Comparison of long-term variability for standard and short-wavelength automated perimetry in stable glaucoma patients.

E Z Blumenthal1, P A Sample, L Zangwill, A C Lee, Y Kono, R N Weinreb.   

Abstract

PURPOSE: To quantify and compare, on a point-by-point basis, the long-term variability of standard and short-wavelength automated perimetry in a group of stable glaucoma patients.
METHODS: From a group of 53 glaucoma patients experienced in visual field testing, we identified one eye, randomly chosen, from each of 25 glaucoma patients whose condition was found to be stable, based on both standard and short-wavelength automated perimetry visual field criteria. On each of three visits during a period of up to 3 months, each patient performed one standard and one short-wavelength automated perimetry 24-2 visual field in a random order on a Humphrey visual field analyzer. The long-term variability (also referred to as test-retest variability) was defined as the SD of the three threshold decibel values at each test location. The long-term variability for each test point (mean +/- SD) was determined separately for both standard visual fields and short-wavelength automated perimetry.
RESULTS: With all 52 test locations of the 24-2 field averaged, the global long-term variability, mean (+/- SD) for standard visual fields and short-wavelength automated perimetry was 2.37 +/- 2.03 dB (95% confidence interval, 2.26-2.48 dB) and 2.92 +/- 2.03 dB (95% confidence interval, 2.81-3.03 dB), respectively (P <.0001). In 16 of the 52 visual field locations, long-term variability on short-wavelength automated perimetry was significantly higher than long-term variability on standard visual fields. In addition, the long-term variability increased with greater distance from the point of fixation for both standard visual fields and short-wavelength automated perimetry. The long-term variability decreased closer to fixation, more for standard visual fields than for short-wavelength automated perimetry.
CONCLUSIONS: In a group of stable glaucoma patients, mean long-term variability was 0.55 dB higher for short-wavelength automated perimetry than for standard visual fields. This needs to be taken into consideration when serial visual fields are evaluated for change.

Entities:  

Mesh:

Year:  2000        PMID: 10704545     DOI: 10.1016/s0002-9394(99)00432-8

Source DB:  PubMed          Journal:  Am J Ophthalmol        ISSN: 0002-9394            Impact factor:   5.258


  9 in total

1.  Frequency doubling illusion VEPs and automated perimetry in multiple sclerosis.

Authors:  Rasa Ruseckaite; Teddy Maddess; Gytis Danta; Andrew Charles James
Journal:  Doc Ophthalmol       Date:  2006-08-12       Impact factor: 2.379

2.  Evaluation of a two-stage neural model of glaucomatous defect: an approach to reduce test-retest variability.

Authors:  Fei Pan; William H Swanson; Mitchell W Dul
Journal:  Optom Vis Sci       Date:  2006-07       Impact factor: 1.973

3.  Frequency of seeing characteristics of the short wavelength sensitive visual pathway in clinically normal subjects and diabetic patients with focal sensitivity loss.

Authors:  E D Gilmore; C Hudson; R K Nrusimhadevara; P T Harvey
Journal:  Br J Ophthalmol       Date:  2005-11       Impact factor: 4.638

4.  Five-year forecasts of the Visual Field Index (VFI) with binocular and monocular visual fields.

Authors:  Ryo Asaoka; Richard A Russell; Rizwan Malik; David F Garway-Heath; David P Crabb
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2012-12-07       Impact factor: 3.117

5.  Is there evidence for continued learning over multiple years in perimetry?

Authors:  Stuart K Gardiner; Shaban Demirel; Chris A Johnson
Journal:  Optom Vis Sci       Date:  2008-11       Impact factor: 1.973

6.  Modeling the relative influence of fixation and sampling errors on retest variability in perimetry.

Authors:  T Maddess
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2014-07-30       Impact factor: 3.117

7.  Long-term variability in short-wavelength automated perimetry compared to standard perimetry in glaucoma patients and normal subjects.

Authors:  Chintan Malhotra
Journal:  Oman J Ophthalmol       Date:  2009-01

Review 8.  Visual function tests for glaucoma practice - What is relevant?

Authors:  Aparna Rao; Debananda Padhy; Anindita Pal; Avik Kumar Roy
Journal:  Indian J Ophthalmol       Date:  2022-03       Impact factor: 2.969

9.  Long- and Short-Term Variability of Perimetry in Glaucoma.

Authors:  Stuart K Gardiner; William H Swanson; Steven L Mansberger
Journal:  Transl Vis Sci Technol       Date:  2022-08-01       Impact factor: 3.048

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.