Literature DB >> 19513917

TNRT profiles with the nucleus research platform 8 system.

W K Lai1, N Dillier, B P Weber, T Lenarz, R Battmer, B Gantz, C Brown, N Cohen, S Waltzman, M Skinner, L Holden, R Cowan, P Busby, M Killian.   

Abstract

This study investigates the effect of the Nucleus CI24RE implant's neural response telemetry (NRT) system, which has less internal noise compared to its predecessor, the CI24M/R implant, on the NRT threshold (TNRT) profile across the array. CI24M/R measurements were simulated by ignoring CI24RE measurements with response amplitudes below 50 uV. Comparisons of the estimated TNRTs from the CI24RE measurements and the CI24M/R simulations suggest that, apart from a constant level difference, the TNRT profiles from the newer implant generally would not have differed very much from those of its predecessor. This view was also reflected by principal component analysis (PCA) results which revealed a 'shift' component similar to that reported by Smoorenburg et al (2002). On the whole, there is no indication that current practices of using the TNRT profiles for assisting with speech processor programming need to be revised for the CI24RE implant.

Entities:  

Mesh:

Year:  2009        PMID: 19513917      PMCID: PMC4255591          DOI: 10.1080/14992020902962413

Source DB:  PubMed          Journal:  Int J Audiol        ISSN: 1499-2027            Impact factor:   2.117


  14 in total

1.  Comparison of EAP thresholds with MAP levels in the nucleus 24 cochlear implant: data from children.

Authors:  M L Hughes; C J Brown; P J Abbas; A A Wolaver; J P Gervais
Journal:  Ear Hear       Date:  2000-04       Impact factor: 3.570

2.  The relationship between EAP and EABR thresholds and levels used to program the nucleus 24 speech processor: data from adults.

Authors:  C J Brown; M L Hughes; B Luk; P J Abbas; A Wolaver; J Gervais
Journal:  Ear Hear       Date:  2000-04       Impact factor: 3.570

3.  Estimation of psychophysical levels using the electrically evoked compound action potential measured with the neural response telemetry capabilities of Cochlear Corporation's CI24M device.

Authors:  K H Franck; S J Norton
Journal:  Ear Hear       Date:  2001-08       Impact factor: 3.570

4.  Neural response telemetry in 12- to 24-month-old children.

Authors:  Karen A Gordon; Kiara A Ebinger; Jan E Gilden; William H Shapiro
Journal:  Ann Otol Rhinol Laryngol Suppl       Date:  2002-05

5.  Speech perception in nucleus CI24M cochlear implant users with processor settings based on electrically evoked compound action potential thresholds.

Authors:  Guido F Smoorenburg; Christina Willeboer; Johannes E van Dijk
Journal:  Audiol Neurootol       Date:  2002 Nov-Dec       Impact factor: 1.854

6.  Normative findings of electrically evoked compound action potential measurements using the neural response telemetry of the Nucleus CI24M cochlear implant system.

Authors:  D Cafarelli Dees; N Dillier; W K Lai; E von Wallenberg; B van Dijk; F Akdas; M Aksit; C Batman; A Beynon; S Burdo; J-M Chanal; L Collet; M Conway; C Coudert; L Craddock; H Cullington; N Deggouj; B Fraysse; S Grabel; J Kiefer; J G Kiss; T Lenarz; A Mair; S Maune; J Müller-Deile; J-P Piron; S Razza; C Tasche; H Thai-Van; F Toth; E Truy; A Uziel; G F Smoorenburg
Journal:  Audiol Neurootol       Date:  2005-01-12       Impact factor: 1.854

7.  Evaluation of the neural response telemetry (NRT) capabilities of the nucleus research platform 8: initial results from the NRT trial.

Authors:  Rolf D Battmer; Norbert Dillier; Wai K Lai; Benno P Weber; Carolyn Brown; Bruce J Gantz; J Thomas Roland; Noel J Cohen; William Shapiro; Joerg Pesch; Matthijs J Killian; Thomas Lenarz
Journal:  Int J Audiol       Date:  2004-12       Impact factor: 2.117

8.  Use of neural response telemetry measures to objectively set the comfort levels in the Nucleus 24 cochlear implant.

Authors:  John E King; Marek Polak; Annelle V Hodges; Stacy Payne; Fred F Telischi
Journal:  J Am Acad Audiol       Date:  2006-06       Impact factor: 1.664

9.  Relation between neural response telemetry thresholds, T- and C-levels, and loudness judgments in 12 adult nucleus 24 cochlear implant recipients.

Authors:  Lisa G Potts; Margaret W Skinner; Brenda D Gotter; Michael J Strube; Chris A Brenner
Journal:  Ear Hear       Date:  2007-08       Impact factor: 3.570

10.  Intra-operative recordings of electrically evoked auditory nerve action potentials in young children by use of neural response telemetry with the nucleus C124M cochlear implant.

Authors:  S M Mason; Y Cope; J Garnham; G M O'Donoghue; K P Gibbin
Journal:  Br J Audiol       Date:  2001-08
View more