B S Uh1, B D Beynnon, D L Churchill, L D Haugh, M A Risberg, B C Fleming, B S Un.
Abstract
The Vermont knee laxity device (VKLD) was developed to evaluate anterior-posterior (A-P) displacement of the tibia relative to the femur (A-P laxity) during weightbearing and non-weightbearing conditions. The purposes of this study were to determine the repeatability and reliability of the VKLD measurements of A-P laxity and to compare them with two devices currently in clinical use: the KT-1000 knee arthrometer and planar stress radiography. Two independent examiners tested six subjects with no history of knee injury. A-P laxity was measured on three separate days with the KT-1000 and the VKLD. With the VKLD, A-P laxity was measured in the weightbearing and non-weightbearing conditions. In addition, one examiner measured A P laxity in each subject on each day using a planar stress radiography technique. Similar A-P laxity values were obtained with the KT-1000 and the VKLD; however, the planar stress radiography technique measured less A-P laxity compared to the VKLD (9.2+/-2.2 mm versus 13.3+/-2.9 mm, P = 0.0004). None of the three devices showed significant differences in measuring A-P laxity between days. During weightbearing, A-P laxity was reduced by 65-70% compared to the non-weightbearing condition (P = 0.0001). Future investigations will use the VKLD to study subjects that have suffered injury to the anterior and posterior cruciate ligaments.Entities:
Mesh:
Year: 2001 PMID: 11781022 DOI: 10.1016/S0736-0266(01)00055-9