Literature DB >> 11527036

Estimation of psychophysical levels using the electrically evoked compound action potential measured with the neural response telemetry capabilities of Cochlear Corporation's CI24M device.

K H Franck1, S J Norton.   

Abstract

OBJECTIVE: The goal of this study was to estimate psychophysical levels using the electrically evoked compound action potential (EAP), measured with the Neural Response Telemetry capabilities of Cochlear Corporation's Nucleus CI24M cochlear implant system.
DESIGN: Twelve postlingually deafened adults with at least 3 mo of implant experience with the CI24M were subjects in this study. EAP growth functions were successfully quantified on each active electrode of every subject. Correlation and regression analyses were performed between EAP measures and cochlear implant fitting psychophysics. Other information including performance, etiology and duration of hearing loss, and individual electrode impedance was considered.
RESULTS: EAP thresholds were found to be highly correlated with psychophysical thresholds. The rate of EAP growth with increasing stimulation levels was also found to be correlated with the dynamic range of loudness limits and psychophysical thresholds in some subjects. No relationship was evident between EAP measures and speech perception tests.
CONCLUSIONS: Information from EAP growth function measurements may be used to estimate psychophysical information used in cochlear implant fitting but not to predict performance with the device.

Entities:  

Mesh:

Year:  2001        PMID: 11527036     DOI: 10.1097/00003446-200108000-00004

Source DB:  PubMed          Journal:  Ear Hear        ISSN: 0196-0202            Impact factor:   3.570


  29 in total

Review 1.  [Intra- and postoperative electrophysiological diagnostics].

Authors:  T Wesarg; S Arndt; A Aschendorff; R Laszig; R Beck; L Jung; S Zirn
Journal:  HNO       Date:  2017-04       Impact factor: 1.284

2.  Psychophysical assessment of stimulation sites in auditory prosthesis electrode arrays.

Authors:  Bryan E Pfingst; Rose A Burkholder-Juhasz; Teresa A Zwolan; Li Xu
Journal:  Hear Res       Date:  2007-11-28       Impact factor: 3.208

3.  Maps created using a new objective procedure (C-NRT) correlate with behavioral, loudness-balanced maps: a study in adult cochlear implant users.

Authors:  Alessandro Scorpecci; Alessandra D'Elia; Paolo Malerba; Italo Cantore; Patrizia Consolino; Franco Trabalzini; Gaetano Paludetti; Nicola Quaranta
Journal:  Eur Arch Otorhinolaryngol       Date:  2016-05-30       Impact factor: 2.503

4.  Recommendations for Measuring the Electrically Evoked Compound Action Potential in Children With Cochlear Nerve Deficiency.

Authors:  Shuman He; Xiuhua Chao; Ruijie Wang; Jianfen Luo; Lei Xu; Holly F B Teagle; Lisa R Park; Kevin D Brown; Michelle Shannon; Cynthia Warner; Angela Pellittieri; William J Riggs
Journal:  Ear Hear       Date:  2020 May/Jun       Impact factor: 3.570

5.  Electrode interaction in pediatric cochlear implant subjects.

Authors:  Marc D Eisen; Kevin H Franck
Journal:  J Assoc Res Otolaryngol       Date:  2005-06-10

6.  Comparison of electrically evoked compound action potential thresholds and loudness estimates for the stimuli used to program the Advanced Bionics cochlear implant.

Authors:  Eun Kyung Jeon; Carolyn J Brown; Christine P Etler; Sara O'Brien; Li-Kuei Chiou; Paul J Abbas
Journal:  J Am Acad Audiol       Date:  2010-01       Impact factor: 1.664

7.  TNRT profiles with the nucleus research platform 8 system.

Authors:  W K Lai; N Dillier; B P Weber; T Lenarz; R Battmer; B Gantz; C Brown; N Cohen; S Waltzman; M Skinner; L Holden; R Cowan; P Busby; M Killian
Journal:  Int J Audiol       Date:  2009       Impact factor: 2.117

8.  Determining electrically evoked compound action potential thresholds: a comparison of computer versus human analysis methods.

Authors:  E Katelyn Glassman; Michelle L Hughes
Journal:  Ear Hear       Date:  2013 Jan-Feb       Impact factor: 3.570

9.  Long-term evolution of the electrical stimulation levels for cochlear implant patients.

Authors:  Jose Luis Vargas; Manuel Sainz; Cristina Roldan; Isaac Alvarez; Angel de la Torre
Journal:  Clin Exp Otorhinolaryngol       Date:  2012-11-13       Impact factor: 3.372

10.  Site of cochlear stimulation and its effect on electrically evoked compound action potentials using the MED-EL standard electrode array.

Authors:  Stefan Brill; Joachim Müller; Rudolf Hagen; Alexander Möltner; Steffi-Johanna Brockmeier; Thomas Stark; Silke Helbig; Jan Maurer; Thomas Zahnert; Clemens Zierhofer; Peter Nopp; Ilona Anderson; Stefan Strahl
Journal:  Biomed Eng Online       Date:  2009-12-16       Impact factor: 2.819

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.