Improvement of the bone-pin interface strength in osteoporotic bone with use of hydroxyapatite-coated tapered external-fixation pins. A prospective, randomized clinical study of wrist fractures.

A Moroni1, C Faldini, S Marchetti, M Manca, V Consoli, S Giannini.   

Abstract

BACKGROUND: Achieving adequate fixation strength in osteoporotic bone is a challenge. In this study, we examined the use of hydroxyapatite-coated tapered external-fixation pins for the fixation of wrist fractures in patients with osteoporosis.
METHODS: Twenty female patients with osteoporosis and a fracture of the wrist were divided into two paired groups and randomized to receive either standard tapered pins or hydroxyapatite-coated tapered pins. Two pins were inserted in the distal part of the radius, two pins were inserted in the second metacarpal, and an external fixation device was mounted. All fixation devices were removed six weeks after surgery.
RESULTS: The mean pin-insertion torque (and standard deviation) was 461 +/- 254 Nmm in the group managed with standard pins and 332 +/- 176 Nmm in the group managed with hydroxyapatite-coated pins (p = 0.01). The mean pin-extraction torque was 191 +/- 155 Nmm in the group managed with standard pins and 600 +/- 214 Nmm in the group managed with hydroxyapatite-coated pins (p < 0.0001, power 95%). The mean extraction torque was lower than the corresponding insertion torque at each pin position in the group managed with standard pins (p < 0.05), whereas the mean extraction torque was higher than the corresponding insertion torque at each pin position in the group managed with hydroxyapatite-coated pins (p = 0.001). Two patients managed with standard pins and no patient managed with hydroxyapatite-coated pins had a pin-track infection. Pain during pin removal did not differ between the two groups.
CONCLUSIONS: The present study showed that hydroxyapatite-coated tapered external-fixation pins provided improved fixation in the treatment of wrist fractures in patients with osteoporosis.

Entities:  

Mesh:

Substances:

Year:  2001        PMID: 11379741     DOI: 10.2106/00004623-200105000-00010

Source DB:  PubMed          Journal:  J Bone Joint Surg Am        ISSN: 0021-9355            Impact factor:   5.284


  13 in total

1.  Microstructural analysis of implant-bone interface of hydroxyapatite-coated and uncoated Schanz screws.

Authors:  M Cimerman; A Cör; M Ceh; A Kristan; J Pizem; M Tonin
Journal:  J Mater Sci Mater Med       Date:  2005-07       Impact factor: 3.896

2.  Limb salvage reconstruction of the ankle with fusion and simultaneous tibial lengthening using the Ilizarov/Taylor spatial frame.

Authors:  Nazzar Tellisi; Austin T Fragomen; Svetlana Ilizarov; S Robert Rozbruch
Journal:  HSS J       Date:  2007-12-08

3.  Bioinspired polydopamine and polyphenol tannic acid functionalized titanium suppress osteoclast differentiation: a facile and efficient strategy to regulate osteoclast activity at bone-implant interface.

Authors:  Chris Steffi; Zhilong Shi; Chee Hoe Kong; Wilson Wang
Journal:  J R Soc Interface       Date:  2019-03-29       Impact factor: 4.118

4.  Effects of Ti2448 half-pin with low elastic modulus on pin loosening in unilateral external fixation.

Authors:  Kai Zheng; Xiaokang Li; Jun Fu; Xiangli Fan; Pan Wang; Yulin Hao; Shujun Li; Hongbin Fan; Zheng Guo
Journal:  J Mater Sci Mater Med       Date:  2011-04-13       Impact factor: 3.896

5.  Surgical approach to bone healing in osteoporosis.

Authors:  Vito Pesce; Domenico Speciale; Giulio Sammarco; Silvio Patella; Antonio Spinarelli; Vittorio Patella
Journal:  Clin Cases Miner Bone Metab       Date:  2009-05

6.  Use of chlorhexidine-impregnated patch at pin site to reduce local morbidity: the ChIPPS Pilot Trial.

Authors:  Stephanie C Wu; Ryan T Crews; Charles Zelen; James S Wrobel; David G Armstrong
Journal:  Int Wound J       Date:  2008-06-01       Impact factor: 3.315

7.  Chitosan-coated stainless steel screws for fixation in contaminated fractures.

Authors:  Alex H Greene; Joel D Bumgardner; Yunzhi Yang; Jon Moseley; Warren O Haggard
Journal:  Clin Orthop Relat Res       Date:  2008-04-29       Impact factor: 4.176

8.  Application of the skin and bone integrated pylon with titanium oxide nanotubes and seeded with dermal fibroblasts.

Authors:  Maxim A Shevtsov; Natalia Yudintceva; Miralda Blinova; George Pinaev; Oleg Galibin; Igor Potokin; Ketul C Popat; Mark Pitkin
Journal:  Prosthet Orthot Int       Date:  2014-09-23       Impact factor: 1.895

9.  Initial clinical trial of pins coated with fibroblast growth factor-2-apatite composite layer in external fixation of distal radius fractures.

Authors:  Yohei Yanagisawa; Atsuo Ito; Yuki Hara; Hirotaka Mutsuzaki; Shinji Murai; Kengo Fujii; Yu Sogo; Motohiro Hirose; Ayako Oyane; Fumiko Kobayashi; Masashi Yamazaki
Journal:  J Orthop       Date:  2018-12-20

Review 10.  Different methods of external fixation for treating distal radial fractures in adults.

Authors:  H H G Handoll; J S Huntley; R Madhok
Journal:  Cochrane Database Syst Rev       Date:  2008-01-23
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.