Literature DB >> 11307380

Influence of rotational speed, torque and operator's proficiency on ProFile failures.

G M Yared1, F E Bou Dagher, P Machtou.   

Abstract

AIM: The purpose of this study was to evaluate the influence of rotational speed, torque, and operator experience with a specific Ni-Ti rotary instrumentation technique on the incidence of locking, deformation and separation of instruments.
METHODOLOGY: ProFile Ni-Ti rotary instruments (PRI) sizes 40-15 with a 6% taper were used in a crown-down technique. In one group of canals (n = 300) speeds of 150, 250 and 350 rpm (subgroups 1, 2 and 3) were used. Each one of the subgroups included 100 canals. In a second group (n = 300) torque was set at 20, 30 and 55 Ncm (subgroups 4, 5 and 6). In the third group (n = 300) three operators with varying experience (subgroups 7, 8 and 9) were also compared. Each subgroup included the use of 10 sets of PRI and 100 canals of extracted human molars. Each set of PRI was used in up to 10 canals and then sterilized before each case. NaOCl 2.5% was used as an irrigant. The number of locked, deformed, and separated instruments for the different groups, and within each part of the study was analysed statistically for significance with chi-squared tests.
RESULTS: In group 1 only one instrument was deformed in the 150-rpm group and no instruments separated or locked. In the 250-rpm group instrument separation did not occur, however, a high incidence of locking, deformation and separation was noted in the 350-rpm group. In general, instrument sizes 30-15 locked, deformed and separated. Chi-squared statistics showed a significant difference between the 150 and 350 rpm groups but no difference between the 150 and 250 rpm groups with regard to instrument separation. Overall, there was a trend toward a higher incidence of instrument deformation and separation in smaller instruments. Locking and separation occurred during the final passage of the instruments, in the last (tenth) canal in each subgroup. In the second group, neither separation nor deformation and locking occurred during the use of the ProFile instruments, at 150 rpm, and at the different torque values. In the third group, chi-squared analysis demonstrated that significantly more instruments separated with the least experienced operator. Instrument locking, deformation, and separation did not occur with the most experienced operator.
CONCLUSIONS: Preclinical training in the use of the PRI technique with crown-down at 150 rpm were crucial in avoiding instrument separation and reducing the incidence of instrument locking and deformation.

Entities:  

Mesh:

Substances:

Year:  2001        PMID: 11307380     DOI: 10.1046/j.1365-2591.2001.00352.x

Source DB:  PubMed          Journal:  Int Endod J        ISSN: 0143-2885            Impact factor:   5.264


  23 in total

Review 1.  Endodontic Rotary Files, What Should an Endodontist Know?

Authors:  Ana-Belén Dablanca-Blanco; Pablo Castelo-Baz; Ramón Miguéns-Vila; Pablo Álvarez-Novoa; Benjamín Martín-Biedma
Journal:  Medicina (Kaunas)       Date:  2022-05-27       Impact factor: 2.948

2.  Evaluation of the efficacy of rotary vs. hand files in root canal preparation of primary teeth in vitro using CBCT.

Authors:  P K Musale; S A V Mujawar
Journal:  Eur Arch Paediatr Dent       Date:  2013-07-27

3.  Canal shaping with WaveOne reciprocating files: influence of operator experience on instrument breakage and canal preparation time.

Authors:  Luigi Generali; Elena Righi; Maria Vittoria Todesca; Ugo Consolo
Journal:  Odontology       Date:  2013-06-28       Impact factor: 2.634

4.  In Vivo Evaluation of Operative Torque Generated by Two Nickel-Titanium Rotary Instruments during Root Canal Preparation.

Authors:  Gianluca Gambarini; Massimo Galli; Marco Seracchiani; Dario Di Nardo; Marco A Versiani; Lucila Piasecki; Luca Testarelli
Journal:  Eur J Dent       Date:  2019-12-31

5.  Cyclic fatigue resistance of Mtwo NiTi rotary instruments used by experienced and novice operators--an in vivo and in vitro study.

Authors:  Gianluca Plotino; Dina Al-Sudani; Stefano Pulino; Nicola M Grande; Piero Alessandro Marcoli; Silvia Pizzi; Luca Testarelli; Gianluca Gambarini
Journal:  Med Sci Monit       Date:  2012-06

6.  An in vitro comparison of root canal system prepared with either hand or rotary instruments.

Authors:  Azza A Dafalla; Neamat Hassan Abubakr; Yahia E Ibrahim
Journal:  Iran Endod J       Date:  2010-11-15

7.  A comparative evaluation of cyclic fatigue resistance for different endodontic NiTi rotary files: An in-vitro study.

Authors:  Akanksha Bhatt; Balakrishnan Rajkumar
Journal:  J Oral Biol Craniofac Res       Date:  2019-01-02

8.  Cyclic Fatigue Resistance and Fractographic Analysis of Race and Protaper Rotary NiTi Instruments.

Authors:  Shahram Azimi; Parisa Delvari; Hamid Cyrus Hajarian; Mohammad Ali Saghiri; Kasra Karamifar; Mehrdad Lotfi
Journal:  Iran Endod J       Date:  2011-05-15

9.  Failure of ProTaper rotary Ni-Ti instruments used by undergraduate students.

Authors:  Marc García-Font; Fernando Duran-Sindreu; Sergio Morello-Castro; Montserrat Mercade-Bellido; Rufino Bueno-Martínez; Miguel Roig-Cayón
Journal:  J Clin Exp Dent       Date:  2012-10-01

10.  Influence of operator's experience level on lifespan of the WaveOne Primary file in extracted teeth.

Authors:  Abdulrahman Mohammed Saleh; Saeid Tavanafar; Pouyan Vakili-Gilani; Noor Jamal Al Sammerraie; Faahim Rashid
Journal:  Restor Dent Endod       Date:  2013-11-12
View more