| Literature DB >> 35098085 |
Kemal Turgay Özbilen1, M Selim Kocabora2.
Abstract
OBJECTIVES: This study aimed to investigate and compare the reliability of Goldmann applanation tonometer (GAT), dynamic contour (DCT), and noncontact (NCT) tonometers in intraocular pressure (IOP) measurement and the affecting parameters in healthy subjects and cases with primary open-angle glaucoma (POAG).Entities:
Keywords: Intraocular Pressure; Ocular; Ocular Pulse Amplitude; Open-Angle Glaucoma; Pascal Dynamic Contour tonometer; Tonometry
Year: 2020 PMID: 35098085 PMCID: PMC8784467 DOI: 10.14744/bej.2020.82713
Source DB: PubMed Journal: Beyoglu Eye J ISSN: 2459-1777
Demographic distribution and characteristic features of cases
| POAG (n=33) | Healthy (n=31) | Total (n=64) | p | |
|---|---|---|---|---|
| Sex Distribution Female ratio | 17/33 | 15/31 | 32/64 | 0.91 (Chi-square) |
| Mean Age | 57.48±10.51 years | 48.96±7.89 years | 53.36±10.20 years | 0.001 |
| CCT | 561.39±46.12 µ | 561.70±45.31 | 561.54±45.36 µ | 0.98 |
| RNFL-T | 76.60±12.74 µ | 100.39±9.24 µ | 88.12±16.33 µ | 0.001 |
| SAP-MD | -10.64±7.94 Db | -2.00±1.1 Db | -6.48±7.16 Db | 0.001 |
| NCT (IOP1) | 15.72±3.02 mmHg | 15.80±3.99 mmHg | 15.76±3.49 mmHg | 0.93 |
| DCT (IOP2) | 17.84±3.29 mmHg | 17.95±3.86 mmHg | 17.89±3.55 mmHg | 0.91 |
| GAT (IOP3) | 16.48±3.41 mmHg | 16.29±4.14 mmHg | 16.39±3.75 mmHg | 0.84 |
| OPA | 2.54±1.14 mmHg | 2.57±0.94 mmHg | 2.56±1.04 mmHg | 0.89 |
: Independent T-test between groups; RNFL: Retinal nerve fiber layer- thickness; CCT: Central corneal thickness; POAG: Primary open-angle glaucoma; n: number of cases; SAP-MD: Standard automated perimetry- mean deviation; dB: Decibel GAT: Goldman applanation tonometer, NCT: Non-contact tonometer, DCT: Dynamic contour tonometer; OPA: Ocular pulse amplitude.
Binary correlations of GAT / DCT / NCT in Control / POAG groups
| Control/POAG | GAT/DCT | GAT/NCT | DCT/NCT |
|---|---|---|---|
| Control | r=0.939 | r=0.878 | r=0.836 |
| n=31 | p<0.001 | p<0,001 | p<0.001 |
| %95 CI:0.876-0.970 | %95 CI:0,760-0,924 | %95 CI:0.685-0.918 | |
| POAG | r=0.878 | r=0.879 | r=0.743 |
| n=33 | p<0.001 | p<0.001 | p<0.001 |
| %95 CI: 0.744-0.932 | %95 CI: 0.767-0.939 | %95 CI: 0.537-0.866 |
POAG: Primary open-angle glaucoma; GAT: Goldman applanation tonometer; NCT: Non-contact tonometer; DCT: Dynamic contour tonometer; N: Number of cases; CI: Reliability range.
Cross distribution of Control, POAG Groups and CCT’s Thin, Medium and Thick Groups
| Groups | CCT | |||
|---|---|---|---|---|
|
| ||||
| Thin | Medium | Thick | Total | |
| Control | ||||
| n | 6 | 17 | 8 | 31 |
| % total | 19.4 | 54.8 | 25.8 | 100.0 |
| POAG | ||||
| n | 7 | 16 | 10 | 33 |
| % total | 21.2 | 48.5 | 30.3 | 100.0 |
| Total | ||||
| n | 13 | 33 | 18 | 64 |
| % total | 20.3 | 51.6 | 28.1 | 100.0 |
CCT: Central corneal thickness, POAG: Primary open-angle glaucoma
n: Number of cases.
Comparison of measurements in groups of different corneal thicknesses, and determination of the cornea group that makes a difference
| Variables | ANOVA | post-Hoc test (p) | ||
|---|---|---|---|---|
|
| ||||
| Thin- Medium | Thin-Thick | Medium-Thick | ||
| DCT | p=0.003, F=6.37 | 0.814 | 0.017 | 0.001 |
| GAT | p<0.001, F=7.88 | 0.514 | 0.002 | 0.001 |
| NCT | p<0.001, F=12.26 | 0.152 | <0.001 | <0.001 |
| OPA | p=0.245 | |||
: between three corneal groups, GAT: Goldman applanation tonometer, NCT: Non-contact tonometer, DCT: Dynamic contour tonometer, OPA: Ocular pulse amplitude.
Comparison of the Methods in order to their thicknesses with CCT Groups (Bland and Altman Plot)
| Groups acc. to CCT | ||||
|---|---|---|---|---|
|
| ||||
| CCT<520µ (n=13) | 520µ<CCT<580µ (n=33) | CCT>580µ (n=18) | Total (n=64) | |
| Mean. Dif. (mmHg) | ||||
| GAT-DCT (mmHg) | -2.31±1.50 | -1.36±1.24 | -1.17±2.00 | -1.51±1.57 |
| %95 CI | -3.22-1.40 | -1.80-0.93 | -2.70-0.18 | -1.90-1.11 |
| GAT-NCT (mmHg) | 1.23±1.59 | 0.49±1.54 | 0.46±1.66 | 0.63±1.55 |
| %95 CI | 0.26–2.19 | 0.06–1.03 | -0.32–1.29 | 0.23–1.03 |
| DCT-NCT (mmHg) | 3.54±1.90 | 1.85±2.00 | 1.69±1.77 | 2.14±2.02 |
| %95 CI | 2.39-4.70 | 1,14-2.56 | 0.76-2.52 | 1.63–2.64 |
DCT: Dynamic contour tonometer, GAT: Goldmann applanation tonometer, NCT: Non-contact tonometer CCT: Central corneal thickness, CI: Confidence interval, mmHg: millimeter mercury, µ: micron, n: Number of cases.