Effect of stenting on coronary flow velocity reserve: comparison of coil and tubular stents.

C J Vrints1, M J Claeys, J Bosmans, V Conraads, J P Snoeck.   

Abstract

OBJECTIVE: To determine whether coil stents are as effective as tubular stents in improving coronary flow velocity reserve (CFVR) after stent deployment.
METHODS: Distal CFVR was measured with a 0. 014 inch Doppler guide wire before and after stenting in 33 patients. A coil stent was implanted in 16 patients and a tubular stent was used in 17 patients. Coronary flow velocity within the stent was also recorded during a slow pullback.
RESULTS: Following placement of the stents, the percentage diameter stenosis was similar for both the tubular and coil stents (mean (SE) 11 (2)% v 13 (2)%, NS). However, distal CFVR was higher after stenting with a tubular stent compared with a coil stent (2.46 (0.13) v 1.96 (0.14), p < 0.05). Furthermore, pullback through the stent detected a major flow velocity increase within coil stents but not in tubular stents (83 (24)% v 5 (5)%, p < 0.05).
CONCLUSIONS: In spite of similar angiographic improvement, placement of coil stents was associated with inferior functional results compared with tubular stents. The flow velocity acceleration within the coil stents suggests the presence of a residual narrowing within the stent, which is not appreciated on angiography.

Entities:  

Mesh:

Year:  1999        PMID: 10490562      PMCID: PMC1760279          DOI: 10.1136/hrt.82.4.465

Source DB:  PubMed          Journal:  Heart        ISSN: 1355-6037            Impact factor:   5.994


  21 in total

1.  Validation of a Doppler guide wire for intravascular measurement of coronary artery flow velocity.

Authors:  J W Doucette; P D Corl; H M Payne; A E Flynn; M Goto; M Nassi; J Segal
Journal:  Circulation       Date:  1992-05       Impact factor: 29.690

2.  The effect of coronary angioplasty on coronary flow reserve.

Authors:  R F Wilson; M R Johnson; M L Marcus; P E Aylward; D J Skorton; S Collins; C W White
Journal:  Circulation       Date:  1988-04       Impact factor: 29.690

3.  Analysis of coronary blood flow velocity dynamics in angiographically normal and stenosed arteries before and after endolumen enlargement by angioplasty.

Authors:  E O Ofili; M J Kern; A J Labovitz; J A St Vrain; J Segal; F V Aguirre; R Castello
Journal:  J Am Coll Cardiol       Date:  1993-02       Impact factor: 24.094

4.  Intracoronary blood flow velocity and transstenotic pressure gradient using sensor-tip pressure and Doppler guidewires: a new technology for the assessment of stenosis severity in the catheterization laboratory.

Authors:  C Di Mario; P J de Feyter; C J Slager; P de Jaegere; J R Roelandt; P W Serruys
Journal:  Cathet Cardiovasc Diagn       Date:  1993-04

5.  Measurement of volumetric coronary blood flow by simultaneous intravascular two-dimensional and Doppler ultrasound: validation in an animal model.

Authors:  T M Chou; K Sudhir; S Iwanaga; K Chatterjee; P G Yock
Journal:  Am Heart J       Date:  1994-08       Impact factor: 4.749

6.  Prognostic value of intracoronary flow velocity and diameter stenosis in assessing the short- and long-term outcomes of coronary balloon angioplasty: the DEBATE Study (Doppler Endpoints Balloon Angioplasty Trial Europe).

Authors:  P W Serruys; C di Mario; J Piek; E Schroeder; C Vrints; P Probst; B de Bruyne; C Hanet; E Fleck; M Haude; E Verna; V Voudris; H Geschwind; H Emanuelsson; V Mühlberger; G Danzi; H O Peels; A J Ford; E Boersma
Journal:  Circulation       Date:  1997-11-18       Impact factor: 29.690

7.  Alterations of phasic coronary artery flow velocity in humans during percutaneous coronary angioplasty.

Authors:  J Segal; M J Kern; N A Scott; S B King; J W Doucette; R R Heuser; E Ofili; R Siegel
Journal:  J Am Coll Cardiol       Date:  1992-08       Impact factor: 24.094

8.  Experimental basis of determining maximum coronary, myocardial, and collateral blood flow by pressure measurements for assessing functional stenosis severity before and after percutaneous transluminal coronary angioplasty.

Authors:  N H Pijls; J A van Son; R L Kirkeeide; B De Bruyne; K L Gould
Journal:  Circulation       Date:  1993-04       Impact factor: 29.690

9.  Delayed recovery of coronary resistive vessel function after coronary angioplasty.

Authors:  N G Uren; T Crake; D C Lefroy; R de Silva; G J Davies; A Maseri
Journal:  J Am Coll Cardiol       Date:  1993-03-01       Impact factor: 24.094

10.  Limitations of the zero crossing detector in the analysis of intracoronary Doppler: a comparison with fast Fourier transform analysis of basal, hyperemic, and transstenotic blood flow velocity measurements in patients with coronary artery disease.

Authors:  C Di Mario; J R Roelandt; P de Jaegere; D T Linker; J Oomen; P W Serruys
Journal:  Cathet Cardiovasc Diagn       Date:  1993-01
View more
  1 in total

1.  Influence of Artery Straightening on Local Hemodynamics in Left Anterior Descending (LAD) Artery after Stent Implantation.

Authors:  Pengfei Liu; Xiaoyan Deng; Xiao Liu; Anqiang Sun; Hongyan Kang
Journal:  Cardiol Res Pract       Date:  2020-05-22       Impact factor: 1.866

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.