Literature DB >> 11132787

Speech perception as a function of electrical stimulation rate: using the Nucleus 24 cochlear implant system.

A E Vandali1, L A Whitford, K L Plant, G M Clark.   

Abstract

OBJECTIVE: To investigate the effect of varying electrical stimulation rate on speech comprehension by cochlear implant users, while keeping the number of stimulated channels constant.
DESIGN: Three average rates of electrical stimulation, 250, 807, and 1615 pulses per second per channel (pps/ch), were compared using a speech processing strategy that employed an electrode selection technique similar to that used in the Spectral Maxima Sound Processor strategy (McDermott, McKay, & Vandali, 1992; McDermott & Vandali, Reference Note 1; McKay, McDermott, Vandali, & Clark, 1991) and the Spectral Peak strategy (Skinner et al., 1994; Whitford et al., 1995). Speech perception tests with five users of the Nucleus 24 cochlear implant system were conducted over a 21-wk period. Subjects were given take-home experience with each rate condition. A repeated ABC evaluation protocol with alternating order was employed so as to account for learning effects and to minimize order effects. Perception of open-set monosyllabic words in quiet and open-set sentences at signal to noise ratios ranging from +20 to 0 dB, depending on the subject's ability, were tested. A comparative performance questionnaire was also administered.
RESULTS: No statistical differences in group performance between the 250 and 807 pps/ch rates were observed in any of the speech perception tests. However, significantly poorer group performance was observed for the 1615 pps/ch rate for some tests due predominantly to the results of one subject. Analysis of individual scores showed considerable variation across subjects. For some subjects, one or more of the three rate conditions evaluated provided benefits on some speech perception tasks. The results of the comparative performance questionnaire indicated a preference for the 250 and 807 pps/ch rates over the 1615 pps/ch rate for most listening situations.
CONCLUSIONS: For the speech processing strategy, implant system, and subjects evaluated in this study, the group results indicated that the use of electrical stimulation rates higher than 250 pps/ch (up to 1615 pps/ch) generally provided no significant improvement to speech comprehension. However, individual results indicated that perceptual benefits could be obtained by adjusting rate of stimulation optimally to suit each subject. Results from one subject, together with tinnitus problems arising from high-rate stimulation for another subject, indicated that high rates of stimulation may in fact be undesirable for some subjects.

Entities:  

Mesh:

Year:  2000        PMID: 11132787     DOI: 10.1097/00003446-200012000-00008

Source DB:  PubMed          Journal:  Ear Hear        ISSN: 0196-0202            Impact factor:   3.570


  76 in total

1.  Studies on bilateral cochlear implants at the University of Wisconsin's Binaural Hearing and Speech Laboratory.

Authors:  Ruth Y Litovsky; Matthew J Goupell; Shelly Godar; Tina Grieco-Calub; Gary L Jones; Soha N Garadat; Smita Agrawal; Alan Kan; Ann Todd; Christi Hess; Sara Misurelli
Journal:  J Am Acad Audiol       Date:  2012-06       Impact factor: 1.664

2.  A point process framework for modeling electrical stimulation of the auditory nerve.

Authors:  Joshua H Goldwyn; Jay T Rubinstein; Eric Shea-Brown
Journal:  J Neurophysiol       Date:  2012-06-06       Impact factor: 2.714

3.  Effect of stimulation rate on cochlear implant users' phoneme, word and sentence recognition in quiet and in noise.

Authors:  Robert V Shannon; Rachel J Cruz; John J Galvin
Journal:  Audiol Neurootol       Date:  2010-07-17       Impact factor: 1.854

4.  Multi-microphone adaptive noise reduction strategies for coordinated stimulation in bilateral cochlear implant devices.

Authors:  Kostas Kokkinakis; Philipos C Loizou
Journal:  J Acoust Soc Am       Date:  2010-05       Impact factor: 1.840

Review 5.  Temporal Considerations for Stimulating Spiral Ganglion Neurons with Cochlear Implants.

Authors:  Jason Boulet; Mark White; Ian C Bruce
Journal:  J Assoc Res Otolaryngol       Date:  2016-02

6.  Evaluation of a spectral subtraction strategy to suppress reverberant energy in cochlear implant devices.

Authors:  Kostas Kokkinakis; Christina Runge; Qudsia Tahmina; Yi Hu
Journal:  J Acoust Soc Am       Date:  2015-07       Impact factor: 1.840

7.  Effect of stimulus level on the temporal response properties of the auditory nerve in cochlear implants.

Authors:  Michelle L Hughes; Sarah A Laurello
Journal:  Hear Res       Date:  2017-06-13       Impact factor: 3.208

8.  Speech recognition by bilateral cochlear implant users in a cocktail-party setting.

Authors:  Philipos C Loizou; Yi Hu; Ruth Litovsky; Gongqiang Yu; Robert Peters; Jennifer Lake; Peter Roland
Journal:  J Acoust Soc Am       Date:  2009-01       Impact factor: 1.840

9.  The relation between auditory-nerve temporal responses and perceptual rate integration in cochlear implants.

Authors:  Michelle L Hughes; Jacquelyn L Baudhuin; Jenny L Goehring
Journal:  Hear Res       Date:  2014-08-02       Impact factor: 3.208

10.  Factors Affecting Outcomes in Cochlear Implant Recipients Implanted With a Perimodiolar Electrode Array Located in Scala Tympani.

Authors:  Laura K Holden; Jill B Firszt; Ruth M Reeder; Rosalie M Uchanski; Noël Y Dwyer; Timothy A Holden
Journal:  Otol Neurotol       Date:  2016-12       Impact factor: 2.311

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.