The x-ray sensitivity of amorphous selenium for mammography.

Mary F Stone1, Wei Zhao, Barbara V Jacak, Paul O'Connor, Bo Yu, Pavel Rehak.   

Abstract

A study of the x-ray sensitivity of amorphous selenium (a-Se) for digital mammography has been performed. A uniform layer of a-Se was deposited on a glass substrate with electrodes on both surfaces. The deposition procedure was identical to that used for a-Se flat-panel detectors. A high voltage was applied to the top surface of the a-Se layer in order to establish an electric field E(Se). Then the sample was exposed to x rays with 27 kVp spectra generated from an x-ray tube with a molybdenum (Mo) target. The mean x-ray energy of the spectrum used was approximately 16.6 keV. The x-ray current generated by the a-Se layer was measured as a function of E(Se). From the current measurement and the estimation of total x-ray energy absorbed in the a-Se, the energy required to create one electron-hole pair (EHP), W, was determined as a function of E(Se). It was found that at the most commonly used E(Se) of 10 V/microm, W was measured as 64 eV. This is considerably higher than the widely accepted typical value of W = 50 eV measured at higher x-ray photon energies (e.g., 50 keV). The dependence of W as a function of E(Se) can be best fitted using the empirical expression of E(Se)-gamma. This relationship is consistent with the results obtained at higher x-ray energies. This article provides an accurate measurement of x-ray sensitivity of a-Se at mammographic energies independent of detector operation, such as the most recently developed flat-panel detectors. The results will be a useful tool for investigation and optimization of a-Se-based x-ray imaging detectors, such as determination of pixel fill-factor and optimal E(Se) during operation.

Entities:  

Mesh:

Substances:

Year:  2002        PMID: 11929014     DOI: 10.1118/1.1449874

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  6 in total

Review 1.  Digital detectors for mammography: the technical challenges.

Authors:  A Noel; F Thibault
Journal:  Eur Radiol       Date:  2004-10-08       Impact factor: 5.315

2.  Flat detectors and their clinical applications.

Authors:  Martin Spahn
Journal:  Eur Radiol       Date:  2005-04-02       Impact factor: 5.315

3.  Evaluation of a hybrid direct-indirect active matrix flat-panel imager using Monte Carlo simulation.

Authors:  Scott Dow; Adrian Howansky; Anthony R Lubinsky; Wei Zhao
Journal:  J Med Imaging (Bellingham)       Date:  2020-05-12

4.  Porous Silicon-Based Quantum Dot Broad Spectrum Radiation Detector.

Authors:  M Urdaneta; P Stepanov; I N Weinberg; I R Pala; S Brock
Journal:  J Instrum       Date:  2011-01-11       Impact factor: 1.415

5.  Development of active matrix flat panel imagers incorporating thin layers of polycrystalline HgI(2) for mammographic x-ray imaging.

Authors:  Hao Jiang; Qihua Zhao; Larry E Antonuk; Youcef El-Mohri; Tapan Gupta
Journal:  Phys Med Biol       Date:  2013-01-14       Impact factor: 3.609

6.  Direct Thermal Growth of Large Scale Cl-doped CdTe Film for Low Voltage High Resolution X-ray Image Sensor.

Authors:  Silah Lee; Jin Sung Kim; Kyeong Rok Ko; Gun Hwan Lee; Dong Jin Lee; Dong Wook Kim; Jin Eui Kim; Ho Kyung Kim; Dong Woon Kim; Seongil Im
Journal:  Sci Rep       Date:  2018-10-04       Impact factor: 4.379

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.