J S Day1, D J Murdoch, G A Dumas.
Abstract
This paper describes a method for calibrating data from a magnetic tracking device. Position and orientation data were collected in a 1. 6x0.8x1.4m(3) volume using a Polhemus Fastrak((R)) in conjunction with both a long-range and standard transmitter. Position and orientation data were calibrated using a locally linear model based on the position of the measurement. After calibration, the average position and angular errors were less than 1.8cm and 1.2 degrees up to 1.8m from the transmitter for the long-range transmitter. For the standard transmitter, even after calibration, errors increased sharply when the sensor was more than 1.2m from the transmitter. Up to that distance, post-calibration errors were less than 1.2cm and 1. 2 degrees, while up to 1.8m they were below 5cm and 4 degrees. These errors could be further reduced by noise filtering. However, use of the standard transmitter is not recommended at distance greater than 1.2m due to orientation-based effects. It was concluded that for the volume investigated, tracking devices could provide similar three-dimensional accuracy to video systems.Entities:
Mesh:
Year: 2000 PMID: 10828336 DOI: 10.1016/s0021-9290(00)00044-0
Source DB: PubMed Journal: J Biomech ISSN: 0021-9290 Impact factor: 2.712