Literature DB >> 10667677

CT of metal implants: reduction of artifacts using an extended CT scale technique.

T M Link1, W Berning, S Scherf, U Joosten, A Joist, K Engelke, H E Daldrup-Link.   

Abstract

PURPOSE: The purpose of this work was to use an extended CT scale technique (ECTS) to reduce artifacts due to metal implants and to optimize CT imaging parameters for metal implants using an experimental model.
METHOD: Osteotomies were performed in 20 porcine femur specimens. One hundred cobalt-base screws and 24 steel plates were used for osteosynthesis in these specimens. Artificial lesions were produced in 50 screws, such as osteolysis near the screws (mimicking lysis due to infection, tumor, or loosening), displacement of the screws, as well as fractures of the screws. All specimens were examined using eight different CT protocols: four conventional (CCT) and four spiral (SCT) CT protocols with different milliampere-second values (130 and 480 mAs for CCT, 130 and 300 mAs for SCT), kilovolt potentials (120 and 140 kVp), and slice thicknesses (2 and 5 mm). The images were analyzed by three observers using a standard window (maximum window width 4,000 HU) and ECTS (maximum window width 40,000 HU). Receiver operating characteristic analysis was performed, and image quality was assessed according to a five level scale.
RESULTS: Metal artifacts were significantly reduced using ECTS (p < 0.05). The highest diagnostic performance was obtained using ECTS with the thinnest slice thickness. Metal artifacts were more pronounced using SCT. In this experimental model, exposure dose and kilovolt potential had no significant impact on diagnostic performance (p > 0.05).
CONCLUSION: ECTS improved imaging of metal implants. In this study, no significant effects of exposure dose and kilovolt potential were noted. Metal artifacts were more prominent using SCT than using CCT.

Entities:  

Mesh:

Substances:

Year:  2000        PMID: 10667677     DOI: 10.1097/00004728-200001000-00029

Source DB:  PubMed          Journal:  J Comput Assist Tomogr        ISSN: 0363-8715            Impact factor:   1.826


  30 in total

1.  Reducing the effects of metal artefact using high keV monoenergetic reconstruction of dual energy CT (DECT) in hip replacements.

Authors:  Mark Lewis; Karen Reid; Andoni P Toms
Journal:  Skeletal Radiol       Date:  2012-06-09       Impact factor: 2.199

2.  Metal artefact reduction in gemstone spectral imaging dual-energy CT with and without metal artefact reduction software.

Authors:  Young Han Lee; Kwan Kyu Park; Ho-Taek Song; Sungjun Kim; Jin-Suck Suh
Journal:  Eur Radiol       Date:  2012-02-04       Impact factor: 5.315

3.  Assessment of CAOS as a training model in spinal surgery: a randomised study.

Authors:  P J Richards; I C Kurta; V Jasani; C H Wynn Jones; A Rahmatalla; G Mackenzie; J Dove
Journal:  Eur Spine J       Date:  2006-05-09       Impact factor: 3.134

Review 4.  Periprosthetic osteolysis after total hip replacement: molecular pathology and clinical management.

Authors:  Donald W Howie; Susan D Neale; David R Haynes; Oksana T Holubowycz; Margaret A McGee; Lucian B Solomon; Stuart A Callary; Gerald J Atkins; David M Findlay
Journal:  Inflammopharmacology       Date:  2013-10-15       Impact factor: 4.473

5.  Value of monoenergetic dual-energy CT (DECT) for artefact reduction from metallic orthopedic implants in post-mortem studies.

Authors:  Laura Filograna; Nicola Magarelli; Antonio Leone; Roman Guggenberger; Sebastian Winklhofer; Michael John Thali; Lorenzo Bonomo
Journal:  Skeletal Radiol       Date:  2015-05-12       Impact factor: 2.199

6.  Iterative algorithms for metal artifact reduction in children with orthopedic prostheses: preliminary results.

Authors:  Seema Toso; Meryle Laurent; Elise Dupuis Lozeron; Pauline Brindel; Marirosa Cristallo Lacalamita; Sylviane Hanquinet
Journal:  Pediatr Radiol       Date:  2018-07-28

7.  Measuring femoral lesions despite CT metal artefacts: a cadaveric study.

Authors:  Daniel F Malan; Charl P Botha; Gert Kraaij; Raoul M S Joemai; Huub J L van der Heide; Rob G H H Nelissen; Edward R Valstar
Journal:  Skeletal Radiol       Date:  2011-07-06       Impact factor: 2.199

8.  Monitoring and risk of progression of osteolysis after total hip arthroplasty.

Authors:  Michael D Ries; Thomas M Link
Journal:  J Bone Joint Surg Am       Date:  2012-11-21       Impact factor: 5.284

9.  Effect of brightness in the evaluation of lumbar pedicular screws position: clinical study.

Authors:  L M Romero-Muñoz; M Alfonso; C Villas; J L Zubieta
Journal:  Musculoskelet Surg       Date:  2013-04-05

10.  Metal artifact reduction image reconstruction algorithm for CT of implanted metal orthopedic devices: a work in progress.

Authors:  Patrick T Liu; William P Pavlicek; Mary B Peter; Mark J Spangehl; Catherine C Roberts; Robert G Paden
Journal:  Skeletal Radiol       Date:  2009-01-14       Impact factor: 2.199

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.