Literature DB >> 11012057

In vitro biomechanical investigation of the stability and stress-shielding effect of lumbar interbody fusion devices.

M Kanayama1, B W Cunningham, C J Haggerty, K Abumi, K Kaneda, P C McAfee.   

Abstract

OBJECT: Interbody fusion devices are rapidly gaining acceptance as a method of ensuring lumbar interbody arthrodesis. Although different types of devices have been developed, the comparative reconstruction stability remains controversial. It also remains unclear how different stress-shielded environments are created within the devices. Using a calf spine model, this study was designed to compare the construct stiffness afforded by 11 differently designed lumbar interbody fusion devices and to quantify their stress-shielding effects by measuring pressure within the devices.
METHODS: Sixty-six lumbar specimens obtained from calves were subjected to anterior interbody reconstruction at L4-5 by using one of the following interbody fusion devices: four different threaded fusion cages (BAK device, BAK Proximity, Ray TFC, and Danek TIBFD), five different nonthreaded fusion devices (oval and circular Harms cages, Brantigan PLIF and ALIF cages, and InFix device); two different types of allograft (femoral ring and bone dowel) were used. Construct stiffness was evaluated in axial compression, torsion, flexion, and lateral bending. Prior to testing, a silicon elastomer was injected into the cages and intracage pressures were measured using pressure needle transducers.
CONCLUSIONS: No statistical differences were observed in construct stiffness among the threaded cages and nonthreaded devices in most of the testing modalities. Threaded fusion cages demonstrated significantly lower intracage pressures compared with nonthreaded cages and structural allografts. Compared with nonthreaded cages and structural allografts, threaded fusion cages afforded equivalent reconstruction stiffness but provided more stress-shielded environment within the devices.

Entities:  

Mesh:

Year:  2000        PMID: 11012057     DOI: 10.3171/spi.2000.93.2.0259

Source DB:  PubMed          Journal:  J Neurosurg        ISSN: 0022-3085            Impact factor:   5.115


  32 in total

Review 1.  [Vertebral body replacement in spine surgery].

Authors:  F Kandziora; K J Schnake; C K Klostermann; N P Haas
Journal:  Unfallchirurg       Date:  2004-05       Impact factor: 1.000

Review 2.  [Interbody metal implants ("cages") for lumbar fusion].

Authors:  G Freiherr von Salis-Soglio; R Scholz; K Seller
Journal:  Orthopade       Date:  2005-10       Impact factor: 1.087

3.  Primary stability of anterior lumbar stabilization: interdependence of implant type and endplate retention or removal.

Authors:  Christian H Flamme; Nadine von der Heide; Caroline Heymann; Christof Hurschler
Journal:  Eur Spine J       Date:  2005-08-10       Impact factor: 3.134

4.  Biomechanical study of a hat type cervical intervertebral fusion cage.

Authors:  Yu-Tong Gu; Lian-Shun Jia; Tong-Yi Chen
Journal:  Int Orthop       Date:  2006-06-09       Impact factor: 3.075

5.  Porous biodegradable lumbar interbody fusion cage design and fabrication using integrated global-local topology optimization with laser sintering.

Authors:  Heesuk Kang; Scott J Hollister; Frank La Marca; Paul Park; Chia-Ying Lin
Journal:  J Biomech Eng       Date:  2013-10-01       Impact factor: 2.097

6.  Assessment of BioPlex interbody fusion device in a sheep lumbar fusion model.

Authors:  Douglas C Fredericks; Anup A Gandhi; Nicole M Grosland; Joseph D Smucker
Journal:  Iowa Orthop J       Date:  2013

Review 7.  Bioresorbable polymers: heading for a new generation of spinal cages.

Authors:  P I J M Wuisman; T H Smit
Journal:  Eur Spine J       Date:  2005-11-15       Impact factor: 3.134

8.  Elastic modulus in the selection of interbody implants.

Authors:  Robert F Heary; Naresh Parvathreddy; Sujitha Sampath; Nitin Agarwal
Journal:  J Spine Surg       Date:  2017-06

9.  Evaluation of the 96/4 PLDLLA polymer resorbable lumbar interbody cage in a long term animal model.

Authors:  Jean Y Lazennec; Abdallah Madi; Marc A Rousseau; Bernard Roger; Gérard Saillant
Journal:  Eur Spine J       Date:  2006-05-31       Impact factor: 3.134

10.  Evolution of Design of Interbody Cages for Anterior Lumbar Interbody Fusion.

Authors:  Kevin Phan; Ralph J Mobbs
Journal:  Orthop Surg       Date:  2016-08       Impact factor: 2.071

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.