Literature DB >> 21751065

In vitro validation and reliability study of electromagnetic skin sensors for evaluation of end range of motion positions of the hip.

E A Audenaert1, L Vigneron, T Van Hoof, K D'Herde, G van Maele, D Oosterlinck, C Pattyn.   

Abstract

There is growing evidence that femoroacetabular impingement (FAI) is a probable risk factor for the development of early osteoarthritis in the nondysplastic hip. As FAI arises with end range of motion activities, measurement errors related to skin movement might be higher than anticipated when using previously reported methods for kinematic evaluation of the hip. We performed an in vitro validation and reliability study of a noninvasive method to define pelvic and femur positions in end range of motion activities of the hip using an electromagnetic tracking device. Motion data, collected from sensors attached to the bone and skin of 11 cadaver hips, were simultaneously obtained and compared in a global reference frame. Motion data were then transposed in the hip joint local coordinate systems. Observer-related variability in locating the anatomical landmarks required to define the local coordinate system and variability of determining the hip joint center was evaluated. Angular root mean square (RMS) differences between the bony and skin sensors averaged 3.2° (SD 3.5°) and 1.8° (SD 2.3°) in the global reference frame for the femur and pelvic sensors, respectively. Angular RMS differences between the bony and skin sensors in the hip joint local coordinate systems ranged at end range of motion and dependent on the motion under investigation from 1.91 to 5.81°. The presented protocol for evaluation of hip motion seems to be suited for the 3-D description of motion relevant to the experimental and clinical evaluation of femoroacetabular impingement.

Entities:  

Mesh:

Year:  2011        PMID: 21751065     DOI: 10.1007/s11517-011-0804-5

Source DB:  PubMed          Journal:  Med Biol Eng Comput        ISSN: 0140-0118            Impact factor:   2.602


  27 in total

1.  3D shoulder position measurements using a six-degree-of-freedom electromagnetic tracking device.

Authors:  C.G.M. Meskers; H.M. Vermeulen; J.H. de Groot; F.C.T. van Der Helm; P.M. Rozing
Journal:  Clin Biomech (Bristol, Avon)       Date:  1998-06       Impact factor: 2.063

2.  Squatting exercises in older adults: kinematic and kinetic comparisons.

Authors:  Sean Flanagan; George J Salem; Man-Ying Wang; Serena E Sanker; Gail A Greendale
Journal:  Med Sci Sports Exerc       Date:  2003-04       Impact factor: 5.411

3.  Coordination and timing of spine and hip joints during full body reaching tasks.

Authors:  James S Thomas; Gary E Gibson
Journal:  Hum Mov Sci       Date:  2006-10-18       Impact factor: 2.161

4.  [Polycyclic azines with heteroatoms in the 1- and 3-position. 30. Synthesis of 6,7-dimethoxy-substituted 3,1-benzothiazin-4-ones].

Authors:  M Gütschow; K Heinecke; W Thiel; S Leistner
Journal:  Arch Pharm (Weinheim)       Date:  1991-07       Impact factor: 3.751

5.  Feasibility of using inertial sensors to assess human movement.

Authors:  Kambiz Saber-Sheikh; Elizabeth C Bryant; Charlotte Glazzard; Alicia Hamel; Raymond Y W Lee
Journal:  Man Ther       Date:  2009-07-25

6.  The effects of impingement and dysplasia on stress distributions in the hip joint during sitting and walking: a finite element analysis.

Authors:  Salman Chegini; Martin Beck; Stephen J Ferguson
Journal:  J Orthop Res       Date:  2009-02       Impact factor: 3.494

7.  Application of a magnetic tracking device to kinesiologic studies.

Authors:  K N An; M C Jacobsen; L J Berglund; E Y Chao
Journal:  J Biomech       Date:  1988       Impact factor: 2.712

8.  Accuracy of an electromagnetic tracking device: a study of the optimal range and metal interference.

Authors:  A D Milne; D G Chess; J A Johnson; G J King
Journal:  J Biomech       Date:  1996-06       Impact factor: 2.712

Review 9.  Osseous abnormalities and early osteoarthritis: the role of hip impingement.

Authors:  Michael Tanzer; Nicolas Noiseux
Journal:  Clin Orthop Relat Res       Date:  2004-12       Impact factor: 4.176

10.  Open-configuration MRI study of femoro-acetabular impingement.

Authors:  Mitsuyoshi Yamamura; Hidenobu Miki; Nobuo Nakamura; Masakazu Murai; Hideki Yoshikawa; Nobuhiko Sugano
Journal:  J Orthop Res       Date:  2007-12       Impact factor: 3.494

View more
  3 in total

1.  Brachial plexus 3D reconstruction from MRI with dissection validation: a baseline study for clinical applications.

Authors:  Joris Van de Velde; Stephanie Bogaert; Pieter Vandemaele; Wouter Huysse; Eric Achten; Joris Leijnse; Wilfried De Neve; Tom Van Hoof
Journal:  Surg Radiol Anat       Date:  2015-08-23       Impact factor: 1.246

2.  Reliability and accuracy assessment of Radiation Therapy Oncology Group-endorsed guidelines for brachial plexus contouring.

Authors:  Joris Van de Velde; Tom Vercauteren; Werner De Gersem; Johan Wouters; Katrien Vandecasteele; Philippe Vuye; Frank Vanpachtenbeke; Katharina D'Herde; Ingrid Kerckaert; Wilfried De Neve; Tom Van Hoof
Journal:  Strahlenther Onkol       Date:  2014-04-09       Impact factor: 3.621

3.  Evaluation of range of motion restriction within the hip joint.

Authors:  Glen A Turley; Mark A Williams; Richard M Wellings; Damian R Griffin
Journal:  Med Biol Eng Comput       Date:  2012-12-22       Impact factor: 2.602

  3 in total