Literature DB >> 715402

The improved Björk-Shiley tilting disc valve prosthesis.

V O Björk.   

Abstract

The Björk-Shiley valve has been improved in three respects: (1) Increased strength by making the inlet strut an integral part of the orifice ring and doubling its cross-section area; (2) Improved hydrodynamics; (3) Elimination of the area of stagnant and low flow behind the disc. By utilizing a convexo-concave disc and by moving the pivot point downstream, the disc in open position is moved further out of the orifice ring. The flow through the smaller hole is thus increased by 40%. A significant clearance is obtained between the disc and the valve ring in open position. 234 of these valves have been inserted and followed-up for a maximum period of two years with excellent results.

Mesh:

Year:  1978        PMID: 715402     DOI: 10.3109/14017437809100353

Source DB:  PubMed          Journal:  Scand J Thorac Cardiovasc Surg        ISSN: 0036-5580


  3 in total

1.  Outlet strut fracture of Björk-Shiley convexo concave heart valves: the UK cohort study.

Authors:  R Z Omar; L S Morton; D A Halliday; E M Danns; M T Beirne; W J Blot; K M Taylor
Journal:  Heart       Date:  2001-07       Impact factor: 5.994

2.  A case of strut fracture of the new Björk-Shiley mitral valve prosthesis.

Authors:  J Ninomiya; N Yamate; T Mizutani; R Soga; T Shoji
Journal:  Jpn J Surg       Date:  1983-09

3.  Early clinical evaluation of a new Björk-Shiley valve prosthesis with integral monostrut.

Authors:  S Nakano; H Hirose; H Matsuda; K Taniguchi; T Kawamoto; Y Kawashima
Journal:  Heart Vessels       Date:  1985-11       Impact factor: 2.037

  3 in total