J T Winterer1, K Scheffler, G Paul, M Hauer, O Schäfer, C Altehoefer, J Laubenberger.
Abstract
Our objective was to optimize bolus administration and sequence setting in gadolinium-enhanced magnetic resonance (MR) angiography of the hands. Elliptically reordered three-dimensional (3D) spoiled gradient-echo sequence with non-slab-selective radio frequency excitation was optimized according to the measurements of arterial and venous time-signal curves in 21 patients. Great variations in bolus arrival time and arterio-venous transit time could be observed. In most patients high-quality arterial depiction could be obtained with minor venous contamination. Contrast-to-noise, spatial resolution, and selective arterial filling is still a challenge for 3D MR angiography of the hand but can be optimized using Gadolinium-BOPTA and a dedicated pulse sequence setting with exact bolus timing.Entities:
Mesh:
Substances:
Year: 2000 PMID: 11105711 DOI: 10.1097/00004728-200011000-00017