Literature DB >> 11222348

Clinical comparison of frequency doubling technology perimetry and Humphrey perimetry.

R Casson, B James, A Rubinstein, H Ali.   

Abstract

AIM: To determine the number of missed points on frequency doubling technology (FDT) perimetry that optimise the sensitivity and specificity of the test and to determine the topographical accuracy of the test in a clinical setting.
METHODS: In a prospective study, the perimetric data from 99 patients who underwent both FDT perimetry in the screening mode and Humphrey 24-2 (H24-2) were used to determine the sensitivity and specificity of the FDT perimetry compared with the full threshold H24-2 as the gold standard.
RESULTS: Missed points on the FDT perimetry correlated with both the mean deviation and the corrected pattern standard deviation on the Humphrey perimetry. A score assigned to abnormal points on the FDT perimetry and the Humphrey total deviation plot showed a significant correlation for both the location and the depth of the defect. In comparing the Humphrey hemifield test with the FDT perimetry results, if at least one missed point on the frequency doubling test was considered as abnormal then the overall sensitivity of the test was 78.1% and the specificity was 89.1%.
CONCLUSION: FDT perimetry in the screening mode performed in a clinical setting was highly specific, exhibited reasonable sensitivity, and accurately determined the location and depth of scotomas when compared with the full threshold Humphrey 24-2.

Entities:  

Mesh:

Year:  2001        PMID: 11222348      PMCID: PMC1723884          DOI: 10.1136/bjo.85.3.360

Source DB:  PubMed          Journal:  Br J Ophthalmol        ISSN: 0007-1161            Impact factor:   4.638


  13 in total

1.  Frequency doubling technology perimetry for detection of glaucomatous visual field loss.

Authors:  K E Cello; J M Nelson-Quigg; C A Johnson
Journal:  Am J Ophthalmol       Date:  2000-03       Impact factor: 5.258

2.  Comparison of frequency doubling perimetry with humphrey visual field analysis in a glaucoma practice.

Authors:  Y Burnstein; N J Ellish; M Magbalon; E J Higginbotham
Journal:  Am J Ophthalmol       Date:  2000-03       Impact factor: 5.258

3.  Algorithm for interpreting the results of frequency doubling perimetry.

Authors:  S C Patel; D S Friedman; P Varadkar; A L Robin
Journal:  Am J Ophthalmol       Date:  2000-03       Impact factor: 5.258

4.  Screening for glaucoma with frequency-doubling technology and Damato campimetry.

Authors:  N Yamada; P P Chen; R P Mills; M M Leen; M F Lieberman; R L Stamper; D C Stanford
Journal:  Arch Ophthalmol       Date:  1999-11

5.  Testing for glaucoma with the frequency-doubling illusion in the whole, macular and eccentric visual fields.

Authors:  T Maddess; W L Severt
Journal:  Aust N Z J Ophthalmol       Date:  1999 Jun-Aug

6.  Frequency doubling perimetry in glaucoma early diagnosis.

Authors:  P Brusini; P Busatto
Journal:  Acta Ophthalmol Scand Suppl       Date:  1998

7.  Chronic human glaucoma causing selectively greater loss of large optic nerve fibers.

Authors:  H A Quigley; G R Dunkelberger; W R Green
Journal:  Ophthalmology       Date:  1988-03       Impact factor: 12.079

8.  Identification of glaucoma-related visual field abnormality with the screening protocol of frequency doubling technology.

Authors:  H A Quigley
Journal:  Am J Ophthalmol       Date:  1998-06       Impact factor: 5.258

9.  Clinical classification of glaucomatous visual field loss by frequency doubling perimetry.

Authors:  W E Sponsel; S Arango; Y Trigo; J Mensah
Journal:  Am J Ophthalmol       Date:  1998-06       Impact factor: 5.258

10.  Differential vulnerability of neurochemically identified subpopulations of retinal neurons in a monkey model of glaucoma.

Authors:  J C Vickers; R A Schumer; S M Podos; R F Wang; B M Riederer; J H Morrison
Journal:  Brain Res       Date:  1995-05-22       Impact factor: 3.252

View more
  15 in total

1.  The distribution of visual field defects per quadrant in standard automated perimetry as compared to frequency doubling technology perimetry.

Authors:  Wadih M Zein; Ziad F Bashshur; Rola F Jaafar; Baha' N Noureddin
Journal:  Int Ophthalmol       Date:  2010-10-06       Impact factor: 2.031

2.  Characteristics of the normative database for the Humphrey matrix perimeter.

Authors:  Andrew John Anderson; Chris A Johnson; Murray Fingeret; John L Keltner; Paul G D Spry; Michael Wall; John S Werner
Journal:  Invest Ophthalmol Vis Sci       Date:  2005-04       Impact factor: 4.799

3.  Performance of the 24-2-5 frequency doubling technology screening test: a prospective case study.

Authors:  P G D Spry; H M Hussin; J M Sparrow
Journal:  Br J Ophthalmol       Date:  2007-03-27       Impact factor: 4.638

Review 4.  Glaucoma medications: use and safety in the elderly population.

Authors:  Elliott Kanner; James C Tsai
Journal:  Drugs Aging       Date:  2006       Impact factor: 3.923

5.  Meta-analysis of the diagnostic efficiency of frequency-doubling technology for primary glaucoma.

Authors:  Ting Liu; Xiangge He
Journal:  Front Med China       Date:  2007-02

6.  Frequency doubling technique perimetry and spectral domain optical coherence tomography in patients with early glaucoma.

Authors:  F K Horn; C Y Mardin; D Bendschneider; A G Jünemann; W Adler; R P Tornow
Journal:  Eye (Lond)       Date:  2010-11-19       Impact factor: 3.775

7.  [Conventional perimetry. Antiquated or indispensable for functional glaucoma diagnostics?].

Authors:  F Tonagel; B Voykov; U Schiefer
Journal:  Ophthalmologe       Date:  2012-04       Impact factor: 1.059

8.  [Sensitivity and specificity of flicker perimetry with Pulsar. Comparison with achromatic (white-on-white) perimetry in glaucoma patients].

Authors:  K Göbel; C Erb
Journal:  Ophthalmologe       Date:  2013-02       Impact factor: 1.059

9.  Using frequency-doubling perimetry to detect optic neuropathy in patients with Graves' orbitopathy.

Authors:  Mário L R Monteiro; André L F Portes; Frederico C Moura; Dina B W Regensteiner
Journal:  Jpn J Ophthalmol       Date:  2008-12-17       Impact factor: 2.447

10.  Comparison of frequency doubling and flicker defined form perimetry in early glaucoma.

Authors:  Folkert K Horn; Vicki Scharch; Christian Y Mardin; Robert Lämmer; Jan Kremers
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2016-02-17       Impact factor: 3.117

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.