VDD pacemaker replacement is safe and reliable independently of the previously implanted lead: a prospective and controlled study.

Alessandro Fabiani1, Antonio Burali, Enrico Manfredini, Giorgio Corbucci, Leonardo Bolognese, S Serge Barold.   

Abstract

UNLABELLED: Theoretically, replacement of a VDD device requires using a similar pacemaker to provide the best match between the filtering characteristics of the pacemaker and the atrial dipole of the lead. This study evaluated the performance of newly implanted Vitatron VDD pacemakers connected to dedicated leads and compared the results with those of the same Vitatron pacemakers used as replacement but connected to a variety of nondedicated leads.
METHODS: 23 consecutive patients (15 M, 8 F, 78 +/- 6 years) in Group 1 underwent pacemaker replacement with a VDD(R) Saphir 3 (Vitatron BV, Arnhem, The Netherlands) device designed for an 8.6 mm atrial dipole. Atrial dipoles of the previously implanted leads ranged from 5 to 30 mm. Another 22 consecutive patients (14 M, 8 F, 80 +/- 7 years) in Group 2 received a Saphir 3 pacemaker with the related dedicated lead.
RESULTS: P-wave amplitude measured by the same Pacing Sensing Analyzer (PSA) at the first implantation was 1.7 +/- 0.8 vs. 1.7 +/- 0.5 mV (P = NS) in groups 1 and 2 respectively. P-wave amplitudes measured at 1 month follow-up after replacement in Group 1 and at 1 month follow-up after implantation in Group 2 were 0.69 +/- 0.5 vs. 0.85 +/- 0.3 mV (P = NS) respectively. The percentage of atrial sensing at the same follow-up was 97 +/- 3 vs. 95 +/- 5% (P = NS) in groups 1 and 2 respectively.
CONCLUSIONS: Replacement of VDD pacemakers with the Saphir 3 model designed for a short dipole is safe and reliable when used in combination with previously implanted nondedicated leads.

Entities:  

Mesh:

Year:  2006        PMID: 16755339     DOI: 10.1007/s10840-006-7743-3

Source DB:  PubMed          Journal:  J Interv Card Electrophysiol        ISSN: 1383-875X            Impact factor:   1.900


  13 in total

1.  Comparative three-year performance of chronic atrial sensing among eight different VDD systems.

Authors:  John A Chiladakis; Nikolaos Patsouras; George Agelopoulos; Nikolaos Nikolopoulos; Antonis S Manolis
Journal:  Am J Cardiol       Date:  2002-05-15       Impact factor: 2.778

2.  Sensing characteristics of unipolar and bipolar orthogonal floating atrial electrodes: morphology and spectral analysis.

Authors:  A E Aubert; H Ector; B G Denys; H De Geest
Journal:  Pacing Clin Electrophysiol       Date:  1986-05       Impact factor: 1.976

3.  Toward optimizing the detection of atrial depolarization with floating bipolar electrodes.

Authors:  R R Brownlee
Journal:  Pacing Clin Electrophysiol       Date:  1989-03       Impact factor: 1.976

4.  Characteristics of intracardiac electrograms II: Atrial endocardial electrograms.

Authors:  V Parsonnet; G H Myers; Y M Kresh
Journal:  Pacing Clin Electrophysiol       Date:  1980-07       Impact factor: 1.976

5.  Atrial sensing and AV synchrony in single lead VDD pacemakers: a prospective comparison to DDD devices with bipolar atrial leads.

Authors:  U K Wiegand; F Bode; R Schneider; G Taubert; A Brandes; W Peters; H A Katus; J Potratz
Journal:  J Cardiovasc Electrophysiol       Date:  1999-04

6.  Long-term reliability of single lead atrial synchronous pacing systems using closely spaced atrial dipoles: five-year experience.

Authors:  L Ansani; G F Percoco; G Guardigli; T Toselli; G E Antonioli
Journal:  Pacing Clin Electrophysiol       Date:  1994-11       Impact factor: 1.976

7.  Avoiding atrial undersensing by assessment of P wave amplitude histogram data.

Authors:  W Boute; B A Albers; V Giele
Journal:  Pacing Clin Electrophysiol       Date:  1994-11       Impact factor: 1.976

Review 8.  Single lead atrial synchronous ventricular pacing: a dream come true.

Authors:  G E Antonioli
Journal:  Pacing Clin Electrophysiol       Date:  1994-09       Impact factor: 1.976

9.  Development of sinus node disease in patients with AV block: implications for single lead VDD pacing.

Authors:  U K Wiegand; F Bode; R Schneider; A Brandes; H Haase; H A Katus; J Potratz
Journal:  Heart       Date:  1999-06       Impact factor: 5.994

10.  Implantation strategy of the atrial dipole impacts atrial sensing performance of single lead VDD pacemakers.

Authors:  Uwe K H Wiegand; Bernd Nowak; Udo Reisp; Torsten Peiffer; Frank Bode; Jürgen Potratz
Journal:  Pacing Clin Electrophysiol       Date:  2002-03       Impact factor: 1.976

View more