Literature DB >> 11130644

Effect of sterilization method and other modifications on the wear resistance of acetabular cups made of ultra-high molecular weight polyethylene. A hip-simulator study.

H McKellop1, F W Shen, B Lu, P Campbell, R Salovey.   

Abstract

BACKGROUND: Wear of ultra-high molecular weight polyethylene acetabular cups in hip prostheses produces billions of submicrometer wear particles annually that can cause osteolysis and loosening of the components. Thus, substantial improvement of the wear resistance of ultra-high molecular weight polyethylene could extend the clinical life span of total hip prostheses. It has become apparent that the conditions under which ultra-high molecular weight polyethylene cups have been sterilized can markedly affect their long-term wear properties, and new sterilization methods and other modifications have been developed to minimize the negative effects.
METHODS: In the present study, a hip-joint simulator was used to assess whether it is preferable to sterilize ultra-high molecular weight polyethylene cups without gamma irradiation, to avoid radiation-induced oxidative degradation, or to sterilize with gamma irradiation while the cups are packaged in a suitable low-oxygen atmosphere to minimize oxidation while retaining the increased wear resistance conferred by the radiation-induced cross-linking. Ion-implanted cups and cups made of a highly crystalline polyethylene (Hylamer) also were investigated. Cups made of each material were subjected to wear-testing prior to and after artificial thermal aging to accelerate oxidative degradation.
RESULTS: The results of the present study demonstrated that the cross-linking induced by gamma irradiation improves the wear resistance of ultra-high molecular weight polyethylene, while oxidation reduces it. Without thermal aging, the two types of cups that were sterilized with gamma irradiation while in low-oxygen packaging exhibited about a 50 percent lower rate of wear than did either the nonsterilized cups or the nonirradiated cups sterilized with gas plasma. There was a comparable advantage in the rate of wear after fourteen days of thermal aging. However, after thirty days of aging, the cups sterilized with gamma irradiation in low-oxygen packaging wore several times faster than did the nonirradiated cups. Ion-implanting improved the wear resistance without thermal aging, but after extensive thermal aging the oxidation and wear were greater than those of the controls. Hylamer cups (that is, those that were sterilized with gas plasma) exhibited wear properties very close to those of the nonsterilized ultra-high molecular weight polyethylene cups (the controls) with or without aging.
CONCLUSIONS: Sterilizing an ultra-high molecular weight polyethylene acetabular cup without radiation (for example, with ethylene oxide or gas plasma) avoids immediate and long-term oxidative degradation of the implant but does not improve the inherent wear resistance of the polyethylene. Sterilizing with use of gamma irradiation with the implant packaged in a low-oxygen atmosphere avoids immediate oxidation and cross-links the polyethylene, thereby increasing its wear resistance, but long-term oxidation of the residual free radicals may markedly reduce the wear resistance. Ideally, cross-linking with gamma irradiation to reduce wear should be done in a manner that avoids both immediate and long-term oxidation.

Entities:  

Mesh:

Substances:

Year:  2000        PMID: 11130644     DOI: 10.2106/00004623-200012000-00004

Source DB:  PubMed          Journal:  J Bone Joint Surg Am        ISSN: 0021-9355            Impact factor:   5.284


  36 in total

1.  The John Charnley Award: an accurate and extremely sensitive method to separate, display, and characterize wear debris: part 2: metal and ceramic particles.

Authors:  Fabrizio Billi; Paul Benya; Aaron Kavanaugh; John Adams; Harry McKellop; Edward Ebramzadeh
Journal:  Clin Orthop Relat Res       Date:  2012-02       Impact factor: 4.176

2.  Is routine mid-term total hip arthroplasty surveillance beneficial?

Authors:  James A Keeney; Bradley S Ellison; William J Maloney; John C Clohisy
Journal:  Clin Orthop Relat Res       Date:  2012-06-06       Impact factor: 4.176

3.  Early polyethylene wear and osteolysis with ABG acetabular cups (7- to 12-year follow-up).

Authors:  Sachin Badhe; Peter Livesley
Journal:  Int Orthop       Date:  2005-11-08       Impact factor: 3.075

4.  Oxidation and wear of 100-Mrad cross-linked polyethylene shelf-aged for 30 years.

Authors:  Hironobu Oonishi; Sok Chol Kim; Hiroyuki Oonishi; Masayuki Kyomoto; Shingo Masuda
Journal:  Clin Orthop Relat Res       Date:  2008-01-10       Impact factor: 4.176

5.  Long-term radiographic assessment of cemented polyethylene acetabular cups.

Authors:  Sophie Williams; Graham Isaac; Neil Porter; John Fisher; John Older
Journal:  Clin Orthop Relat Res       Date:  2008-01-10       Impact factor: 4.176

6.  Gamma inert sterilization: a solution to polyethylene oxidation?

Authors:  Francisco J Medel; Steven M Kurtz; William J Hozack; Javad Parvizi; James J Purtill; Peter F Sharkey; Daniel MacDonald; Matthew J Kraay; Victor Goldberg; Clare M Rimnac
Journal:  J Bone Joint Surg Am       Date:  2009-04       Impact factor: 5.284

Review 7.  Periprosthetic osteolysis after total hip replacement: molecular pathology and clinical management.

Authors:  Donald W Howie; Susan D Neale; David R Haynes; Oksana T Holubowycz; Margaret A McGee; Lucian B Solomon; Stuart A Callary; Gerald J Atkins; David M Findlay
Journal:  Inflammopharmacology       Date:  2013-10-15       Impact factor: 4.473

8.  Highly crosslinked polyethylene is safe for use in total knee arthroplasty.

Authors:  Jeffrey T Hodrick; Erik P Severson; Deborah S McAlister; Brian Dahl; Aaron A Hofmann
Journal:  Clin Orthop Relat Res       Date:  2008-09-10       Impact factor: 4.176

9.  THA with the ABG I prosthesis at 15 years. Excellent survival with minimal osteolysis.

Authors:  P N Baker; I A McMurtry; G Chuter; A Port; J Anderson
Journal:  Clin Orthop Relat Res       Date:  2010-07       Impact factor: 4.176

10.  THA with highly cross-linked polyethylene in patients 50 years or younger.

Authors:  Derek S Shia; John C Clohisy; Mark F Schinsky; John M Martell; William J Maloney
Journal:  Clin Orthop Relat Res       Date:  2009-01-14       Impact factor: 4.176

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.