Respiration-induced B0 fluctuations and their spatial distribution in the human brain at 7 Tesla.

Pierre-François Van de Moortele1, Josef Pfeuffer, Gary H Glover, Kamil Ugurbil, Xiaoping Hu.   

Abstract

In functional magnetic resonance imaging (fMRI), it is known that physiological influences such as cardiac pulsation, respiration, and brain motion can induce fluctuations in signal intensity and phase. Some of the mechanisms potentially involved in those phenomena are expected to be amplified at higher magnetic fields. This study addresses the issue of B(0) fluctuations induced by susceptibility changes during respiration attributed to movements of chest and diaphragm, and variations in the oxygen concentration. It is demonstrated that respiration-induced resonance offsets (RIROs) are significant at 7T. Data were acquired with an RF pulse (no gradients), multislice echo-planar imaging (EPI), and dynamic 3D fast low-angle shot (3D- FLASH) imaging. Three main observations summarize the experimental findings. First, in FIDs measured after a single RF pulse, a RIRO with a large amplitude was consistently detected, although the average amplitude varied between subjects from 1.45 Hz to 4 Hz. Second, in transverse EPI images obtained in the occipital lobe, the RIRO amplitude showed a monotonic increase along the z axis toward the lungs. Third, a more detailed spatial analysis with 3D-FLASH phase maps revealed that a previously published analytical model can accurately describe the spatial distribution of RIRO. Consequential apparent motions in the EPI series, as well as the implications of slice orientation for correction strategies are discussed. Copyright 2002 Wiley-Liss, Inc.

Entities:  

Mesh:

Year:  2002        PMID: 11979567     DOI: 10.1002/mrm.10145

Source DB:  PubMed          Journal:  Magn Reson Med        ISSN: 0740-3194            Impact factor:   4.668


  68 in total

Review 1.  Magnetic resonance imaging at ultrahigh fields.

Authors:  Kamil Ugurbil
Journal:  IEEE Trans Biomed Eng       Date:  2014-03-25       Impact factor: 4.538

2.  Respiration artifact correction in three-dimensional proton resonance frequency MR thermometry using phase navigators.

Authors:  Bryant T Svedin; Allison Payne; Dennis L Parker
Journal:  Magn Reson Med       Date:  2015-08-13       Impact factor: 4.668

3.  Removal of high frequency contamination from motion estimates in single-band fMRI saves data without biasing functional connectivity.

Authors:  Caterina Gratton; Ally Dworetsky; Rebecca S Coalson; Babatunde Adeyemo; Timothy O Laumann; Gagan S Wig; Tania S Kong; Gabriele Gratton; Monica Fabiani; Deanna M Barch; Daniel Tranel; Oscar Miranda-Dominguez; Damien A Fair; Nico U F Dosenbach; Abraham Z Snyder; Joel S Perlmutter; Steven E Petersen; Meghan C Campbell
Journal:  Neuroimage       Date:  2020-04-20       Impact factor: 6.556

4.  Methodology of H NMR Spectroscopy of the Human Brain at Very High Magnetic Fields.

Authors:  I Tkáč; R Gruetter
Journal:  Appl Magn Reson       Date:  2005-03       Impact factor: 0.831

5.  Magnetic field shift due to mechanical vibration in functional magnetic resonance imaging.

Authors:  Bernd U Foerster; Dardo Tomasi; Elisabeth C Caparelli
Journal:  Magn Reson Med       Date:  2005-11       Impact factor: 4.668

Review 6.  Special considerations for functional magnetic resonance imaging of pediatric populations.

Authors:  Eleni Kotsoni; Dana Byrd; B J Casey
Journal:  J Magn Reson Imaging       Date:  2006-06       Impact factor: 4.813

7.  Hybrid two-dimensional navigator correction: a new technique to suppress respiratory-induced physiological noise in multi-shot echo-planar functional MRI.

Authors:  Robert L Barry; L Martyn Klassen; Joy M Williams; Ravi S Menon
Journal:  Neuroimage       Date:  2007-10-11       Impact factor: 6.556

8.  Correction of breathing-induced errors in magnetic resonance thermometry of hyperthermia using multiecho field fitting techniques.

Authors:  Cory R Wyatt; Brian J Soher; James R MacFall
Journal:  Med Phys       Date:  2010-12       Impact factor: 4.071

9.  Improving robustness and reliability of phase-sensitive fMRI analysis using temporal off-resonance alignment of single-echo timeseries (TOAST).

Authors:  Andrew D Hahn; Andrew S Nencka; Daniel B Rowe
Journal:  Neuroimage       Date:  2008-10-18       Impact factor: 6.556

10.  Distortion and signal loss in medial temporal lobe.

Authors:  Cheryl A Olman; Lila Davachi; Souheil Inati
Journal:  PLoS One       Date:  2009-12-03       Impact factor: 3.240

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.