Literature DB >> 10966725

Multiphasic injection method for uniform prolonged vascular enhancement at CT angiography: pharmacokinetic analysis and experimental porcine model.

K T Bae1, H Q Tran, J P Heiken.   

Abstract

PURPOSE: To determine if multiphasic injection provides uniform, prolonged vascular contrast medium enhancement at computed tomographic (CT) angiography.
MATERIALS AND METHODS: With a computer-based, compartmental model of the cardiovascular system, theoretic analysis was performed to estimate an injection algorithm for uniform, prolonged vascular enhancement. For algorithm validation, four pigs were scanned after intravenous injection of 50 or 70 mL of contrast medium (282 mg of iodine per milliliter). Uni-, bi-, and multiphasic injection schemes were tested. In most cases, the initial injection rate was 2 mL/sec. In each CT study, 27 dynamic images were acquired every 2 seconds at a fixed mid-abdominal aortic level. Time-enhancement curves were calculated. Injection duration, peak aortic enhancement, and enhancement uniformity (duration of enhancement achieved within 90% of the peak [90% DCE]) were evaluated.
RESULTS: Theoretic and experimental results agreed well. Compared with uniphasic injection, biphasic injection resulted in more prolonged enhancement but generated two enhancement peaks with a valley between, and multiphasic injection yielded more uniform and prolonged enhancement. With 50- and 70-mL multiphasic injections, respectively, injection duration increased by 32% and 51%, peak enhancement decreased by 19% and 18%, and 90% DCE increased by 81% and 94%.
CONCLUSION: Uniform, prolonged vascular enhancement, which is desirable for CT angiography and essential for steady-state quantification of blood volume in organs, can be achieved with multiphasic injection.

Entities:  

Mesh:

Substances:

Year:  2000        PMID: 10966725     DOI: 10.1148/radiology.216.3.r00au43872

Source DB:  PubMed          Journal:  Radiology        ISSN: 0033-8419            Impact factor:   11.105


  19 in total

1.  Optimisation of contrast medium volume and injection-related factors in CT pulmonary angiography: 64-slice CT study.

Authors:  Selma Uysal Ramadan; Pinar Kosar; Iclal Sonmez; Sevilay Karahan; Ugur Kosar
Journal:  Eur Radiol       Date:  2010-05-01       Impact factor: 5.315

2.  64-Slice multidetector row CT angiography of the abdomen: comparison of low versus high concentration iodinated contrast media in a porcine model.

Authors:  N-S Holalkere; K Matthes; S P Kalva; W R Brugge; D V Sahani
Journal:  Br J Radiol       Date:  2010-11-16       Impact factor: 3.039

3.  Optimizing scan timing of hepatic arterial phase by physiologic pharmacokinetic analysis in bolus-tracking technique by multi-detector row computed tomography.

Authors:  Isao Yamaguchi; Eiji Kidoya; Masayuki Suzuki; Hirohiko Kimura
Journal:  Radiol Phys Technol       Date:  2010-09-25

4.  Quantitative prediction of contrast enhancement from test bolus data in cardiac MSCT.

Authors:  Andreas H Mahnken; Annabella Rauscher; Ernst Klotz; Georg Mühlenbruch; Marco Das; Rolf W Günther; Joachim E Wildberger
Journal:  Eur Radiol       Date:  2006-11-18       Impact factor: 5.315

5.  Preliminary examination to determine the suitable contrast material injection protocol for CT angiography of the pelvis and lower extremities with a multidetector row helical scanner.

Authors:  Minoru Honda; Hideharu Sugimoto; Masao Obuchi; Takashi Narisawa
Journal:  Radiat Med       Date:  2006-05

Review 6.  Cardiac CT: coronary arteries and beyond.

Authors:  Andreas H Mahnken; Georg Mühlenbruch; Rolf W Günther; Joachim E Wildberger
Journal:  Eur Radiol       Date:  2006-10-26       Impact factor: 5.315

7.  Effect of varying injection rates of a saline chaser on aortic enhancement in CT angiography: phantom study.

Authors:  Sebastian T Schindera; Rendon C Nelson; Laurens Howle; Eli Nichols; David M DeLong; Elmar M Merkle
Journal:  Eur Radiol       Date:  2008-03-20       Impact factor: 5.315

8.  Multislice CT angiography.

Authors:  U Joseph Schoepf; Christoph R Becker; Lars K Hofmann; Marco Das; Thomas Flohr; Bernd M Ohnesorge; Bernhard Baumert; Joshua Rolnick; Jean M Allen; Vassilios Raptopoulos
Journal:  Eur Radiol       Date:  2003-02-26       Impact factor: 5.315

9.  A new method with variable injection parameters in contrast-enhanced CT: a phantom study for evaluating an aortic peak enhancement.

Authors:  Kazuaki Terasawa; Atsunori Maruyama; Tomohiro Tsukimata
Journal:  Radiol Phys Technol       Date:  2015-05-01

10.  Whole-body computed tomography for multiple traumas using a triphasic injection protocol.

Authors:  Christos Loupatatzis; Sebastian Schindera; Jan Gralla; Hanno Hoppe; Jan Bittner; Ralph Schröder; Sudesh Srivastav; Harald Marcel Bonel
Journal:  Eur Radiol       Date:  2008-02-13       Impact factor: 5.315

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.