Clinical comparison of the Keeler Pulsair 3000 with Goldmann applanation tonometry.

V A Parker1, J Herrtage, N J Sarkies.   

Abstract

AIM: To confirm the accuracy of the Pulsair 3000 before introducing the instrument into clinical practice.
METHOD: A masked study by two experienced tonometrists comparing the mean of the Goldmann intraocular pressure (IOP) readings against the Pulsair 3000 reading (average of four puffs). Results of 150 eyes were compared with an IOP range of 10 mm Hg-44 mm Hg.
RESULTS: Correlation between the two Goldmann tonometry results was 0.9830 with a standard deviation of 1.1085 mm Hg. Correlation between the mean of the two Goldmann readings against the Pulsair 3000 reading was 0.982 with a standard deviation of 1.1179 mm Hg. Bland-Altman analysis confirms a satisfactory outcome.
CONCLUSION: The Pulsair 3000 provides an accurate and objective method of measuring IOP with many advantages over traditional Goldmann tonometry.

Mesh:

Year:  2001        PMID: 11673293      PMCID: PMC1723760          DOI: 10.1136/bjo.85.11.1303

Source DB:  PubMed          Journal:  Br J Ophthalmol        ISSN: 0007-1161            Impact factor:   4.638


  8 in total

1.  Clinical comparison of air-puff and Goldmann tonometers.

Authors:  H L Brencher; P Kohl; A R Reinke; R L Yolton
Journal:  J Am Optom Assoc       Date:  1991-05

2.  Standardizing the measurement of intraocular pressure for clinical research. Guidelines from the Eye Care Technology Forum.

Authors:  M A Kass
Journal:  Ophthalmology       Date:  1996-01       Impact factor: 12.079

3.  Comparison of a new non-contact tonometer with Goldmann applanation.

Authors:  M J Moseley; N M Evans; A R Fielder
Journal:  Eye (Lond)       Date:  1989       Impact factor: 3.775

4.  Effects of repeated tonometry: genuine and sham measurements.

Authors:  K Wilke
Journal:  Acta Ophthalmol (Copenh)       Date:  1972

5.  Is applanation tonometry a risk factor for transmission of Creutzfeldt-Jakob disease?

Authors:  M Rizzo; J J Corbett; H S Thompson
Journal:  Arch Ophthalmol       Date:  1987-03

6.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

7.  Reproducibility with the Keeler Pulsair 2000 non-contact tonometer.

Authors:  S A Vernon
Journal:  Br J Ophthalmol       Date:  1995-06       Impact factor: 4.638

8.  Clinical evaluation of five portable tonometers.

Authors:  T A Wingert; C J Bassi; W H McAlister; J C Galanis
Journal:  J Am Optom Assoc       Date:  1995-11
  8 in total
  10 in total

1.  A comparison of four methods of tonometry: method agreement and interobserver variability.

Authors:  P-A Tonnu; T Ho; K Sharma; E White; C Bunce; D Garway-Heath
Journal:  Br J Ophthalmol       Date:  2005-07       Impact factor: 4.638

2.  Comparison of the iCare rebound tonometer and the Goldmann applanation tonometer over a wide IOP range.

Authors:  S Munkwitz; A Elkarmouty; E M Hoffmann; N Pfeiffer; H Thieme
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2008-01-12       Impact factor: 3.117

Review 3.  Tonometers-which one should I use?

Authors:  Kanza Aziz; David S Friedman
Journal:  Eye (Lond)       Date:  2018-02-19       Impact factor: 3.775

4.  Measuring intraocular pressure with the Pulsair 3000 and Rebound tonometers in elderly patients without an anesthetic.

Authors:  Antti Kontiola; Päivi Puska
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2003-11-21       Impact factor: 3.117

5.  Detecting IOP Fluctuations in Glaucoma Patients.

Authors:  Brenda Nuyen; Kaweh Mansouri
Journal:  Open Ophthalmol J       Date:  2016-02-29

6.  A comparison of NCT, Goldman application tonometry values with and without fluorescein.

Authors:  Hakika Erdogan; Ziya Akingol; Ozlem Cam; Sadik Sencan
Journal:  Clin Ophthalmol       Date:  2018-10-29

7.  Measurement of Intraocular Pressure with Applanation, Dynamic Contour, and Air-Puff Tonometers: A Comparative Study in Primary Open-Angle Glaucoma and Healthy Cases.

Authors:  Kemal Turgay Özbilen; M Selim Kocabora
Journal:  Beyoglu Eye J       Date:  2020-12-28

8.  Association of intraocular pressure and postoperative nausea and vomiting after microvascular decompression - a prospective cohort study.

Authors:  Yuantao Hou; Hansheng Liang; Cungang Fan; Ruen Liu; Yi Feng
Journal:  BMC Anesthesiol       Date:  2022-04-30       Impact factor: 2.217

9.  Applanation tonometry: a comparison of the Perkins handheld and Goldmann slit lamp-mounted methods.

Authors:  R Arora; H Bellamy; Mw Austin
Journal:  Clin Ophthalmol       Date:  2014-03-26

10.  Comparison of three methods of tonometry in normal subjects: Goldmann applanation tonometer, non-contact airpuff tonometer, and Tono-Pen XL.

Authors:  Ihsan Yilmaz; Cigdem Altan; Ebru Demet Aygit; Cengiz Alagoz; Okkes Baz; Sibel Ahmet; Semih Urvasizoglu; Dilek Yasa; Ahmet Demirok
Journal:  Clin Ophthalmol       Date:  2014-06-07
  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.