Haemodynamic performance of aortic pericardial bioprostheses and bileaflet prostheses at rest and during exercise: implications for the surgical management of patients with small aortic roots.

J R González-Juanatey1, M V Fernández, F G Sampedro, J M García-Acuña, J B García-Bengoechea, A A Cendon, M G de La Peña.   

Abstract

OBJECTIVE: To determine the haemodynamic behaviour, at rest and during exercise, of aortic valve pericardial bioprostheses and different sizes of bileaflet prosthesis.
DESIGN: Observational study.
SETTING: Tertiary medical centre. PATIENTS AND
INTERVENTIONS: 74 patients (33 women, 41 men; mean age 64 years) in whom 40 pericardial bioprostheses and 34 bileaflet prostheses sized 19, 21, or 23 mm had been implanted to replace aortic valves. MAIN OUTCOME MEASURES: Doppler echocardiography at rest and at peak exercise, between 12 and 47 months after surgery.
RESULTS: All patients achieved a significant increase in heart rate, systolic blood pressure, and cardiac output with exercise. Transvalvar pressure fall, valve area, and left ventricular systolic and diastolic function indices also underwent significant changes with exercise. Reductions in peak and mean transvalvar pressure, at rest and at peak exercise, were greater in patients with small valves (p < 0.05). Valve areas and effective area index were greater in the patients with larger valves (p < 0.001). There were no significant differences between patients with mechanical and biological prostheses with regard to transvalvar pressure fall and valve areas at rest and at peak exercise.
CONCLUSIONS: 19 mm and 21 mm aortic prostheses and bioprostheses continue to create significant obstruction, particularly with exercise.

Entities:  

Mesh:

Year:  1999        PMID: 10409527      PMCID: PMC1729145          DOI: 10.1136/hrt.82.2.149

Source DB:  PubMed          Journal:  Heart        ISSN: 1355-6037            Impact factor:   5.994


  21 in total

1.  Echocardiographic determination of left ventricular mass in man. Anatomic validation of the method.

Authors:  R B Devereux; N Reichek
Journal:  Circulation       Date:  1977-04       Impact factor: 29.690

2.  Doppler echocardiographic assessment of the St. Jude Medical prosthetic valve in the aortic position using the continuity equation.

Authors:  E R Chafizadeh; W A Zoghbi
Journal:  Circulation       Date:  1991-01       Impact factor: 29.690

3.  Hemodynamic evaluation of the CarboMedics prosthetic heart valve in the aortic position: comparison of noninvasive and invasive techniques.

Authors:  H Ihlen; P Mølstad; S Simonsen; K Vatne; E Ovrum; O Geiran; P Laake; T Frøysaker
Journal:  Am Heart J       Date:  1992-01       Impact factor: 4.749

4.  Cardiac output measured by Doppler echocardiography in patients with aortic prosthetic valves.

Authors:  H Ihlen; P Mølstad
Journal:  Eur Heart J       Date:  1990-05       Impact factor: 29.983

5.  Rest and exercise evaluation of St. Jude Medical and Medtronic Hall prostheses. Influence of primary lesion, valvular type, valvular size, and left ventricular function.

Authors:  S Tatineni; H B Barner; A C Pearson; D Halbe; R Woodruff; A J Labovitz
Journal:  Circulation       Date:  1989-09       Impact factor: 29.690

6.  Echocardiographic description of the CarboMedics bileaflet prosthetic heart valve.

Authors:  J Chambers; J Cross; P Deverall; E Sowton
Journal:  J Am Coll Cardiol       Date:  1993-02       Impact factor: 24.094

7.  Clinical and hemodynamic performance of the Ionescu-Shiley valve in the small aortic root. Results in 117 patients with 17 and 19 mm valves.

Authors:  R M Bojar; J T Diehl; M Moten; D D Payne; H Rastegar; J J Stetz; N G Pandian; R J Cleveland
Journal:  J Thorac Cardiovasc Surg       Date:  1989-12       Impact factor: 5.209

8.  St. Jude Medical cardiac valves in small aortic roots: follow-up to sixteen years.

Authors:  D Sawant; A K Singh; W C Feng; A A Bert; F Rotenberg
Journal:  J Thorac Cardiovasc Surg       Date:  1997-03       Impact factor: 5.209

9.  The risk of small St. Jude aortic valve prostheses.

Authors:  J M Kratz; R M Sade; F A Crawford; A J Crumbley; M R Stroud
Journal:  Ann Thorac Surg       Date:  1994-05       Impact factor: 4.330

10.  Accurate noninvasive quantification of stenotic aortic valve area by Doppler echocardiography.

Authors:  W A Zoghbi; K L Farmer; J G Soto; J G Nelson; M A Quinones
Journal:  Circulation       Date:  1986-03       Impact factor: 29.690

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.