UNLABELLED: Iterative reconstruction of SPECT images has recently become clinically available as an alternative to filtered backprojection (FBP). However, there is conflicting evidence on whether iterative reconstruction, such as with the ordered-subsets expectation maximization (OSEM) algorithm, improves diagnostic performance over FBP. The study objective was to determine if the detection and localization of small lesions in simulated thoracic gallium SPECT images are better with OSEM reconstruction than with FBP, both with and without attenuation correction (AC). METHODS: Images were simulated using an analytic projector acting on the mathematic cardiac torso computer phantom. Perfect scatter rejection was assumed. Lesion detection accuracy was assessed using localization receiver operating characteristic methodology. The images were read by 5 nuclear medicine physicians. For each reconstruction strategy and for each observer, data were collected in 2 viewing sessions of 100 images. Two-way ANOVA and, when indicated, the Scheffé multiple comparisons test were applied to check for significant differences. RESULTS: Little difference in the accuracy of detection or localization was seen between FBP with and without AC. OSEM with AC extended the contrast range for accurate lesion detection and localization over that of the other methods investigated. Without AC, no significant difference between OSEM and FBP reconstruction was detected. CONCLUSION: OSEM with AC may improve the detection and localization of thoracic gallium-labeled lesions over FBP reconstruction.
UNLABELLED: Iterative reconstruction of SPECT images has recently become clinically available as an alternative to filtered backprojection (FBP). However, there is conflicting evidence on whether iterative reconstruction, such as with the ordered-subsets expectation maximization (OSEM) algorithm, improves diagnostic performance over FBP. The study objective was to determine if the detection and localization of small lesions in simulated thoracic gallium SPECT images are better with OSEM reconstruction than with FBP, both with and without attenuation correction (AC). METHODS: Images were simulated using an analytic projector acting on the mathematic cardiac torso computer phantom. Perfect scatter rejection was assumed. Lesion detection accuracy was assessed using localization receiver operating characteristic methodology. The images were read by 5 nuclear medicine physicians. For each reconstruction strategy and for each observer, data were collected in 2 viewing sessions of 100 images. Two-way ANOVA and, when indicated, the Scheffé multiple comparisons test were applied to check for significant differences. RESULTS: Little difference in the accuracy of detection or localization was seen between FBP with and without AC. OSEM with AC extended the contrast range for accurate lesion detection and localization over that of the other methods investigated. Without AC, no significant difference between OSEM and FBP reconstruction was detected. CONCLUSION:OSEM with AC may improve the detection and localization of thoracic gallium-labeled lesions over FBP reconstruction.
Authors: O Honda; M Yanagawa; A Inoue; A Kikuyama; S Yoshida; H Sumikawa; K Tobino; M Koyama; N Tomiyama Journal: Br J Radiol Date: 2010-11-16 Impact factor: 3.039
Authors: N F Pereira; H C Gifford; P H Pretorius; T Farncombe; M Smyczynski; R Licho; P Schneider; M A King Journal: IEEE Nucl Sci Symp Conf Rec (1997) Date: 2007
Authors: John M Floberg; Aaron F Struck; Brooke K Peters; Christine J Jaskowiak; Scott B Perlman; Lance T Hall Journal: Am J Nucl Med Mol Imaging Date: 2012-07-10
Authors: Viktor Haase; Katharina Hahn; Harald Schöndube; Karl Stierstorfer; Andreas Maier; Frédéric Noo Journal: Med Phys Date: 2019-12 Impact factor: 4.071
Authors: Nicholas F Pereira; Howard C Gifford; P Hendrik Pretorius; Mark Smyczynski; Robert Licho; Peter Schneider; Troy Farncombe; Michael A King Journal: Med Phys Date: 2008-11 Impact factor: 4.071
Authors: Mark S Smyczynski; Howard C Gifford; Andre Lehovich; Joseph E McNamara; W Paul Segars; Eric A Hoffman; Benjamin M W Tsui; Michael A King Journal: IEEE Trans Nucl Sci Date: 2016-02-15 Impact factor: 1.679
Authors: Andre Lehovich; Philippe P Bruyant; Howard S Gifford; Peter B Schneider; Shayne Squires; Robert Licho; Gene Gindi; Michael A King Journal: IEEE Trans Med Imaging Date: 2009-03-24 Impact factor: 10.048