Literature DB >> 10611870

Fixation, polyethylene wear, and pelvic osteolysis in primary total hip replacement.

W J Maloney1, J O Galante, M Anderson, V Goldberg, W H Harris, J Jacobs, M Kraay, P Lachiewicz, H E Rubash, S Schutzer, S T Woolson.   

Abstract

A multicenter retrospective review was performed analyzing 1081 primary total hip replacements in 944 patients using the Harris Galante-I cementless acetabular component with screw fixation. All patients were followed up for a minimum of 5 years with a mean followup of 81 months. Linear polyethylene wear averaged 0.11 mm/year (range, 0-0.86 mm/year). Pelvic osteolysis was seen in 25 patients (2.3%). Migration of the acetabular component was seen in four hips. A subgroup of patients was reanalyzed at a minimum followup of 10 years. The mean linear polyethylene wear rate remained 0.11 mm/year. In this group, only one socket had migrated. There was an association between wear rate and age. On average, younger patients had higher wear rates. The risk for having pelvic osteolysis develop and the need for revision surgery also was age-related. Twenty-two percent of hip replacements (15 hips) in patients younger than 50 years of age at the time of their index operation had pelvic osteolysis develop. In contrast, for patients older than 50 years of age at the time of surgery only 7.8% (eight hips) had osteolysis of the pelvis develop. For patients older than 70 years of age at the time of primary total hip replacement, none had pelvic osteolysis develop.

Entities:  

Mesh:

Substances:

Year:  1999        PMID: 10611870     DOI: 10.1097/00003086-199912000-00016

Source DB:  PubMed          Journal:  Clin Orthop Relat Res        ISSN: 0009-921X            Impact factor:   4.176


  23 in total

1.  Is routine mid-term total hip arthroplasty surveillance beneficial?

Authors:  James A Keeney; Bradley S Ellison; William J Maloney; John C Clohisy
Journal:  Clin Orthop Relat Res       Date:  2012-06-06       Impact factor: 4.176

2.  Highly crosslinked polyethylene reduces wear in total hip arthroplasty at 5 years.

Authors:  Jonathan Mutimer; Peter A Devane; Kathryn Adams; J Geoffrey Horne
Journal:  Clin Orthop Relat Res       Date:  2010-12       Impact factor: 4.176

3.  Ten-year follow-up of an anatomical hydroxyapatite-coated total hip prosthesis.

Authors:  V Canales Cortés; J J Panisello Sebastiá; A Herrera Rodríguez; A Peguero Bona; A Martínez Martín; L Herrero Barcos; L García-Dihinx
Journal:  Int Orthop       Date:  2006-02-28       Impact factor: 3.075

Review 4.  Sports activity after total hip and knee arthroplasty : specific recommendations concerning tennis.

Authors:  Thorsten M Seyler; Michael A Mont; Phillip S Ragland; Munaf M Kachwala; Ronald E Delanois
Journal:  Sports Med       Date:  2006       Impact factor: 11.136

5.  Radiographic texture analysis in the characterization of trabecular patterns in periprosthetic osteolysis.

Authors:  Joel R Wilkie; Maryellen L Giger; Charles A Engh; Robert H Hopper; John M Martell
Journal:  Acad Radiol       Date:  2008-02       Impact factor: 3.173

6.  Wear is reduced in THA performed with highly cross-linked polyethylene.

Authors:  Burak Beksaç; Antonio Salas; Alejandro González Della Valle; Eduardo A Salvati
Journal:  Clin Orthop Relat Res       Date:  2008-12-12       Impact factor: 4.176

7.  Durable fixation achieved with medialized, high hip center cementless THAs for Crowe II and III dysplasia.

Authors:  Danyal H Nawabi; Morteza Meftah; Denis Nam; Amar S Ranawat; Chitranjan S Ranawat
Journal:  Clin Orthop Relat Res       Date:  2014-02       Impact factor: 4.176

8.  THA with highly cross-linked polyethylene in patients 50 years or younger.

Authors:  Derek S Shia; John C Clohisy; Mark F Schinsky; John M Martell; William J Maloney
Journal:  Clin Orthop Relat Res       Date:  2009-01-14       Impact factor: 4.176

9.  Similar wear in total hip arthroplasties with metallic or zirconia femoral heads.

Authors:  Randy M Cohn; Alejandro González Della Valle; Margaret Peterson; Charles N Cornell
Journal:  HSS J       Date:  2008-07-15

10.  Polyethylene in total hip arthroplasty: half a century in the limelight.

Authors:  Eduardo García-Rey; Eduardo García-Cimbrelo
Journal:  J Orthop Traumatol       Date:  2010-05-27