Literature DB >> 10379979

Computer-assisted spine surgery.

P Merloz1, J Tonetti, L Pittet, M Coulomb, S Lavallée, J Troccaz, P Cinquin, P Sautot.   

Abstract

The aim of this study was to improve the reliability of pedicle screw insertion. Transpedicle screw insertion may cause neurological, vascular, and mechanical complications. Previous studies of surgical procedures have shown a significant rate of incorrect placement of the screw ranging from 10 to 40%. A new technique that combines preoperative computed tomography (CT) imaging with intraoperative passive navigation was used to perform 64 pedicle screw insertions in the thoracolumbar region. At the same time, 64 pedicle screw insertions were performed manually in the same region and on the same vertebral levels. Surgery was followed in all cases by postoperative radiographs and computed tomography examination, which allowed measurements of screw position relative to pedicle position to be performed. A comparison between the two groups showed that six screws in 64 vertebra (9%) had incorrect placement with the computer-assisted technique whereas 28 screws in 64 vertebra (44%) had incorrect placement with manual insertion. The intraoperative accuracy provided by the computer after registration was better than 1 mm. The good results obtained are similar to those reported in the literature. The cortex penetration observed with the computer-assisted technique was not imputed to computer failure. Errors by the surgeon in acquiring data in the pre- and perioperative steps may explain the six incorrect screw placements. This clinical experience confirms that the accuracy and the reliability of this computer-assisted technique are good.

Entities:  

Mesh:

Year:  1998        PMID: 10379979     DOI: 10.1002/(SICI)1097-0150(1998)3:6<297::AID-IGS3>3.0.CO;2-8

Source DB:  PubMed          Journal:  Comput Aided Surg        ISSN: 1092-9088


  21 in total

1.  [Precision of navigation-assisted surgery of the thoracic and lumbar spine].

Authors:  M Arand; M Schempf; D Hebold; S Teller; L Kinzl; F Gebhard
Journal:  Unfallchirurg       Date:  2003-11       Impact factor: 1.000

2.  Computer-assisted posterior instrumentation of the cervical and cervico-thoracic spine.

Authors:  Marcus Richter; Thomas Mattes; Balkan Cakir
Journal:  Eur Spine J       Date:  2003-11-22       Impact factor: 3.134

3.  [Accuracy of CT-based navitation of pedicle screws in the thoracic spine compared with conventional technique].

Authors:  K J Schnake; B König; U Berth; R J Schroeder; F Kandziora; U Stöckle; M Raschke; N P Haas
Journal:  Unfallchirurg       Date:  2004-02       Impact factor: 1.000

4.  Factors influencing the accuracy of iliosacral screw placement in trauma patients.

Authors:  Lars Grossterlinden; Johannes Rueger; Philip Catala-Lehnen; Martin Rupprecht; Wolfgang Lehmann; Andreas Rücker; Daniel Briem
Journal:  Int Orthop       Date:  2010-07-20       Impact factor: 3.075

5.  Accuracy of biopsy needle navigation using the Medarpa system--computed tomography reality superimposed on the site of intervention.

Authors:  M Fawad Khan; Selami Dogan; Adel Maataoui; Jessen Gurung; Mirko Schiemann; Hanns Ackermann; Stefan Wesarg; Georgios Sakas; Thomas J Vogl
Journal:  Eur Radiol       Date:  2005-03-08       Impact factor: 5.315

6.  [A-mode ultrasonic pointer for navigated pelvic surgery].

Authors:  M Oszwald; M Citak; J Kowal; C Amstutz; D Kendoff; T Kirchhoff; L P Nolte; C Krettek; T Hüfner
Journal:  Unfallchirurg       Date:  2008-03       Impact factor: 1.000

Review 7.  Computer-assisted Surgery for Scaphoid Fracture.

Authors:  Zi-Run Xiao; Ge Xiong
Journal:  Curr Med Sci       Date:  2018-12-07

Review 8.  Functional outcome of computer-assisted spinal pedicle screw placement: a systematic review and meta-analysis of 23 studies including 5,992 pedicle screws.

Authors:  Rajeev Verma; Sonal Krishan; Kurt Haendlmayer; A Mohsen
Journal:  Eur Spine J       Date:  2010-01-06       Impact factor: 3.134

9.  [Best matching. Experimental comparison of different matching procedures for use in computer navigation].

Authors:  A Schäffler; B König; N P Haas; U Stöckle
Journal:  Unfallchirurg       Date:  2009-09       Impact factor: 1.000

Review 10.  Image-guided spine surgery: state of the art and future directions.

Authors:  Thorsten Tjardes; Sven Shafizadeh; Dieter Rixen; Thomas Paffrath; Bertil Bouillon; Eva S Steinhausen; Holger Baethis
Journal:  Eur Spine J       Date:  2009-09-11       Impact factor: 3.134