| Literature DB >> 25435806 |
Mao-Kai Chen1, Hung-Te Hsu1, I-Cheng Lu2, Chih-Kai Shih3, Ya-Chun Shen1, Kuang-Yi Tseng1, Kuang-I Cheng4.
Abstract
BACKGROUND: Many tools have been developed to facilitate the insertion of the ProSeal laryngeal mask airway (LMA) insertion, which can be impeded by folding of its soft cuff. The aim of this study was to compare the efficiency of ProSeal LMA insertion guided by a soft, direct optical Foley Airway Stylet Tool (FAST) with the standard introducer tool (IT).Entities:
Keywords: Fiberoptic bronchoscope; Foley Airway Stylet Tool (FAST); Introducer tool (IT); ProSeal laryngeal mask airway (LMA)
Mesh:
Year: 2014 PMID: 25435806 PMCID: PMC4247224 DOI: 10.1186/1471-2253-14-105
Source DB: PubMed Journal: BMC Anesthesiol ISSN: 1471-2253 Impact factor: 2.217
Figure 1Consolidated Standards of Reporting Trials (CONSORT) diagram for the study.
Figure 2Settings of the Foley Airway Stylet Tool® (FAST) in a ProSeal LMA. The atraumatic tip of the stylet can be seen protruding from the distal end of the drainage tube; the flexible portion of the stylet extends 5–6 cm beyond the cuff.
Grading of the views of the laryngeal field from the distal aperture of the airway tube and the esophagus from the drain tube using fiberoptic bronchoscopy
| Viewing | Grading | Description |
|---|---|---|
| Laryngeal field |
|
|
|
|
| |
|
|
| |
|
|
| |
|
|
| |
| Esophagus |
|
|
|
|
| |
|
|
|
Patient characteristics, ASA classification, LMA size, and duration and specialty of surgery
| FAST group (n = 69) | IT group (n = 75) | P values | |
|---|---|---|---|
| Male/female | 16/53 | 28/47 | 0.066 |
| Age (yr)# | 41 (20-65) | 43 (22-65) | 0.452 |
| Height (cm) | 159.5 (8.4) | 160.7 (8.0) | 0.360 |
| Weight (kg) | 61.4 (13.3) | 62.6 (11.2) | 0.573 |
| BMI | 24.0 (4.0) | 24.2 (3.5) | 0.863 |
| ASA I/ II | 35/34 | 31/44 | 0.259 |
| LMA size NO:3/NO:4 | 26/43 | 23/52 | 0.375 |
| Duration of surgery (min) | 117 (53) | 120 (51) | 0.717 |
| Type of surgery | |||
| Otorhinolaryngology | 3 | 1 | 0.182 |
| Gynecology | 29 | 26 | |
| Orthopedic | 21 | 29 | |
| General | 16 | 15 | |
| Urology | 0 | 4 |
#Median [95% confidence interval]; BMI: body mass index; ASA, American Society of Anesthesiologists; SD: standard deviation.
All data are shown as mean (SD) unless otherwise stated.
Hemodynamic responses to ProSeal laryngeal mask airway insertion
| FAST group | -Tool group | P value | |
|---|---|---|---|
| Number of patients | 66 | 74 | |
| Heart rate | |||
| Before | 80.8 (14.1) | 79.1 (15.6) | 0.97 |
| After | 77.8 (12.8) | 75.5 (14.7) | 0.33 |
| Mean arterial pressure | |||
| Before | 99.2 (13.9) | 99.1 (15.4) | 0.82 |
| After | 101.1 (15.4) | 98.3 (17.8) | 0.27 |
ProSeal LMA insertion parameters and incidence of postoperative sore throat
| FAST group | IT group | P value | |
|---|---|---|---|
| Insertion time | 17.4 ± 6.1 | 12.6 ± 4.7 | <0.001*** |
| Insertion condition | |||
| Smooth | 62 (89.9%) | 67 (89.4%) | |
| Mild resistance | 2 (2.9%) | 1 (1.3%) | 0.413 |
| Two attempts | 2 (2.9%) | 6 (8.0%) | |
| Failed insertion# | 3 (4.3%) | 1 (1.3%) | |
| Viewing glottic field | |||
| 1 | 0 (0%) | 0 (0%) | 0.201 |
| 2 | 4 (6.1%) | 2 (2.7%) | |
| 3 | 42 (63.6%) | 57 (77.0%) | |
| 4 | 20 (30.3%) | 15 (20.3%) | |
| 5 | 0 (0%) | 0 (0%) | |
| Viewing esophagus | |||
| A | 59 (89.4%) | 69 (93.2%) | 0.435 |
| B | 6 (9.1%) | 3 (4.1%) | |
| C | 1 (1.5%) | 2 (2.7%) | |
| Sore throat in PACU | |||
| None | 44 (66.7%) | 56 (75.7%) | 0.377 |
| Mild | 20 (30.1.1%) | 18 (24.3%) | |
| Moderate | 1 (1.5%) | 0 (0%) | |
| Severe | 1 (1.5%) | 0 (0%) | |
| Sore throat on the first | |||
| Postoperative morning | |||
| None | 52 (78.8%) | 69 (93.2%) | 0.035* |
| Mild | 11 (16.7%) | 3 (4.1%) | |
| Moderate | 3 (4.5%) | 2 (2.7%) | |
| Severe | 0 (0%) | 0 (0%) |
PACU, post-anesthesia care unit; #: Due to failed insertion, ProSeal LMA post-insertion assessments were excluded. *: P< 0.05; ***: P< 0.001.