Tissue damage and calcification may be independent mechanisms of bioprosthetic heart valve failure.

I Vesely1, J E Barber, N B Ratliff.   

Abstract

BACKGROUND AND AIM OF THE STUDY: Porcine bioprosthetic valves have excellent hemodynamics and do not require anticoagulation, but have limited durability. Cusp tearing is a major cause of bioprosthetic valve failure. It has been suggested that the mechanism of bioprosthetic valve failure is stiffening by calcification, which leads to elevated stresses and secondary collagen fiber damage and leaflet tearing. This thesis was tested in explanted porcine bioprostheses.
METHODS: A total of 60 explanted porcine bioprosthetic valves was tested mechanically, and 15 explanted valves were examined grossly and histologically. Circumferentially and radially oriented samples of cusp tissue were tested uniaxially in a materials testing machine and compared with five controls.
RESULTS: Mean (+/-SD) duration of implantation was 10.9+/-5.6 years. Circumferential specimens from explants were less extensible than controls (11.0+/-5.5% versus 24.5+/-2.8% strain, p <0.001), and failed at lower tensions (973+/-733 versus 3075+/-911 N/m, p = 0.001) and at lower strains (21.2+/-8.1% versus 47.3+/-7.1% strain, p <0.001). Radial specimens from explants were less extensible (28.7+/-6.8% versus 39.2+/-5.9% strain, p = 0.002) and failed at lower strains (60.3+/-17.3% versus 112.2+/-24.9% strain, p <0.001) than the controls. The stiffness of the explants was unchanged from controls in both circumferential and radial samples. There were no differences between explants and controls in radial and circumferential stiffness, and in radial failure strength. Calcification was mild and diffuse in most of the tested samples. Tears were found in areas without calcific deposits, along with breaks in collagen fiber bundles.
CONCLUSION: These results do not support the thesis that calcification stiffens glutaraldehyde-fixed porcine bioprostheses, except when the entire cusp is transformed into a solid mass of mineral. Rather, leaflet tears may develop as a result of accumulated mechanical damage that is independent of calcification.

Entities:  

Mesh:

Substances:

Year:  2001        PMID: 11499593

Source DB:  PubMed          Journal:  J Heart Valve Dis        ISSN: 0966-8519


  14 in total

1.  Development and validation of implantable sensors for monitoring function of prosthetic heart valves: in vitro studies.

Authors:  C Lanning; R Shandas
Journal:  Med Biol Eng Comput       Date:  2003-07       Impact factor: 2.602

2.  A novel fibre-ensemble level constitutive model for exogenous cross-linked collagenous tissues.

Authors:  Michael S Sacks; Will Zhang; Silvia Wognum
Journal:  Interface Focus       Date:  2016-02-06       Impact factor: 3.906

Review 3.  Tissue Valve Degeneration and Mechanical Valve Failure.

Authors:  Andrew C W Baldwin; George Tolis
Journal:  Curr Treat Options Cardiovasc Med       Date:  2019-06-14

4.  Evaluation of bioprosthetic heart valve failure using a matrix-fibril shear stress transfer approach.

Authors:  Afshin Anssari-Benam; Asa H Barber; Andrea Bucchi
Journal:  J Mater Sci Mater Med       Date:  2015-12-29       Impact factor: 3.896

5.  Immunoproteomic Identification of Noncarbohydrate Antigens Eliciting Graft-Specific Adaptive Immune Responses in Patients with Bovine Pericardial Bioprosthetic Heart Valves.

Authors:  Katherine V Gates; Qi Xing; Leigh G Griffiths
Journal:  Proteomics Clin Appl       Date:  2018-12-19       Impact factor: 3.494

6.  Effects of cyclic flexural fatigue on porcine bioprosthetic heart valve heterograft biomaterials.

Authors:  Ali Mirnajafi; Brett Zubiate; Michael S Sacks
Journal:  J Biomed Mater Res A       Date:  2010-07       Impact factor: 4.396

Review 7.  Biomechanical Behavior of Bioprosthetic Heart Valve Heterograft Tissues: Characterization, Simulation, and Performance.

Authors:  Joao S Soares; Kristen R Feaver; Will Zhang; David Kamensky; Ankush Aggarwal; Michael S Sacks
Journal:  Cardiovasc Eng Technol       Date:  2016-08-09       Impact factor: 2.495

8.  Neomycin fixation followed by ethanol pretreatment leads to reduced buckling and inhibition of calcification in bioprosthetic valves.

Authors:  Devanathan Raghavan; Sagar R Shah; Naren R Vyavahare
Journal:  J Biomed Mater Res B Appl Biomater       Date:  2010-01       Impact factor: 3.368

Review 9.  On the biomechanical function of scaffolds for engineering load-bearing soft tissues.

Authors:  John A Stella; Antonio D'Amore; William R Wagner; Michael S Sacks
Journal:  Acta Biomater       Date:  2010-01-07       Impact factor: 8.947

10.  Characterizing the collagen fiber orientation in pericardial leaflets under mechanical loading conditions.

Authors:  S Hamed Alavi; Victor Ruiz; Tatiana Krasieva; Elliot L Botvinick; Arash Kheradvar
Journal:  Ann Biomed Eng       Date:  2012-11-21       Impact factor: 3.934

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.