The Clarion electrode positioner: temporal bone studies.

J N Fayad1, W Luxford, F H Linthicum.   

Abstract

OBJECTIVE: To study the relationship of the Clarion electrode to the modiolus when using an intracochlear positioner.
BACKGROUND: There are theoretical advantages to positioning a cochlear implant electrode in close proximity to the modiolus. This may allow more focused, discrete fields of electrical current, reducing both requirements to achieve threshold and the channel interactions associated with the simultaneous and nonsimultaneous stimulation of closely spaced electrodes.
METHODS: Ten fresh temporal bones were used to assess the position of the electrode in the scala tympani with the positioner in place. The bones were X-rayed after implantation. The relationship of the electrode to the modiolus was studied by calculating a ratio between the curve assumed by the electrode in relationship to the outer wall of the cochlea. The depth of insertion was evaluated in degrees or number of turns around the modiolus.
RESULTS: The electrode was brought closer to the modiolus and a greater depth of insertion was achieved in all cases with the positioner.
CONCLUSION: The intracochlear positioner is capable of bringing the electrode consistently closer to the neural elements within the modiolus.

Mesh:

Year:  2000        PMID: 10733188     DOI: 10.1016/s0196-0709(00)80013-7

Source DB:  PubMed          Journal:  Am J Otol        ISSN: 0192-9763


  8 in total

1.  Implications of deep electrode insertion on cochlear implant fitting.

Authors:  Mathieu Gani; Gregory Valentini; Alain Sigrist; Maria-Izabel Kós; Colette Boëx
Journal:  J Assoc Res Otolaryngol       Date:  2007-01-11

2.  Considerations for design of future cochlear implant electrode arrays: electrode array stiffness, size, and depth of insertion.

Authors:  Stephen J Rebscher; Alexander Hetherington; Ben Bonham; Peter Wardrop; David Whinney; Patricia A Leake
Journal:  J Rehabil Res Dev       Date:  2008

3.  Partial withdrawal of deeply inserted cochlear electrodes: observations of two patients.

Authors:  Maria-Izabel Kos; Colette Boex; Jean-Philippe Guyot; Marco Pelizzone
Journal:  Eur Arch Otorhinolaryngol       Date:  2007-06-12       Impact factor: 2.503

4.  Cochlear Histopathology as Observed in Two Patients With a Cochlear Implant Electrode With Positioner.

Authors:  Takefumi Kamakura; Joseph B Nadol
Journal:  Otol Neurotol       Date:  2016-07       Impact factor: 2.311

5.  Electrode interaction in pediatric cochlear implant subjects.

Authors:  Marc D Eisen; Kevin H Franck
Journal:  J Assoc Res Otolaryngol       Date:  2005-06-10

6.  Acoustic to electric pitch comparisons in cochlear implant subjects with residual hearing.

Authors:  Colette Boëx; Lionel Baud; Grégoire Cosendai; Alain Sigrist; Maria-Izabel Kós; Marco Pelizzone
Journal:  J Assoc Res Otolaryngol       Date:  2006-02-01

Review 7.  The cochlear implant: historical aspects and future prospects.

Authors:  Adrien A Eshraghi; Ronen Nazarian; Fred F Telischi; Suhrud M Rajguru; Eric Truy; Chhavi Gupta
Journal:  Anat Rec (Hoboken)       Date:  2012-10-08       Impact factor: 2.064

8.  Histopathology of the Clarion cochlear implant electrode positioner in a human subject.

Authors:  M Seyyedi; B J Burgess; D K Eddington; B J Gantz; J B Nadol
Journal:  Audiol Neurootol       Date:  2013-06-14       Impact factor: 1.854

  8 in total