Literature DB >> 11552787

Does the cleaning technique influence the durability of the <9F flexible ureteroscope?

E M McDougall1, G Alberts, K J Deal, J M Nagy.   

Abstract

BACKGROUND AND
PURPOSE: The fragility of the <9F flexible ureteroscope limits its availability to general urology practice. The purpose of this study was to determine whether the technique used to clean the flexible ureteroscope or the number of persons handling the instrument during the cleaning process influenced endoscope breakage or deterioration during regular endourologic use. PATIENTS AND METHODS: A new Olympus URF/P3 flexible 7.5F ureteroscope was used for each of two 30-day study periods during which a single surgeon used the endoscope for a variety of upper urinary tract procedures. During the first 30-day period (Group 1), the endoscope was leak-proof-pressure tested and cleaned by the endourology support team using the Steris 20 (peroxyacetic acid 35%) technique. During the second 30-day period (Group 2), the endoscope was leak-proof tested and cleaned only by the surgeon using the Cidex (glutaraldehyde 2.4%) technique. A record was kept for each ureteroscopic case to document the patient position, access technique, time the endoscope was in the urinary tract, instruments passed through the ureteroscope, and the maximum irrigant pressure used. In addition, a record was made of the number of broken fibers, the degree of flexion and deflexion of the endoscope, and the problems encountered with the endoscope during the case.
RESULTS: The two study groups were similar in terms of the total number of cases performed, the mean time the endoscope was in the urinary tract per case, the access approach used, and the use of the ureteral access sheath and ancillary equipment. In Group 2, the endoscope was used for a longer total time (618 minutes v 457 minutes), and access to a lower pole calix was more than twice as common as in Group 1. This may explain why more broken fibers were noted in the instrument used in Group 2 over the study period (eight v four broken fibers) than in Group 1. The only breakage occurred as a result of the surgeon accidentally activating the laser probe inside the working channel of the endoscope in Group 2.
CONCLUSION: The technique and number of personnel involved in the maintenance and cleaning of the flexible ureteroscope does not have a significant effect on the durability and function of these instruments. It is the arduous demands of the endourologic procedure that influence the durability of these fragile endoscopes.

Entities:  

Mesh:

Substances:

Year:  2001        PMID: 11552787     DOI: 10.1089/089277901750426409

Source DB:  PubMed          Journal:  J Endourol        ISSN: 0892-7790            Impact factor:   2.942


  9 in total

Review 1.  [Technical innovations in endourological stone therapy].

Authors:  P Honeck; U Nagele; M S Michel
Journal:  Urologe A       Date:  2008-05       Impact factor: 0.639

Review 2.  Which flexible ureteroscope is the best for upper tract urothelial carcinoma treatment?

Authors:  Etienne Xavier Keller; Steeve Doizi; Luca Villa; Olivier Traxer
Journal:  World J Urol       Date:  2019-02-15       Impact factor: 4.226

3.  In vitro evaluation of ScopeSafe fibers and the scope guardian sheath in prevention of ureteroscope endolumenal working damage.

Authors:  Achim Lusch; Corollos Abdelshehid; Michael A Liss; Reza Alipanah; Elspeth M McDougall; Jaime Landman
Journal:  J Endourol       Date:  2013-03-08       Impact factor: 2.942

4.  Defining the Costs of Reusable Flexible Ureteroscope Reprocessing Using Time-Driven Activity-Based Costing.

Authors:  Dylan Isaacson; Tessnim Ahmad; Ian Metzler; David T Tzou; Kazumi Taguchi; Manint Usawachintachit; Samuel Zetumer; Benjamin Sherer; Marshall Stoller; Thomas Chi
Journal:  J Endourol       Date:  2017-09-20       Impact factor: 2.942

Review 5.  Handling and protecting your flexible ureteroscope: how to maximise scope usage.

Authors:  Khaled Hosny; Jennifer Clark; Shalom J Srirangam
Journal:  Transl Androl Urol       Date:  2019-09

6.  FlexGuard: a new laser insertion sheath: functional aspects in ureterorenoscopy (URS).

Authors:  T R W Herrmann; T Bach; F Imkamp; H Tezval; C Klot; U Jonas; A J Gross; M Burchardt
Journal:  World J Urol       Date:  2007-06-05       Impact factor: 3.661

Review 7.  A comprehensive literature-based equation to compare cost-effectiveness of a flexible ureteroscopy program with single-use versus reusable devices.

Authors:  Giovanni S Marchini; Fábio C Torricelli; Carlos A Batagello; Manoj Monga; Fábio C Vicentini; Alexandre Danilovic; Miguel Srougi; William C Nahas; Eduardo Mazzucchi
Journal:  Int Braz J Urol       Date:  2019 Jul-Aug       Impact factor: 1.541

Review 8.  Repair Rate and Associated Costs of Reusable Flexible Ureteroscopes: A Systematic Review and Meta-analysis.

Authors:  Dinah K Rindorf; Thomas Tailly; Guido M Kamphuis; Sara Larsen; Bhaskar K Somani; Olivier Traxer; Kevin Koo
Journal:  Eur Urol Open Sci       Date:  2022-01-29

9.  Durability of current generation flexible ureteroscopes: the experience from a high-volume centre.

Authors:  Chandra Mohan Vaddi; Paidakula Ramakrishna; Soundarya Ganesan; Siddalinga Swamy; Hemnath Anandan; Manas Babu
Journal:  Cent European J Urol       Date:  2022-05-20
  9 in total