M Kim; M M Lee; J K Chung; M C Lee
Journal:  J Nucl Med       Date:  2000-05       Impact factor: 10.057

3.  Reverend Bayes' silent majority. An alternative factor affecting sensitivity and specificity of exercise electrocardiography.

Authors:  G A Diamond
Journal:  Am J Cardiol       Date:  1986-05-01       Impact factor: 2.778

4.  Analysis of probability as an aid in the clinical diagnosis of coronary-artery disease.

Authors:  G A Diamond; J S Forrester
Journal:  N Engl J Med       Date:  1979-06-14       Impact factor: 91.245

5.  How to detect and avoid myocardial perfusion SPECT artifacts.

Authors:  E G DePuey
Journal:  J Nucl Med       Date:  1994-04       Impact factor: 10.057

6.  Comparative diagnostic accuracy of Tl-201 and Tc-99m sestamibi SPECT imaging (perfusion and ECG-gated SPECT) in detecting coronary artery disease in women.

Authors:  R Taillefer; E G DePuey; J E Udelson; G A Beller; Y Latour; F Reeves
Journal:  J Am Coll Cardiol       Date:  1997-01       Impact factor: 24.094

7.  Combined corrections for attenuation, depth-dependent blur, and motion in cardiac SPECT: a multicenter trial.

Authors:  J M Links; L C Becker; P Rigo; R Taillefer; L Hanelin; F Anstett; D Burckhardt; L Mixon
Journal:  J Nucl Cardiol       Date:  2000 Sep-Oct       Impact factor: 5.952

8.  Quantitative evaluation of a comprehensive motion, resolution, and attenuation correction program: initial experience.

Authors:  P Rigo; P Van Boxem; J Foulon; M Safi; J Engdahl; J Links
Journal:  J Nucl Cardiol       Date:  1998 Sep-Oct       Impact factor: 5.952

9.  Using gated technetium-99m-sestamibi SPECT to characterize fixed myocardial defects as infarct or artifact.

Authors:  E G DePuey; A Rozanski
Journal:  J Nucl Med       Date:  1995-06       Impact factor: 10.057

10.  American Society of Nuclear Cardiology position statement on electrocardiographic gating of myocardial perfusion SPECT scintigrams.

Authors:  T M Bateman; D S Berman; G V Heller; K A Brown; M D Cerqueira; M S Verani; J E Udelson
Journal:  J Nucl Cardiol       Date:  1999 Jul-Aug       Impact factor: 3.872

View more
  49 in total

1.  Evaluation of right ventricular function by NuSMUGA software: gated blood-pool SPECT vs. first-pass radionuclide angiography.

Authors:  R H J A Slart; L Poot; D A Piers; D J van Veldhuisen; P L Jager
Journal:  Int J Cardiovasc Imaging       Date:  2003-10       Impact factor: 2.357

2.  Effect of attenuation correction on the interpretation of 99mTc-sestamibi myocardial perfusion scintigraphy: the impact of 1 year's experience.

Authors:  Riemer H J A Slart; Tjin H Que; Dirk J van Veldhuisen; Lieke Poot; Paul K Blanksma; D Albert Piers; Pieter L Jager
Journal:  Eur J Nucl Med Mol Imaging       Date:  2003-08-27       Impact factor: 9.236

3.  Clinical significance of apical thinning after attenuation correction.

Authors:  Jonathan M Links; Lewis C Becker; Frank Anstett
Journal:  J Nucl Cardiol       Date:  2004 Jan-Feb       Impact factor: 5.952

4.  Come on, baby, let's do the twist: detecting and correcting cardiac torsion effects in myocardial perfusion SPECT.

Authors:  Jonathan M Links; Lewis C Becker
Journal:  J Nucl Cardiol       Date:  2002 Sep-Oct       Impact factor: 5.952

5.  American Society of Nuclear Cardiology and Society of Nuclear Medicine joint position statement: attenuation correction of myocardial perfusion SPECT scintigraphy.

Authors:  Gary V Heller; Jonathan Links; Timothy M Bateman; Jack A Ziffer; Edward Ficaro; Mylan C Cohen; Robert C Hendel
Journal:  J Nucl Cardiol       Date:  2004 Mar-Apr       Impact factor: 5.952

6.  Yield of a novel ultra-low-dose computed tomography device mounted on a dedicated cardiac SPECT system in improving the accuracy of myocardial perfusion imaging and the detection of chest abnormalities.

Authors:  Nili Zafrir; Gigeon Shafir; Gil Kovalski; Israel Mats; Jean-Paul Bouhnik; Alexander Battler; Alejandro Solodky
Journal:  J Nucl Cardiol       Date:  2011-12-28       Impact factor: 5.952

7.  Quantitative Nuclear Cardiology: we are almost there!

Authors:  Ernest V Garcia
Journal:  J Nucl Cardiol       Date:  2012-03-31       Impact factor: 5.952

8.  Simplified normal limits and automated quantitative assessment for attenuation-corrected myocardial perfusion SPECT.

Authors:  Piotr J Slomka; Mathews B Fish; Santiago Lorenzo; Hidetaka Nishina; James Gerlach; Daniel S Berman; Guido Germano
Journal:  J Nucl Cardiol       Date:  2006-09       Impact factor: 5.952

9.  Diagnostic accuracy of myocardial perfusion imaging in a study population without post-test referral bias.

Authors:  Allan Johansen; Poul Flemming Høilund-Carlsen; Henrik Wulff Christensen; Werner Vach; Henrik Boel Jørgensen; Annegrete Veje; Torben Haghfelt
Journal:  J Nucl Cardiol       Date:  2005 Sep-Oct       Impact factor: 5.952

10.  Diagnostic accuracy of gated Tc-99m sestamibi stress myocardial perfusion SPECT with combined supine and prone acquisitions to detect coronary artery disease in obese and nonobese patients.

Authors:  Daniel S Berman; Xingping Kang; Hidetaka Nishina; Piotr J Slomka; Leslee J Shaw; Sean W Hayes; Ishac Cohen; John D Friedman; James Gerlach; Guido Germano
Journal:  J Nucl Cardiol       Date:  2006 Mar-Apr       Impact factor: 5.952

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.