Results of preliminary clinical trials of the positron emission mammography system PEM-I: a dedicated breast imaging system producing glucose metabolic images using FDG.

K Murthy1, M Aznar, C J Thompson, A Loutfi, R Lisbona, J H Gagnon.   

Abstract

UNLABELLED: Early detection of breast cancer is crucial for efficient and effective treatment. We have developed an instrument for positron emission mammography (PEM) called PEM-I that performs high-resolution metabolic imaging of breast cancer. Images of glucose metabolism are obtained after injection of 75 MBq FDG. The PEM detectors are integrated into a conventional mammography system, allowing acquisition of the emission images immediately after the mammogram, without subject repositioning, and accurate coregistration of images from the 2 modalities. In this article, we present the results of the first clinical pilot study with the instrument.
METHODS: Sixteen subjects (age range, 34-76 y) were studied. All subjects were nondiabetic, nonpregnant, and without a history of cancer. They had recently been found to have suggestive mammography findings or a palpable breast mass and underwent lumpectomy or mastectomy within 2 wk of the study. Results from the PEM study were compared with those from mammography and pathology. A PEM test was classified positive (indicating the presence of cancer) if significant focal uptake was seen in the image or if the counting rate in the breast with suggestive findings was significantly higher than in the contralateral breast.
RESULTS: Of the 16 subjects studied, 14 were evaluable. Ten cancerous tumors and 4 benign tumors were confirmed by pathologic examination after complete removal of the tumor. PEM correctly detected the presence of disease in 8 of 10 subjects. Findings were false-negative in 2 instances and false-positive in none, giving the instrument 80% sensitivity, 100% specificity, and 86% accuracy.
CONCLUSION: Our preliminary results suggest that PEM can offer a noninvasive method for the diagnosis of breast cancer. Metabolic images from PEM contain unique information not available from conventional morphologic imaging techniques and aid in expeditiously establishing the diagnosis of cancer. In all subjects, the PEM images were of diagnostic quality, with an imaging time of 2-5 min.

Entities:  

Mesh:

Substances:

Year:  2000        PMID: 11079494

Source DB:  PubMed          Journal:  J Nucl Med        ISSN: 0161-5505            Impact factor:   10.057


  28 in total

1.  Investigating the limit of detectability of a positron emission mammography device: a phantom study.

Authors:  Nicholas A Shkumat; Adam Springer; Christopher M Walker; Eric M Rohren; Wei T Yang; Beatriz E Adrada; Elsa Arribas; Selin Carkaci; Hubert H Chuang; Lumarie Santiago; Osama R Mawlawi
Journal:  Med Phys       Date:  2011-09       Impact factor: 4.071

Review 2.  Positron emission tomography for benign and malignant disease.

Authors:  Anthony Visioni; Julian Kim
Journal:  Surg Clin North Am       Date:  2011-02       Impact factor: 2.741

3.  Study of the performance of a novel 1 mm resolution dual-panel PET camera design dedicated to breast cancer imaging using Monte Carlo simulation.

Authors:  Jin Zhang; Peter D Olcott; Garry Chinn; Angela M K Foudray; Craig S Levine
Journal:  Med Phys       Date:  2007-02       Impact factor: 4.071

4.  Quantification with a dedicated breast PET/CT scanner.

Authors:  Spencer L Bowen; Andrea Ferrero; Ramsey D Badawi
Journal:  Med Phys       Date:  2012-05       Impact factor: 4.071

5.  Molecular Imaging in Breast Cancer - Potential Future Aspects.

Authors:  Katja Pinker; Wolfgang Bogner; Stephan Gruber; Peter Brader; Siegfried Trattnig; Georgios Karanikas; Thomas H Helbich
Journal:  Breast Care (Basel)       Date:  2011-04-29       Impact factor: 2.860

Review 6.  History and future technical innovation in positron emission tomography.

Authors:  Terry Jones; David Townsend
Journal:  J Med Imaging (Bellingham)       Date:  2017-03-31

Review 7.  Breast cancer imaging: a perspective for the next decade.

Authors:  Andrew Karellas; Srinivasan Vedantham
Journal:  Med Phys       Date:  2008-11       Impact factor: 4.071

Review 8.  Advantages and limitations of FDG PET in the follow-up of breast cancer.

Authors:  Peter Lind; Isabel Igerc; Thomas Beyer; Peter Reinprecht; Klaus Hausegger
Journal:  Eur J Nucl Med Mol Imaging       Date:  2004-04-15       Impact factor: 9.236

9.  Dual-time-point [18F]-FDG PET/CT in the diagnostic evaluation of suspicious breast lesions.

Authors:  M G Caprio; A Cangiano; M Imbriaco; F Soscia; G Di Martino; A Farina; G Avitabile; L Pace; P Forestieri; M Salvatore
Journal:  Radiol Med       Date:  2009-12-16       Impact factor: 3.469

10.  Tumor margin detection using quantitative NIRF molecular imaging targeting EpCAM validated by far red gene reporter iRFP.

Authors:  Banghe Zhu; Grace Wu; Holly Robinson; Nathaniel Wilganowski; Mary A Hall; Sukhen C Ghosh; Kenneth L Pinkston; Ali Azhdarinia; Barrett R Harvey; Eva M Sevick-Muraca
Journal:  Mol Imaging Biol       Date:  2013-10       Impact factor: 3.488

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.